Evaluation and characterization of parallel plate microchamber's functionalities in small beam dosimetry

被引:25
作者
Lee, HR
Pankuch, M
Chu, JC
Spokas, JJ
机构
[1] Rush Presbyterian St Lukes Med Ctr, Dept Phys Med, Chicago, IL 60612 USA
[2] Benedictine Univ, Dept Phys, Lisle, IL 60532 USA
关键词
small beam; dosimetry; parallel plate microchamber; stereotactic; ionization chamber;
D O I
10.1118/1.1514576
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A parallel plate microchamber (PPMC) has been designed to specifically address the problems of small beam dosimetry. The chamber's extremely small volume and tissue equivalency theoretically make it possible for the chamber to perform an ideal measurement for small field dosimetry. Results show the PPMC to be a simple and reproducible detector for the measurements of total scattering factors, percentage depth doses, and off-axis ratios. Even with its unique geometry, the PPMC requires a correction factor when measuring total scatter factors of fields smaller than 2.5 cm in diameter. Results obtained with the PPMC for fields greater than 2.5 cm diameter closely match those of alternative measurement modalities. The exceptionally small volume of the chamber increases the effect of radiation-induced cable currents. With careful experimental technique, this problem can be resolved. Monte Carlo simulations of a Sun Nuclear QED low build-up diode were done to show that no correction factor is needed for the diode in measuring total scatter factors of small fields. However, the scattering factors measured with the PPMC should be corrected for cone fields smaller than 2.5 cm in diameter. With the correction factor, the scattering factor obtained with the PPMC matches that with the QED diode within 0.7%. The percent depth dose data taken with the PPMC for a 40x40 cm(2) field closely matches that taken with the PTW chamber with the largest deviation being similar to1.2% at a depth of 30 cm. For a measurement of the off-axis ratio with stereotactic cones of diameter 1.25 and 4.0 cm, the data obtained with the PPMC have a good agreement (less than 0.5% difference) with the film measurement. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:2489 / 2496
页数:8
相关论文
共 26 条
[1]  
[Anonymous], 1983, MED PHYS, V10, P741
[2]   ABSORBED DOSE PERTURBATION CAUSED BY DIODES FOR SMALL-FIELD PHOTON DOSIMETRY [J].
BEDDAR, AS ;
MASON, DJ ;
OBRIEN, PF .
MEDICAL PHYSICS, 1994, 21 (07) :1075-1079
[3]   Dosimetric verification of the 95% isodose surface for a conformal irradiation technique [J].
Brugmans, MJP ;
van der Horst, A ;
Lebesque, JV ;
Mijnheer, BJ .
MEDICAL PHYSICS, 1998, 25 (04) :424-434
[4]   The effect of detector size to the broadening of the penumbra - A computer simulated study [J].
Chang, KS ;
Yin, FF ;
Nie, KW .
MEDICAL PHYSICS, 1996, 23 (08) :1407-1411
[5]   The use of deconvolution and total least squares in recovering a radiation detector line spread function [J].
Charland, P ;
El-Khatib, E ;
Wolters, J .
MEDICAL PHYSICS, 1998, 25 (02) :152-160
[6]   ANALYSIS OF PHYSICAL PARAMETERS ASSOCIATED WITH THE MEASUREMENT OF HIGH-ENERGY X-RAY PENUMBRA [J].
DAWSON, DJ ;
HARPER, JM ;
AKINRADEWO, AC .
MEDICAL PHYSICS, 1984, 11 (04) :491-497
[7]   PENUMBRAL MEASUREMENTS IN WATER FOR HIGH-ENERGY X-RAYS [J].
DAWSON, DJ ;
SCHROEDER, NJ ;
HOYA, JD .
MEDICAL PHYSICS, 1986, 13 (01) :101-104
[8]   Use of a new type of radiochromic film, a new parallel-plate micro-chamber, MOSFETs, and TLD 800 microcubes in the dosimetry of small beams [J].
Francescon, P ;
Cora, S ;
Cavedon, C ;
Scalchi, P ;
Reccanello, S ;
Colombo, F .
MEDICAL PHYSICS, 1998, 25 (04) :503-511
[9]   Experimental determination of the convolution kernel for the study of the spatial response of a detector [J].
Garcia-Vicente, F ;
Delgado, JM ;
Peraza, C .
MEDICAL PHYSICS, 1998, 25 (02) :202-207
[10]   A comparison of dosimetry techniques in stereotactic radiosurgery [J].
Heydarian, M ;
Hoban, PW ;
Beddoe, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :93-110