Estimating local noise power spectrum from a few FBP-reconstructed CT scans

被引:17
作者
Zeng, Rongping [1 ]
Gavrielides, Marios A. [1 ]
Petrick, Nicholas [1 ]
Sahiner, Berkman [1 ]
Li, Qin [1 ]
Myers, Kyle J. [1 ]
机构
[1] US FDA, Div Imaging Diagnost & Software Reliabil, Off Sci & Engn Labs, CDRH, Silver Spring, MD 20993 USA
关键词
local noise power spectrum; CT; FBP; detectability; RAY COMPUTED-TOMOGRAPHY; MODERN DIAGNOSTIC MDCT; NODULE SIZE ESTIMATION; DETECTABILITY; COVARIANCE; IMAGES;
D O I
10.1118/1.4939061
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Traditional ways to estimate 2D CT noise power spectrum (NPS) involve an ensemble average of the power spectrums of many noisy scans. When only a few scans are available, regions of interest are often extracted from different locations to obtain sufficient samples to estimate the NPS. Using image samples from different locations ignores the nonstationarity of CT noise and thus cannot accurately characterize its local properties. The purpose of this work is to develop a method to estimate local NPS using only a few fan-beam CT scans. Methods: As a result of FBP reconstruction, the CT NPS has the same radial profile shape for all projection angles, with the magnitude varying with the noise level in the raw data measurement. This allows a 2D CT NPS to be factored into products of a 1D angular and a 1D radial function in polar coordinates. The polar separability of CT NPS greatly reduces the data requirement for estimating the NPS. The authors use this property and derive a radial NPS estimation method: in brief, the radial profile shape is estimated from a traditional NPS based on image samples extracted at multiple locations. The amplitudes are estimated by fitting the traditional local NPS to the estimated radial profile shape. The estimated radial profile shape and amplitudes are then combined to form a final estimate of the local NPS. We evaluate the accuracy of the radial NPS method and compared it to traditional NPS methods in terms of normalized mean squared error (NMSE) and signal detectability index. Results: For both simulated and real CT data sets, the local NPS estimated with no more than six scans using the radial NPS method was very close to the reference NPS, according to the metrics of NMSE and detectability index. Even with only two scans, the radial NPS method was able to achieve a fairly good accuracy. Compared to those estimated using traditional NPS methods, the accuracy improvement was substantial when a few scans were available. Conclusions: The radial NPS method was shown to be accurate and efficient in estimating the local NPS of FBP-reconstructed 2D CT images. It presents strong advantages over traditional NPS methods when the number of scans is limited and can be extended to estimate the in-plane NPS of cone-beam CT and multislice helical CT scans.
引用
收藏
页码:568 / 582
页数:15
相关论文
共 23 条
[1]  
[Anonymous], 1999, MODERN SPECTRAL ESTI
[2]  
[Anonymous], 2004, Foundations of Image Science
[3]   The noise power spectrum in CT with direct fan beam reconstruction [J].
Baek, Jongduk ;
Pelc, Norbert J. .
MEDICAL PHYSICS, 2010, 37 (05) :2074-2081
[4]   Application of the noise power spectrum in modern diagnostic MDCT: Part II. Noise power spectra and signal to noise [J].
Boedeker, K. L. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4047-4061
[5]   Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta [J].
Boedeker, K. L. ;
Cooper, V. N. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4027-4046
[6]   Visual signal detection with two-component noise: low-pass spectrum effects [J].
Burgess, AE .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1999, 16 (03) :694-704
[7]   Prewhitening revisited [J].
Burgess, AE .
IMAGE PERCEPTION: MEDICAL IMAGING 1998, 1998, 3340 :55-64
[8]   ANALYSIS OF X-RAY COMPUTED-TOMOGRAPHY IMAGES USING THE NOISE POWER SPECTRUM AND AUTO-CORRELATION FUNCTION [J].
FAULKNER, K ;
MOORES, BM .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (11) :1343-1352
[9]   Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation [J].
Gang, Grace J. ;
Stayman, J. Webster ;
Zbijewski, Wojciech ;
Siewerdsen, Jeffrey H. .
MEDICAL PHYSICS, 2014, 41 (08) :260-278
[10]   Analysis of Fourier-domain task-based detectability index in tomosynthesis and cone-beam CT in relation to human observer performance [J].
Gang, Grace J. ;
Lee, Junghoon ;
Stayman, J. Webster ;
Tward, Daniel J. ;
Zbijewski, W. ;
Prince, Jerry L. ;
Siewerdsen, Jeffrey H. .
MEDICAL PHYSICS, 2011, 38 (04) :1754-1768