Estimation of the aortic pressure waveform and beat-to-beat relative cardiac output changes from multiple peripheral artery pressure waveforms

被引:17
作者
Swamy, Gokul [1 ]
Mukkamala, Ramakrishna [1 ]
机构
[1] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
关键词
aortic pressure (AP); arterial tree; blood pressure; cardiac output (CO); generalized transfer function; hemodynamics; multichannel blind system identification (MBSI); peripheral artery pressure (PAP); wave reflections; Windkessel model;
D O I
10.1109/TBME.2007.913408
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We introduce a patient- and time-specific technique to estimate the clinically more relevant aortic pressure (AP) waveform and beat-to-beat relative changes in cardiac output (CO) from multiple peripheral artery pressure (PAP) waveforms distorted by wave reflections. The basic idea of the technique is to first estimate the AP waveform by applying a new multichannel blind system identification method that we have developed (rather than the conventional generalized transfer function) to the PAP waveforms and then estimate the beat-to-beat proportional CO by fitting a Windkessel model to the estimated waveform in which wave distortion should be attenuated. We present an evaluation of the technique with respect to four swine datasets including simultaneous measurements of two peripheral AP waveforms, a reference AP waveform, and reference aortic flow probe CO during diverse hemodynamic interventions. Our results show an overall AP waveform error of 3.5 mmHg and an overall beat-to-beat CO error of 12.9% (after a single CO calibration in each animal). These estimation errors represent substantial improvements compared to those,obtained with several alternative PAP waveform analysis techniques. With further successful testing, the new technique may ultimately be employed for automated and less invasive monitoring of central hemodynamics in various cardiovascular patients.
引用
收藏
页码:1521 / 1529
页数:9
相关论文
共 31 条
[1]  
[Anonymous], 1998, The ICU Book
[2]   REPEATED REFLECTION OF WAVES IN THE SYSTEMIC ARTERIAL SYSTEM [J].
BERGER, DS ;
LI, JKJ ;
LASKEY, WK ;
NOORDERGRAAF, A .
AMERICAN JOURNAL OF PHYSIOLOGY, 1993, 264 (01) :H269-H281
[3]   TRANSFER-FUNCTION ANALYSIS OF AUTONOMIC REGULATION .1. CANINE ATRIAL RATE RESPONSE [J].
BERGER, RD ;
SAUL, JP ;
COHEN, RJ .
AMERICAN JOURNAL OF PHYSIOLOGY, 1989, 256 (01) :H142-H152
[4]   CHARACTERISTICS OF AORTIC DIASTOLIC PRESSURE DECAY WITH APPLICATION TO CONTINUOUS MONITORING OF CHANGES IN PERIPHERAL VASCULAR-RESISTANCE [J].
BOURGEOIS, MJ ;
GILBERT, BK ;
DONALD, DE ;
WOOD, EH .
CIRCULATION RESEARCH, 1974, 35 (01) :56-66
[5]   CONTINUOUS DETERMINATION OF BEAT-TO-BEAT STROKE VOLUME FROM AORTIC PRESSURE PULSES IN DOG [J].
BOURGEOIS, MJ ;
GILBERT, BK ;
BERNUTH, GV ;
WOOD, EH .
CIRCULATION RESEARCH, 1976, 39 (01) :15-24
[6]  
Chen CH, 1997, CIRCULATION, V95, P1827
[7]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[8]  
DONALD AM, 1960, BLOOD FLOW ARTERIES
[9]   Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry [J].
Fetics, B ;
Nevo, E ;
Chen, CH ;
Kass, DA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (06) :698-706
[10]  
Guyton AC., 1996, TXB MED PHYSL