In vivo oxide-induced stress corrosion cracking of Ti-6Al-4V in a neck-stem modular taper: Emergent behavior in a new mechanism of in vivo corrosion

被引:80
作者
Gilbert, Jeremy L. [1 ,2 ]
Mali, Sachin [1 ,2 ]
Urban, Robert M. [3 ]
Silverton, Craig D. [4 ]
Jacobs, Joshua J. [3 ]
机构
[1] Syracuse Univ, Syracuse Biomat Inst, Syracuse, NY 13244 USA
[2] Syracuse Univ, Dept Biomed & Chem Engn, Syracuse, NY USA
[3] Rush Univ, Med Ctr, Dept Orthoped Surg, Chicago, IL 60612 USA
[4] Henry Ford Med Grp, Detroit, MI USA
关键词
corrosion; modular; failure analysis; hip prosthesis; titanium (alloys); CREVICE CORROSION; TITANIUM;
D O I
10.1002/jbm.b.31943
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In vivo modular taper corrosion in orthopedic total joint replacements has been documented to occur for headneck tapers, modular-body tapers, and neckstem tapers. While the fretting corrosion mechanism by which this corrosion occurs has been described in the literature, this report shows new and as yet unreported mechanisms at play. A retrieved Ti-6Al-4V/Ti-6Al-4V neckstem taper interface, implanted for 6 years is subjected to failure analysis to document taper corrosion processes that lead to oxide driven crack formation on the medial side of the taper. Metallurgical sectioning techniques and scanning electron microscopy analysis are used to document the taper corrosion processes. The results show large penetrating pitting attack of both sides of the taper interface where corrosion selectively attacks the beta phase of the microstructure and eventually consumes the alpha phase. The pitting attack evolves into plunging pits that ultimately develop into cracks where the crack propagation process is one of corrosion resulting in oxide formation and subsequent reorganization. This process drives open the crack and advances the front by a combination of oxide-driven crack opening stresses and corrosion attack at the tip. The oxide that forms has a complex evolving structure including a network of transport channels that provide access of fluid to the crack tip. This emergent behavior does not appear to require continued fretting corrosion to propagate the pitting and cracking. This new mechanism is similar to stress corrosion cracking where the crack tip stresses arise from the oxide formation in the crack and not externally applied tensile stresses. (C) 2011 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 100B: 584594, 2012.
引用
收藏
页码:584 / 594
页数:11
相关论文
共 18 条
[1]  
[Anonymous], 1986, BASIC CORROSION OXID
[2]   FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS [J].
BROWN, SA ;
FLEMMING, CAC ;
KAWALEC, JS ;
PLACKO, HE ;
VASSAUX, C ;
MERRITT, K ;
PAYER, JH ;
KRAAY, MJ .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :19-26
[3]  
COLLIER JP, 1992, CLIN ORTHOP RELAT R, P129
[4]   The Effect of Static Applied Potential on the 24-Hour Impedance Behavior of Commercially Pure Titanium in Simulated Biological Conditions [J].
Ehrensberger, Mark T. ;
Gilbert, Jeremy L. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2010, 93B (01) :106-112
[5]  
Gilbert J., 2006, ASM INT HDB CORROSIO, P826
[6]  
Gilbert J L, 2011, COMPREHENSIVE BIOMAT
[7]   IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[8]   The mechanical and electrochemical processes associated with taper fretting crevice corrosion: A review [J].
Gilbert, JL ;
Jacobs, JJ .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :45-59
[9]   INTERGRANULAR CORROSION-FATIGUE FAILURE OF COBALT-ALLOY FEMORAL STEMS - A FAILURE ANALYSIS OF 2 IMPLANTS [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ ;
BERTIN, KC ;
ZERNICH, MR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (01) :110-115
[10]  
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149