A durable, low-cost electrogoniometer for dynamic measurement of joint trajectories

被引:43
作者
Wang, Po T. [2 ]
King, Christine E. [2 ]
Do, An H. [1 ]
Nenadic, Zoran [2 ,3 ]
机构
[1] Univ Calif Irvine, Med Ctr, Dept Neurol, Orange, CA 92868 USA
[2] Univ Calif Irvine, Dept Biomed Engn, Irvine, CA 92697 USA
[3] Univ Calif Irvine, Dept Elect Engn & Comp Sci, Irvine, CA 92697 USA
关键词
Articular goniometry; Transducers; Applied kinesiology; Joint range of motion; Regression analysis; GONIOMETER;
D O I
10.1016/j.medengphy.2010.12.008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This article introduces a method and step-by-step instructions for the design of a low-cost, flexible electrogoniometer, suitable for kinesiology, rehabilitation, and biometric applications. Two unidirectional flexible sensors are placed back-to-back, and a multivariate linear regression model was used to combine measurements from the two sensors. Following a short calibration procedure, the electrogoniometer can be reliably used for measurement of flexion/extension angles of various hinge joints. The performance of the goniometer has been tested on a population of 21 healthy subjects performing flexion/extension of index finger, wrist and elbow. The proposed device achieves the quality of joint angle measurements comparable to that of commercial electrogoniometers, while having a significantly higher durability-to-cost ratio. (C) 2010 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:546 / 552
页数:7
相关论文
共 14 条
[1]  
Assal M, 2003, J REHABIL RES DEV, V40, P235
[2]  
BEVINGTON PR, 1992, DATA REDUCTION ERROR, P101
[3]  
*BIOM LTD, 2010, GON TORS
[4]  
*BIOP SYST INC, 2009, FING GON
[5]  
Gentile CT, 1992, US Patent, Patent No. [5086785, 5 086 785]
[6]   ROBUST REGRESSION USING ITERATIVELY RE-WEIGHTED LEAST-SQUARES [J].
HOLLAND, PW ;
WELSCH, RE .
COMMUNICATIONS IN STATISTICS PART A-THEORY AND METHODS, 1977, 6 (09) :813-827
[7]   INDEXES OF RELATIVE WEIGHT AND OBESITY [J].
KEYS, A ;
KARVONEN, MJ ;
KIMURA, N ;
FIDANZA, F ;
TAYLOR, HL .
JOURNAL OF CHRONIC DISEASES, 1972, 25 (6-7) :329-&
[8]   A model of an electro-goniometer and its calibration for biomechanical applications [J].
Legnani, G ;
Zappa, B ;
Casolo, F ;
Adamini, R ;
Magnani, PL .
MEDICAL ENGINEERING & PHYSICS, 2000, 22 (10) :711-722
[9]   Classification of tetraplegics through automatic movement evaluation [J].
Maksimovic, R ;
Popovic, M .
MEDICAL ENGINEERING & PHYSICS, 1999, 21 (05) :313-327
[10]  
NEUMAN MR, 1998, BIOPOTENTIAL AMPLIFI, V3