Performance of arthroscopic irrigation systems assessed with automatic blood detection

被引:15
作者
Tuijthof, G. J. M. [1 ,2 ]
de Vaal, M. M. [1 ]
Sierevelt, I. N. [1 ]
Blankevoort, L. [1 ]
van der List, M. P. J. [3 ,4 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Orthoped Surg, Orthoped Res Ctr Amsterdam, NL-1105 AZ Amsterdam, Netherlands
[2] Delft Univ Technol, Fac 3ME, Dept BioMech Engn, NL-2628 CD Delft, Netherlands
[3] Bergman Med Care, NL-3723 MB Bilthoven, Netherlands
[4] Former Meander Med Ctr, Baarn, Netherlands
关键词
Arthroscopy; Shoulder; Performance; Bleeding episode; View; Irrigation system; PUMP SYSTEMS; SKILLS; PRESSURE; SURGERY;
D O I
10.1007/s00167-011-1495-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose During arthroscopies, bleeding episodes occur as a result of tissue damage. Irrigation systems assist in minimizing these disturbances. The performance of three arthroscopic irrigation systems in clearing bleeding episodes was evaluated objectively. Methods One surgeon performed 99 shoulder arthroscopies using three irrigation systems: gravity infusion, single roller, and double roller pump. The three irrigation systems groups were matched postoperatively for type of surgery-acromioplasty, SLAP, release, rotator cuff repair and capsule repair, and operation duration. The recorded arthroscopies were analyzed for the presence of bleeding episodes with a special computer program that automatically detects the tinctures of red-colored blood. A least 20% of an arthroscopic image had to be covered with blood to qualify as bleeding episode. Results The median (min-max) presence of bleeding episodes as a percentage of the operation time was 6.6% (0.0-43.6%) for gravity infusion, 3.7% (0.2-46.4%) for the single roller, and 3.3% (0.0-19.3%) for the double roller pump, respectively. The large variation could be attributed to the occurrence of arterial bleeding episodes during some procedures. No significant differences were found between the irrigation systems. For a subgroup including acromioplasties and releases, significant differences were found in favor of both roller pumps (P < 0.05). Conclusions Overall, the roller pumps did not outperform gravity infusion. However, from the results, high-risk procedures for bleeding episodes were identified (acromioplasty and release of a frozen shoulder) that can benefit from the use of roller pumps. A clear view is essential to perform an arthroscopic procedure safely and efficiently. Level of evidence III.
引用
收藏
页码:1948 / 1954
页数:7
相关论文
共 18 条
  • [1] Ampat G, 1997, ANN ROY COLL SURG, V79, P341
  • [2] Haptic feedback can provide an objective assessment of arthroscopic skills
    Chami, George
    Ward, James W.
    Phillips, Roger
    Sherman, Kevin P.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (04) : 963 - 968
  • [3] Di Gennaro Giovanni Luigi, 2008, Orthopedics, V31, P809
  • [4] DOLK T, 1989, Arthroscopy, V5, P307, DOI 10.1016/0749-8063(89)90147-3
  • [5] Transferring simulated arthroscopic skills to the operating theatre - A randomised blinded study
    Howells, N. R.
    Gill, H. S.
    Carr, A. J.
    Price, A. J.
    Rees, J. L.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (04): : 494 - 499
  • [6] Motion analysis: A validated method for showing skill levels in arthroscopy
    Howells, Nick R.
    Brinsden, Mark D.
    Gill, Richie S.
    Carr, Andrew J.
    Rees, Jonathan L.
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2008, 24 (03) : 335 - 342
  • [7] The Development of an Objective Model to Assess Arthroscopic Performance
    Insel, Aaron
    Carofino, Bradley
    Leger, Robin
    Arciero, Robert
    Mazzocca, Augustus D.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (09) : 2287 - 2295
  • [8] Morgan C D, 1987, Arthroscopy, V3, P288, DOI 10.1016/S0749-8063(87)80126-3
  • [9] A controlled trial of arthroscopic surgery for osteoarthritis of the knee
    Moseley, JB
    O'Malley, K
    Petersen, NJ
    Menke, TJ
    Brody, BA
    Kuykendall, DH
    Hollingsworth, JC
    Ashton, CM
    Wray, NP
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2002, 347 (02) : 81 - 88
  • [10] Accuracy of pressure and flow capacities of four arthroscopic fluid management systems
    Muellner, T
    Menth-Chiari, WA
    Reihsner, R
    Eberhardsteiner, J
    Engebretsen, L
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2001, 17 (07) : 760 - 764