Effective Use of Radiation Shields to Minimize Operator Dose During Invasive Cardiology Procedures

被引:96
作者
Fetterly, Kenneth A. [1 ]
Magnuson, Dixon J. [2 ]
Tannahill, Gordon M.
Hindal, Mark D. [2 ]
Mathew, Verghese [1 ]
机构
[1] Mayo Clin, Div Cardiovasc Dis, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
关键词
radiation protection; radiation safety; radiation shield; PERFORMING CARDIAC-CATHETERIZATION; SCATTER RADIATION; EXPOSURE; ANGIOGRAPHY; PROTECTION; DEVICE; IMPACT; STAFF; TUBE; LENS;
D O I
10.1016/j.jcin.2011.05.027
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives This study sought to measure the protection from scatter radiation offered to the primary physician by a variety of available shields and to provide best practice guidelines for shield use during invasive cardiology procedures. Background It is accepted that exposure to radiation includes a predicted increase in cancer risk. In the cardiac interventional laboratories, radiation shields are widely available; however, proper use of the shields to optimize protection during cardiac interventional procedures is not well understood. Methods The protection from scatter radiation offered by a variety of shields used alone and in combination was measured. Protection was assessed from air-kerma measurements of scatter radiation from a phantom performed without and with the shields. Protection was assessed for 3 patient-access locations (right jugular vein, right femoral artery, and left anterior chest) and for elevations ranging from 25 to 175 cm from the floor. The influence of precise placement of the ceiling-mounted upper body shield was specifically assessed. Results The utility and protection of shielding varied for the 3 access points and with elevation. For femoral artery access locations, the shields can provide at least 80% protection from scatter at all elevations; however, protection depends substantially on upper body shield position. A disposable radiation-absorbing pad can provide 35% to 70% upper body protection for procedures during which the upper body shield cannot be used effectively. Conclusions Radiation shields can provide substantial protection from radiation during cardiac interventional procedures. Shields must be thoughtfully and actively managed to provide optimum protection. Best practice guidelines for shield use are provided. (J Am Coll Cardiol Intv 2011;4: 1133-9) (C) 2011 by the American College of Cardiology Foundation
引用
收藏
页码:1133 / 1139
页数:7
相关论文
共 19 条
[1]  
[Anonymous], 2006, BEIR VII: Health Risks from Exposure to Low Levels of Ionizing Radiation
[2]  
[Anonymous], 1993, 116 NCRP MEAS
[3]   RADIATION EXPOSURE TO OPERATOR PERFORMING CARDIAC ANGIOGRAPHY WITH U-ARM SYSTEMS [J].
BALTER, S ;
SONES, FM ;
BRANCATO, R .
CIRCULATION, 1978, 58 (05) :925-932
[4]   Decreasing Operators' Radiation Exposure During Coronary Procedures: The Transradial Radiation Protection Board [J].
Behan, Miles ;
Haworth, Peter ;
Colley, Paul ;
Brittain, Michael ;
Hince, Andrew ;
Clarke, Michael ;
Ghuran, Azad ;
Saha, Mrinal ;
Hildick-Smith, David .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (01) :79-84
[5]   Medical personnel and patient dosimetry during coronary angiography and intervention [J].
Efstathopoulos, EP ;
Makrygiannis, SS ;
Kottou, S ;
Karvouni, E ;
Giazitzoglou, E ;
Korovesis, S ;
Tzanalaridou, E ;
Raptou, PD ;
Katritsis, DG .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (18) :3059-3068
[6]   ADJUSTABLE LEAD GLASS SHIELDING DEVICE FOR USE WITH AN OVER-TABLE X-RAY TUBE [J].
EUBIG, C ;
GROVES, BM ;
DAVEY, G .
RADIOLOGY, 1978, 129 (03) :816-818
[7]   IMPROVED RADIATION PROTECTION FOR PHYSICIANS PERFORMING CARDIAC-CATHETERIZATION [J].
GERTZ, EW ;
WISNESKI, JA ;
GOULD, RG ;
AKIN, JR .
AMERICAN JOURNAL OF CARDIOLOGY, 1982, 50 (06) :1283-1286
[8]  
*INT COMM RAD PROT, 2006, ANN ICRP
[9]  
International Commission on Radiological Protection, 2007, ANN ICRP, V103
[10]   Occupational radiation doses to operators performing cardiac catheterization procedures [J].
Kim, Kwang Pyo ;
Miller, Donald L. ;
Balter, Stephen ;
Kleinerman, Ruth A. ;
Linet, Martha S. ;
Kwon, Deukwoo ;
Simon, Steven L. .
HEALTH PHYSICS, 2008, 94 (03) :211-227