Noninvasive Hemoglobin Monitoring A Rapid, Reliable, and Cost-Effective Method Following Total Joint Replacement

被引:8
|
作者
Martin, J. Ryan [1 ,2 ]
Camp, Christopher L. [1 ,2 ]
Stitz, Amber [1 ,2 ,3 ]
Young, Ernest Y. [1 ,2 ]
Abdel, Matthew P. [1 ,2 ]
Taunton, Michael J. [1 ,2 ]
Trousdale, Robert T. [1 ,2 ]
机构
[1] Mayo Clin & Mayo Fdn, 200 1st St SW, Rochester, MN 55905 USA
[2] Mayo Clin & Mayo Fdn, Dept Orthoped Surg, 200 1st St SW, Rochester, MN 55905 USA
[3] Mayo Clin & Mayo Fdn, Dept Nursing, 200 1st St SW, Rochester, MN 55905 USA
关键词
CONCORDANCE CORRELATION-COEFFICIENT; KNEE ARTHROPLASTY; ACCURACY; OXIMETRY; POINT;
D O I
10.2106/JBJS.O.00820
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Noninvasive hemoglobin (nHgb) monitoring was initially introduced in the intensive care setting as a means of rapidly assessing Hgb values without performing a blood draw. We conducted a prospective analysis to compare reliability, cost, and patient preference between nHgb monitoring and invasive Hgb (iHgb) monitoring performed via a traditional blood draw. Methods: We enrolled 100 consecutive patients undergoing primary or revision total hip or total knee arthroplasty. On postoperative day 1, nHgb and iHgb values were obtained within thirty minutes of one another. iHgb and nHgb values, cost, patient satisfaction, and the duration of time required to obtain each reading were recorded. The concordance correlation coefficient (CCC) was utilized to evaluate the agreement of the two Hgb measurement methods. Paired t tests and Wilcoxon signed-rank tests were utilized to compare mean Hgb values, time, and pain for all readings. Results: The mean Hgb values did not differ significantly between the two measurement methods: the mean iHgb value (and standard deviation) was 11.3 +/- 1.4 g/dL (range, 8.2 to 14.3 g/dL), and the mean nHgb value was 11.5 +/- 1.8 g/dL (range, 7.0 to 16.0 g/dL) (p = 0.11). The CCC between the two Hgb methods was 0.69. One hundred percent of the patients with an nHgb value of >= 10.5 g/dL had an iHgb value of >8.0 g/dL. The mean time to obtain an Hgb value was 0.9 minute for the nHgb method and 51.1 minutes for the iHgb method (p < 0.001). At our institution, the cost of iHgb monitoring is approximately $28 per blood draw compared with $2 for each nHgb measurement, resulting in a savings of $26 per Hgb assessment when the noninvasive method is used. Conclusions: Noninvasive Hgb monitoring was found to be more efficient, less expensive, and preferred by patients compared with iHgb monitoring. Providers could consider screening total joint arthroplasty patients with nHgb monitoring and only order iHgb measurement if the nHgb value is <10.5 g/dL. If this protocol had been applied to the first blood draw in our 100 patients, approximately $2000 would have been saved. Extrapolated to the U.S. total joint arthroplasty practice, approximately $20 million could be saved annually.
引用
收藏
页码:349 / 355
页数:7
相关论文
共 4 条
  • [1] Prospective Noninvasive Hemoglobin Monitoring in the Outpatient Total Joint Arthroplasty Setting
    Dugdale, Evan M.
    Mallinger, Benjamin D.
    Bedard, Nicholas A.
    Couch, Cory G.
    Mabry, Tad M.
    Perry, Kevin I.
    Sierra, Rafael J.
    Taunton, Michael J.
    Trousdale, Robert T.
    Abdel, Matthew P.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2025, 107 (04) : 364 - 371
  • [2] Cost-effective Framework for Rapid Underwater Mapping with Digital Camera and Color Correction Method
    Chang, Anjin
    Jung, Jinha
    Um, Dugan
    Yeom, Junho
    Hanselmann, Frederick
    KSCE JOURNAL OF CIVIL ENGINEERING, 2019, 23 (04) : 1776 - 1785
  • [3] Monitoring Marine Habitats With Photogrammetry: A Cost-Effective, Accurate, Precise and High-Resolution Reconstruction Method
    Marre, Guilhem
    Holon, Florian
    Luque, Sandra
    Boissery, Pierre
    Deter, Julie
    FRONTIERS IN MARINE SCIENCE, 2019, 6
  • [4] A New, Rapid, Cost-Effective, Easy and Validated RP-HPLC Method for Determination of Antiviral (Sofosbuvir) in Bulk Forms
    Mehmood, Yasir
    Khan, Ikram Ullah
    Shahzad, Yasser
    Khalid, Syed Haroon
    Asghar, Sajid
    Yousaf, Abid Mehmood
    Hussain, Talib
    Irfan, Muhammad
    Khalid, Ikrima
    PAKISTAN JOURNAL OF ANALYTICAL & ENVIRONMENTAL CHEMISTRY, 2019, 20 (01) : 11 - 16