Monosegmented flow system exploiting multicommutation applied to spectrophotometric determination of manganese in soybean digests

被引:19
|
作者
Smiderle, M [1 ]
Reis, BF [1 ]
Rocha, FRP [1 ]
机构
[1] Univ Sao Paulo, Ctr Energia Nucl Agr, BR-13400970 Piracicaba, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
flow analysis; multicommutation; binary sampling; monosegmented flow analysis; soybean; manganese determination;
D O I
10.1016/S0003-2670(99)00015-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow system combining binary sampling and monosegmented how implemented by multicommutation was developed to improve sensitivity and sampling rate of spectrophotometric methods based on slow chemical reactions. The flow network was assembled from a set of computer controlled three-way solenoid valves. The spectrophotometric method was based on manganese (II) oxidation by periodate in phosphoric acid medium to form the permanganate anion. The reaction is relatively slow and at 47 degrees C a sample residence time of ca. 5 min is necessary to attain equilibrium conditions. However, by exploiting binary sampling and monosegmented flow analysis, quantitative conversion with a sampling rate of 50 determinations per hour was achieved. The procedure was applied to Mn determination in soybean digests, showing as profitable features a linear response for sample concentrations ranging from 2.50 to 40.00 mg l(-1) Mn2+, a relative standard deviation of 0.27% (n=9) for a sample solution containing 17.1 mg l(-1) Mn2+, a low reagent consumption (30 mg NaIO4 per determination) and a detection limit of 1.2 mg l(-1) (99.7% confidence level). Results agreed with those obtained by inductively coupled argon plasma atomic emission spectrometry at 95% confidence level. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:129 / 135
页数:7
相关论文
共 50 条
  • [1] An improved flow system for spectrophotometric determination of anions exploiting multicommutation and multidetection
    Rocha, FRP
    Martelli, PB
    Reis, BF
    ANALYTICA CHIMICA ACTA, 2001, 438 (1-2) : 11 - 19
  • [2] Multicommutation in flow analysis exploiting a multizone trapping approach: spectrophotometric determination of boron in plants
    Tumang, CD
    de Luca, GC
    Fernandes, RN
    Reis, BF
    Krug, FJ
    ANALYTICA CHIMICA ACTA, 1998, 374 (01) : 53 - 59
  • [3] Multicommutation in flow analysis .5. Binary sampling for sequential spectrophotometric determination of ammonium and phosphate in plant digests
    Kronka, EAM
    Reis, BF
    Korn, M
    Bergamin, H
    ANALYTICA CHIMICA ACTA, 1996, 334 (03) : 287 - 293
  • [4] Automatic multicommutation flow system for wide range spectrophotometric calcium determination
    Rocha, FRP
    Martelli, PB
    Frizzarin, RM
    Reis, BF
    ANALYTICA CHIMICA ACTA, 1998, 366 (1-3) : 45 - 53
  • [5] MULTICOMMUTATION IN FLOW-ANALYSIS .1. BINARY SAMPLING - CONCEPTS, INSTRUMENTATION AND SPECTROPHOTOMETRIC DETERMINATION OF IRON IN PLANT DIGESTS
    REIS, BF
    GINE, MF
    ZAGATTO, EAG
    LIMA, JLFC
    LAPA, RA
    ANALYTICA CHIMICA ACTA, 1994, 293 (1-2) : 129 - 138
  • [6] Spectrophotometric Determination of Thiocyanate in Human Saliva Employing Micropumping Multicommutation Flow System
    Silva Junior, Joilton J.
    Farias, Marcelo A.
    Silva, Valdinete L.
    Montenegro, Maria C. B. S. M.
    Araujo, Alberto N.
    Lavorante, Andre F.
    Paim, Ana Paula S.
    SPECTROSCOPY LETTERS, 2010, 43 (03) : 213 - 219
  • [7] An improved flow system for phenols determination exploiting multicommutation and long pathlength spectrophotometry
    Lupetti, KO
    Rocha, FRP
    Fatibello, O
    TALANTA, 2004, 62 (03) : 463 - 467
  • [8] SPECTROPHOTOMETRIC DETERMINATION OF CREATININE BY MONOSEGMENTED CONTINUOUS-FLOW ANALYSIS
    DEFARIA, LC
    PASQUINI, C
    JOURNAL OF AUTOMATIC CHEMISTRY, 1992, 14 (03): : 97 - 100
  • [9] A monosegmented flow titration for the spectrophotometric determination of total acidity in vinegar
    Honorato, RS
    Araújo, MCU
    Veras, G
    Zagatto, EAG
    Lapa, RAS
    Lima, JLFC
    ANALYTICAL SCIENCES, 1999, 15 (07) : 665 - 668
  • [10] A Monosegmented Flow Titration for the Spectrophotometric Determination of Total Acidity in Vinegar
    Ricardo S. Honorato
    Mário C. U. Araújo
    Germano Veras
    Elias A. G. Zagatto
    Rui A. S. Lapa
    José L. F. C. Lima
    Analytical Sciences, 1999, 15 : 665 - 668