Use of the aerosizer® aerodynamic particle size analyzer to characterize aerosols from pressurized metered-dose inhalers (pMDIs) for medication delivery

被引:9
作者
Mitchell, JP
Nagel, MW
Cheng, YS
机构
[1] Trudell Med Int, London, ON N5Z 3M5, Canada
[2] Inhalat Toxicol Res Inst, Albuquerque, NM 87185 USA
关键词
D O I
10.1016/S0021-8502(98)00072-X
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The Aerosizer(R) aerodynamic particle size analyzer is widely used for the rapid assessment of medical aerosols. Its real-time capability to make measurements offers a significant advantage in terms of user convenience compared with aerodynamic particle size analysis by cascade impactor or multi-stage liquid impinger. However, the present study has highlighted a limitation in its use in the measurement of aerosols from many types of pressurized metered-dose inhaler (pMDI) in which the measured total number concentration exceeds about 10(10) particles m(-3). The causes are complex, but appear primarily to be associated with the loss of electrical signals from individual particle-detection beam interactions, associated with timer insensitivity of the signal correlation system. This phenomenon occurs where several particles are present simultaneously within the measurement zone. Subsequent processing of the time-of-flight data may result in significant distortion of the reported size distribution, including in extreme cases the absence of portions of the data where the AUTOCOMBINE feature is used to create the full size distribution from the low and high detection sensitivity sub-ranges. (C) 1999 Elsevier Science Ltd. All rights reserved
引用
收藏
页码:467 / 477
页数:11
相关论文
共 14 条
[1]   CHOOSING THE METERED-DOSE INHALER SPACER OR HOLDING CHAMBER THAT MATCHES THE PATIENTS NEED - EVIDENCE THAT THE SPECIFIC DRUG BEING DELIVERED IS AN IMPORTANT CONSIDERATION [J].
AHRENS, R ;
LUX, C ;
BAHL, T ;
HAN, SH .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1995, 96 (02) :288-294
[2]  
*API INC, 1990, AER US MAN VERS 5 01
[3]  
BARON PA, 1993, AEROSOL MEASUREMENT, P381
[4]  
BOHAN J, 1994, RESPIRATORY DRUG DEL, V4, P330
[5]   AEROSOL BEAM SPECTROMETRY [J].
DAHNEKE, B .
NATURE-PHYSICAL SCIENCE, 1973, 244 (134) :54-55
[6]   PROPERTIES OF CONTINUUM SOURCE PARTICLE BEAMS .1. CALCULATION METHODS AND RESULTS [J].
DAHNEKE, BE ;
CHENG, YS .
JOURNAL OF AEROSOL SCIENCE, 1979, 10 (03) :257-274
[7]  
Dalby RN, 1996, INHALATION AEROSOLS, P441
[8]  
DOLOVICH MB, 1995, EUR RESPIR J, V8, P201
[9]  
*EUR PHARM, 1995, PREP INH
[10]  
GEOFFROY, 1998, COMMUNICATION