Thermoluminescence measurements of neutron dose around a medical linac

被引:37
作者
Barquero, R [1 ]
Méndez, R
Iñiguez, MP
Vega, HR
Voytchev, M
机构
[1] Hosp Univ Rio Hortega, Dept Protecc Radiol, Valladolid, Spain
[2] Univ Valladolid, Dept Fis Teor & Atom Mol & Nucl, Valladolid, Spain
[3] Ctr Reg Estudios Nucl, Zacatecas, Mexico
关键词
D O I
10.1093/oxfordjournals.rpd.a006035
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The photoneutron ambient dose around a 18 MV medical electron lineal accelerator has been measured with LiF:Mg,Ti chips of 3 X 3 X I nun inside moderating spheres. During the measurements a water phantom was irradiated in a field of 40 X 40 cm(2). Two methods have been considered for comparison. In the first, a TLD-600/TLD-700 pair at the centre of a 25 cm diameter paraffine sphere was used, with the system behaving as a rein meter. In the second method, TLD-600/TLD-700 pairs, bare and at the centre of 7.6, 12.7, 20.3 25.4, and 30.5 cm diameter polyethylene Bonner spheres Were used to obtain the neutron spectrum. This was unfolded using the BUNKIUT code with the SPUNIT algorithm and the UTA4 and ARKI response functions. The neutron dose was followed by Multiplying the unfolded neutron spectrum by the ambient dose equivalent to neutron fluence conversion factors. Both methods result in 0.5 mSv. Gy(-1) 1 m away from the isocentre.
引用
收藏
页码:493 / 496
页数:4
相关论文
共 17 条
[1]  
AAPM, 1986, 19 AAPM
[2]   A NEW TYPE OF NEUTRON SPECTROMETER [J].
BRAMBLETT, RL ;
EWING, RI ;
BONNER, TW .
NUCLEAR INSTRUMENTS & METHODS, 1960, 9 (01) :1-12
[3]   In vivo neutron dosimetry during high-energy bremsstrahlung radiotherapy [J].
D'Errico, F ;
Nath, R ;
Silvano, G ;
Tana, L .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (05) :1185-1192
[4]   NEW METHODS FOR MEASURING NEUTRON-SPECTRA WITH ENERGY FROM 0.4EV TO 10 MEV BY TRACK AND ACTIVATION DETECTORS [J].
DOROSHENKO, JJ ;
KRAITOR, SN ;
KUZNETSOVA, TV ;
KUSHNEREVA, KK ;
LEONOV, ES .
NUCLEAR TECHNOLOGY, 1977, 33 (03) :296-304
[5]  
GOLDHAGEN P, 2001, IN PRESS NUCL I ME A
[6]  
GRIFFITH RV, 1990, TECH REPORT SERIES, V318
[7]   THE RESPONSE OF BONNER SPHERES TO NEUTRONS FROM THERMAL ENERGIES TO 17.3 MEV [J].
HERTEL, NE ;
DAVIDSON, JW .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 238 (2-3) :509-516
[8]  
HOLEMAN G R, 1977, Medical Physics (Woodbury), V4, P508, DOI 10.1118/1.594339
[9]  
Lemley E. C., 1994, Transactions of the American Nuclear Society, V70, P335
[10]   MODIFICATIONS TO RECURSION UNFOLDING ALGORITHMS TO FIND MORE APPROPRIATE NEUTRON-SPECTRA [J].
LOWRY, KA ;
JOHNSON, TL .
HEALTH PHYSICS, 1984, 47 (04) :587-593