Fluorometric determination of ethanol in liquor samples by flow-injection analysis using an immobilized enzyme-reactor column with packing prepared by coupling alcohol oxidase and peroxidase onto chitosan beads

被引:0
作者
Taniai, T [1 ]
Sukuragawa, A [1 ]
Okutani, T [1 ]
机构
[1] Nihon Univ, Coll Sci & Technol, Dept Mat & Appl Chem, Chiyoda Ku, Tokyo 1018308, Japan
关键词
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-injection system was developed for the determination of ethanol with an immobilized enzyme-reactor column. This system, which consisted of hand-made reactor columns packed with alcohol oxidase and horseradish peroxidase immobilized onto chitosan beads, and a fluorometric detector, was applied to the determination of ethanol in liquor samples. Under the recommended conditions, the ethanol, which was present in the pretreated samples, was converted to hydrogen peroxide when it was passed through the immobilized alcohol oxidase (AOD) column with 0.1 mol/dm(3) phosphate buffer (pH 7.0). A sample can be analyzed with this system in < 10 min. The calibration curve for ethanol was linear from 2.0 to 0.1 mg/dm(3). The determination limit, which was defined by the difference between the sample peak and blank peak, was estimated to be 50 mug/dm(3) for ethanol. Interferences from some substances present in actual liquor samples decreased the analytical response and activity of the immobilized AOD-reactor column, but they were removed by dilution and pretreatment with an octyidecylsilane cartridge.
引用
收藏
页码:1475 / 1483
页数:9
相关论文
共 27 条
[1]   2-(4-carboxyphenyl)-6-N,N-diethylaminobenzofuran:: a useful reagent for the sensitive determination of alcohols by high-performance liquid chromatography with fluorimetric detection [J].
Assaf, P ;
Katzhendler, J ;
Haj-Yehia, AI .
JOURNAL OF CHROMATOGRAPHY A, 2000, 869 (1-2) :243-250
[2]  
BRANDENBERGER H, 1974, CLIN BIOCH PRINCIPLE, V2, P1435
[3]   Enzymatic determination of ethanol based on the intrinsic fluorescence of alcohol dehydrogenase [J].
deMarcos, S ;
Galban, J ;
Albajez, R ;
Castillo, JR .
ANALYTICA CHIMICA ACTA, 1997, 343 (1-2) :117-123
[4]   Enzymic flow injection determination of free L-carnitine in infant formulae [J].
Ferreira, IMPLVO ;
Macedo, MN ;
Ferreira, MA .
ANALYST, 1997, 122 (12) :1539-1541
[5]  
Glazkov VV, 1997, J ANAL CHEM+, V52, P71
[6]   SELECTIVE DETECTION IN FLOW-ANALYSIS BASED ON THE COMBINATION OF IMMOBILIZED ENZYMES AND CHEMICALLY MODIFIED ELECTRODES [J].
GORTON, L ;
CSOREGI, E ;
DOMINGUEZ, E ;
EMNEUS, J ;
JONSSONPETTERSSON, G ;
MARKOVARGA, G ;
PERSSON, B .
ANALYTICA CHIMICA ACTA, 1991, 250 (01) :203-248
[7]   MICROENZYMATIC FLUORESCENCE ASSAY FOR SERUM-CHOLESTEROL [J].
GRAY, MC ;
PLANT, AL ;
NICHOLSON, JM ;
MAY, WE .
ANALYTICAL BIOCHEMISTRY, 1995, 224 (01) :286-292
[8]   DEVELOPMENT OF ENZYME CARTRIDGE FLOW-INJECTION ANALYSIS FOR INDUSTRIAL-PROCESS MONITORING .2. APPLICATION FOR MONITORING OF MICROORGANISM CULTIVATIONS [J].
JURGENS, H ;
BRANDES, L ;
JOPPIEN, R ;
SIEBOLD, M ;
SCHUBERT, J ;
WU, X ;
KRETZMER, G ;
SCHUGERL, K .
ANALYTICA CHIMICA ACTA, 1995, 302 (2-3) :289-296
[9]  
Keesey J., 1987, BIOCH INFORM, P58
[10]   Highly sensitive flow-injection determination of glucose in plasma using an immobilized pyranose oxidase and a chemiluminometric peroxidase sensor [J].
Kiba, N ;
Itagaki, A ;
Fukumura, S ;
Saegusa, K ;
Furusawa, M .
ANALYTICA CHIMICA ACTA, 1997, 354 (1-3) :205-210