Characteristics of orthovoltage x-ray therapy beams at extended SSD for applicators with end plates

被引:12
作者
Li, XA
Salhani, D
Ma, CM
机构
[1] UNIV OTTAWA,DEPT RADIOL,OTTAWA,ON K1Y 4K7,CANADA
[2] NATL RES COUNCIL CANADA,INST NATL MEASUREMENT STAND,IONIZING RADIAT STAND,OTTAWA,ON K1A 0R6,CANADA
关键词
D O I
10.1088/0031-9155/42/2/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effects of extended SSD (ESSD) on orthovoltage x-ray therapy dosimetry have been studied for a Therapax DXT300 treatment unit with several open- and close-ended cones. For the open-ended cones, the beam output and percentage depth dose (PDD) at ESSD can be accurately estimated from the measured data at the nominal SSD (NSSD) by an inverse square law (ISL) and a divergence factor for gaps (distance from cone end to point of interest) up to 15 cm. In the case of close-ended applicators (with an end plate of 3.2 mm thick PMMA), extending SSD results in significant changes in output and in PDDs, but has little effect on beam penumbra and flatness for gaps up to 15 cm. The output at ESSD calculated by applying the ISL with the NSSD is found to be as much as 9% different from the measurement for the close-ended cone. The output at ESSD can, however, be calculated accurately by applying the ISL with an effective SSD or using a measured gap factor. The change in the PDDs with SSD can be calculated from a geometric divergence using a pseudo SSD which is approximately equal to 60% of the effective SSD determined in air. The calculated PDD curves at ESSD using this divergence factor agree well with measurements (within +/-1% for beams of 100-300 kVp with cone size ranging from 10x10 to 20x20 cm(2) and gap distance up to 15 cm).
引用
收藏
页码:357 / 370
页数:14
相关论文
共 15 条
[1]  
*AAPM, 1991, MED PHYS, V18, P73
[2]  
Aldrich J E, 1992, Med Dosim, V17, P69
[3]  
BURNS JE, 1983, BR J RADIOL S, V17
[4]   BEAM CHARACTERISTICS OF THE THERAPAX DXT300 ORTHOVOLTAGE THERAPY UNIT [J].
GERIG, L ;
SOUBRA, M ;
SALHANI, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (09) :1377-1392
[5]   DEPENDENCE OF THE PHOTON BACKSCATTER FACTOR FOR WATER ON SOURCE-TO-PHANTOM DISTANCE AND IRRADIATION FIELD SIZE [J].
GROSSWENDT, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1233-1245
[6]  
GROSSWENDT B, 1993, PHYS MED BIOL, V38, P306
[7]  
*IAEA, 1987, 277 IAEA
[8]  
ICRU, 1973, 23 ICRU
[9]  
IPSM, 1991, PHYS MED BIOL, V36, P1027
[10]  
JOHNS HE, 1983, PHYSICS RADIOLOGY