Comparative Analysis of Different Methods of Modeling the Thermal Effect of Circulating Blood Flow During RF Cardiac Ablation

被引:76
作者
Gonzalez-Suarez, A. [1 ,2 ]
Berjano, E. [1 ]
机构
[1] Univ Politecn Valencia, Dept Elect Engn, Biomed Synergy, E-46022 Valencia, Spain
[2] Basque Ctr Appl Math, Bilbao 48009, Spain
关键词
Blood flow; cardiac ablation; catheter ablation; computer modeling; radio frequency (RF) ablation; thermal modeling; RADIOFREQUENCY-CATHETER ABLATION; 3-DIMENSIONAL FINITE-ELEMENT; LESION SIZE; TEMPERATURE DISTRIBUTIONS; EXPERIMENTAL VALIDATION; MYOCARDIAL TISSUE; LETHAL ISOTHERM; CURRENT-DENSITY; ELECTRODE; ENDOCARDIUM;
D O I
10.1109/TBME.2015.2451178
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Our aim was to compare the different methods of modeling the effect of circulating blood flow on the thermal lesion dimensions created by radio frequency (RF) cardiac ablation and on the maximum blood temperature. Computational models were built to study the temperature distributions and lesion dimensions created by a nonirrigated electrode by two RF energy delivery protocols (constant voltage and constant temperature) under high and low blood flow conditions. Four methods of modeling the effect of circulating blood flowon lesion dimensions and temperature distribution were compared. Three of them considered convective coefficients at the electrode-blood and tissue-blood interfaces to model blood flow: 1) without including blood as a part of the domain; 2) constant electrical conductivity of blood; and 3) temperature-dependent electrical conductivity of blood (+2%/degrees C). Method 4) included blood motion andwas considered to be a reference method for comparison purposes. Only Method 4 provided a realistic blood temperature distribution. The other three methods predicted lesion depth values similar to those of the reference method (differences smaller than 1 mm), regardless of ablation mode and blood flow conditions. Considering the aspects of lesion size and maximum temperature reached in blood and tissue, Method 2 seems to be the most suitable alternative to Method 4 in order to reduce the computational complexity. Our findings could have an important implication in future studies of RF cardiac ablation, in particular, in choosing the most suitable method to model the thermal effect of circulating blood.
引用
收藏
页码:250 / 259
页数:10
相关论文
共 39 条
[1]   A thermal-ablation bioheat model including liquid-to-vapor phase change, pressure- and necrosis-dependent perfusion, and moisture-dependent properties [J].
Abraham, J. P. ;
Sparrow, E. M. .
INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2007, 50 (13-14) :2537-2544
[2]   Could it be advantageous to tune the temperature controller during radiofrequency ablation? A feasibility study using theoretical models [J].
Alba-Martinez, Jose ;
Trujillo, Macarena ;
Blasco-Gimenez, Ramon ;
Berjano, Enrique .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2011, 27 (06) :539-548
[3]   Ablation time efficiency and lesion volume - in vitro comparison of 4 mm, non irrigated, gold- and platinum-iridium-tip radiofrequency ablation catheters [J].
Balazs, Tibor ;
Laczko, Romola ;
Bognar, Eszter ;
Akman, Serkan ;
Nagy, Peter ;
Zima, Endre ;
Dobranszky, Janos ;
Szili-Torok, Tamas .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2013, 36 (01) :13-18
[4]   Theoretical modeling for radiofrequency ablation: state-of-the-art and challenges for the future [J].
Berjano, Enrique J. .
BIOMEDICAL ENGINEERING ONLINE, 2006, 5 (1)
[5]   Flow effect on lesion formation in RF cardiac catheter ablation [J].
Cao, H ;
Vorperian, VR ;
Tungjitkusolmun, S ;
Tsai, JZ ;
Haemmerich, D ;
Bin Choy, Y ;
Webster, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (04) :425-433
[6]   CALCULATION OF ELECTRIC-FIELDS IN CONDUCTIVE MEDIA [J].
DOSS, JD .
MEDICAL PHYSICS, 1982, 9 (04) :566-573
[7]  
Gallagher N. P., 2012, THESIS CALGARY U CAL
[8]   Contact Geometry Affects Lesion Formation in Radio-Frequency Cardiac Catheter Ablation [J].
Gallagher, Neal ;
Fear, Elise C. ;
Byrd, Israel A. ;
Vigmond, Edward J. .
PLOS ONE, 2013, 8 (09)
[9]   Radiofrequency cardiac ablation with catheters placed on opposing sides of the ventricular wall: Computer modelling comparing bipolar and unipolar modes [J].
Gonzalez-Suarez, Ana ;
Trujillo, Macarena ;
Koruth, Jacob ;
d'Avila, Andre ;
Berjano, Enrique .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2014, 30 (06) :372-384
[10]   Hepatic radiofrequency ablation with internally cooled probes: Effect of coolant temperature on lesion size [J].
Haemmerich, D ;
Chachati, L ;
Wright, AS ;
Mahvi, DM ;
Lee, FT ;
Webster, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2003, 50 (04) :493-500