Survival rates as a guide to implanted cardioverter-defibrillator replacement strategies for device recalls - Adding statistical insight to clinical intuition

被引:12
作者
Gula, Lorne J. [1 ]
Massel, David [1 ]
Krahn, Andrew D. [1 ]
Skanes, Allan C. [1 ]
Yee, Raymond [1 ]
Klein, George J. [1 ]
机构
[1] Univ Western Ontario, Dept Med, Div Cardiol, London, ON N6A 5A5, Canada
关键词
D O I
10.1016/j.ahj.2006.10.029
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Recalls and advisories of implanted cardioverter-defibrillators (ICDs) have become an unfortunate reality of cardiac rhythm management. With a paucity of data available on which to base replacement decisions, our goal is to model the potential risks and benefits of ICD generator replacement. The estimated risks are varied through a wide range to determine the potential range of outcomes. Methods and Results Using initial estimates of risk derived from real data on 2915 advisory devices from 17 implanting centers, a decision analysis and Markov model were used to estimate survival according to device replacement decision. Survival rates at 5 years with and without device replacement were estimated at 60.38% and 60.66%, respectively. This difference was not significantly different on comparative analysis, using variability determined by Monte Carlo simulation. One-way and two-way sensitivity analyses are presented, demonstrating the minimal effect of varying estimates of risk. Only variation in risk of device failure had a differential effect on survival, with a survival benefit at 7 years if annual risk of device failure is at least 1.8%. Little differential effect on survival was demonstrated by variation of estimates of arrhythmia risk, nonarrhythmic mortality, and postprocedure infection rate. Conclusions Survival rates with a generator replacement or nonreplacement strategy in response to ICD recalls are similar and decrease nearly in parallel over time. The main factor with differential effect on survival is risk of device failure, although the level of this risk required to confer a survival advantage to a replacement strategy is quite large.
引用
收藏
页码:253 / 259
页数:7
相关论文
共 17 条
[1]   Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure [J].
Bardy, GH ;
Lee, KL ;
Mark, DB ;
Poole, JE ;
Packer, DL ;
Boineau, R ;
Domanski, M ;
Troutman, C ;
Anderson, J ;
Johnson, G ;
McNulty, SE ;
Clapp-Channing, N ;
Davidson-Ray, LD ;
Fraulo, ES ;
Fishbein, DP ;
Luceri, RM ;
Ip, JH .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (03) :225-237
[2]   Clinical study of the laser sheath for lead extraction: The total experience in the United States [J].
Byrd, CL ;
Wilkoff, BL ;
Love, CJ ;
Sellers, TD ;
Reiser, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (05) :804-808
[3]   Intravascular extraction of problematic or infected permanent pacemaker leads: 1994-1996 [J].
Byrd, CL ;
Wilkoff, BL ;
Love, CJ ;
Sellers, TD ;
Turk, KT ;
Reeves, R ;
Young, R ;
Crevey, B ;
Kutalek, SP ;
Freedman, R ;
Friedman, R ;
Trantham, J ;
Watts, M ;
Schutzman, J ;
Oren, J ;
Wilson, J ;
Gold, F ;
Fearnot, NE ;
Van Zandt, HJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (09) :1348-1357
[4]  
Curnis Antonio, 2005, Ital Heart J, V6, P216
[5]  
GORSETT A, 2005, COMMUNICATION 0617
[6]  
GORSETT A, 2005, COMMUNICATION 0624
[7]  
GORSETT A, 2005, COMMUNICATION 0523
[8]   Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories [J].
Gould, PA ;
Krahn, AD .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (16) :1907-1911
[9]   Lessons from the failure and recall of an implantable cardioverter-defibrillator [J].
Hauser, RG ;
Maron, BJ .
CIRCULATION, 2005, 112 (13) :2040-2042
[10]   Deaths associated with implantable cardioverter defibrillator failure and deactivation reported in The United States Food and Drug Administration Manufacturer and User Facility Device Experience Database [J].
Hauser, RG ;
Kallinen, L .
HEART RHYTHM, 2004, 1 (04) :399-405