Electrical Properties of Rat Skin and Design of Implantable Antennas for Medical Wireless Telemetry

被引:77
作者
Karacolak, Tutku [1 ]
Cooper, Robert [2 ]
Topsakal, Erdem [1 ]
机构
[1] Mississippi State Univ, Dept Elect & Comp Engn, Mississippi State, MS 39762 USA
[2] Mississippi State Univ, Coll Vet Med, Mississippi State, MS 39762 USA
关键词
Implantable antenna; medical implant communications service (MICS) band; scientific and medical (ISM) band; skin-mimicking material; DIELECTRIC-PROPERTIES; BIOLOGICAL TISSUES; MICROWAVE-FREQUENCIES; MICROSTRIP ANTENNA; MODELS; RANGE; BAND;
D O I
10.1109/TAP.2009.2027197
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The design and test is described of a small size dual band implantable antenna operating in Medical Implant Communications Service (MICS) (402 MHz-405 MHz) and Industrial, Scientific and Medical (ISM) (2.4 GHz-2.48 GHz) bands to be used in animal studies for medical research. The antenna is intended for wireless medical monitoring of the physiological parameters such as glucose, pressure, temperature, etc. First, the electrical properties (epsilon(r) and sigma) of skin samples from donor rats are measured at Mississippi State University's (MSU) College of Veterinary Medicine. A dual band antenna is then designed using an in-house finite element boundary integral solver in conjunction with particle swarm optimization algorithm. Finally, the antenna is tested using both skin-mimicking materials and real skin samples. The development details of the skin-mimicking materials are also given. Results regarding the S-11 and gain of the designed antenna are given and discussed in detail.
引用
收藏
页码:2806 / 2812
页数:7
相关论文
共 25 条
[1]  
Ballen M., 2001, PROC 23 ANN BIOELECT, V14, P80
[2]   Subminiature implantable potentiostat and modified commercial telemetry device for remote glucose monitoring [J].
Beach, RD ;
Von Küster, F ;
Moussy, F .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (06) :1239-1245
[3]   Towards a miniature implantable in vivo telemetry monitoring system dynamically configurable as a potentiostat or galvanostat for two- and three-electrode biosensors [J].
Beach, RD ;
Conlan, RW ;
Godwin, MC ;
Moussy, F .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2005, 54 (01) :61-72
[4]  
CHOU CK, 1984, BIOELECTROMAGNETICS, V5, P435, DOI 10.1002/bem.2250050408
[5]  
Duck F.A., 1990, PHYS PROPERTIES TISS, P20
[6]  
FOSTER KR, 1989, CRIT REV BIOMED ENG, V17, P25
[7]   Dielectric properties of tissue-equivalent liquids and their effects on specific absorption rate [J].
Fukunaga, K ;
Watanabe, S ;
Yamanaka, Y .
IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2004, 46 (01) :126-129
[8]   The dielectric properties of biological tissues .1. Literature survey [J].
Gabriel, C ;
Gabriel, S ;
Corthout, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2231-2249
[9]   The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2251-2269
[10]   The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2271-2293