Correction of artefacts in optical projection tomography

被引:75
作者
Walls, JR
Sled, JG
Sharpe, J
Henkelman, RM
机构
[1] Univ Toronto, Hosp Sick Children, Mouse Imaging Ctr, Toronto, ON M5G 1X8, Canada
[2] Western Gen Hosp, MRC, Human Genet Unit, Edinburgh EH4 2XU, Midlothian, Scotland
关键词
D O I
10.1088/0031-9155/50/19/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new imaging technique called optical projection tomography (OPT), essentially an optical version of x-ray computed tomography (CT), provides molecular specificity, cellular resolution and larger specimen coverage (approximate to 1 cubic centimetre) than was previously possible with other imaging techniques. It is ideally suited to gene expression studies in small animals. Reconstructed OPT images demonstrate several artefacts which reduce the overall image quality. In this paper, we describe methods to prevent smear artefacts due to illumination intensity fluctuation, ring artefacts due to CCD pixel sensitivity variation and a new 'detector edge' artefact caused by non-zero background signal. We also present an automated method to align the position of the rotational axis during image reconstruction. Finally, we propose a method to eliminate bowl artefacts due to projection truncation using a lower resolution OPT scan of the same specimen. This solution also provides OPT with the ability to obtain a high-resolution reconstruction from a region of interest of a specimen that is larger than the field of view. Implementation of these corrections and modifications increases the accuracy of the OPT imaging technique and extends its capabilities to obtain higher resolution data from within a whole specimen.
引用
收藏
页码:4645 / 4665
页数:21
相关论文
共 35 条
[31]   Phenotyping transgenic embryos: a rapid 3-D screening method based on episcopic fluorescence image capturing [J].
Weninger, WJ ;
Mohun, T .
NATURE GENETICS, 2002, 30 (01) :59-65
[32]   ENHANCED DC ARC LAMP PERFORMANCE FOR SPECTROSCOPIC APPLICATIONS [J].
WOODLEE, RL ;
FUH, MRS ;
PATONAY, G ;
WARNER, IM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (12) :3640-3646
[33]   Molecular contrast optical coherence tomography: A review [J].
Yang, CH .
PHOTOCHEMISTRY AND PHOTOBIOLOGY, 2005, 81 (02) :215-237
[34]  
Zucker RM, 1998, CYTOMETRY, V33, P348, DOI 10.1002/(SICI)1097-0320(19981101)33:3<348::AID-CYTO9>3.0.CO
[35]  
2-C