Sensitive determination of tetrodotoxin using column-switching liquid chromatography-mass spectrometry with electrospray ionization in mouse serum

被引:9
作者
Hayashida, M [1 ]
Hayakawa, H [1 ]
Wada, K [1 ]
Nihira, M [1 ]
Ohno, Y [1 ]
机构
[1] Nippon Med Coll, Dept Legal Med, Bunkyo Ku, Tokyo 1138602, Japan
关键词
D O I
10.1093/jat/28.1.46
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A liquid chromatography-electrospray ionization-mass spectrometry (LC-ESI-MS) method coupled with a column-switching technique was developed for the determination of tetrodotoxin (TTX) in serum. An on-column, column-switching technique was employed to analyze TTX without pretreatment of the serum. The combination of a multimode column with reversed phases and cation exchange for TTX provided successful separation and MS determination in the ESI-positive mode. A 100-μL serum sample was injected directly into a precolumn. For TTX monitored at m/z 320.1 in the selective ion monitoring mode, the calibration curve was linear within the range 0.1-100 ng/mL, and the limit of detection was 0.5 ng/mL. Recoveries were around 90% for TTX. The present method allows successful analysis of TTX in serum. In conclusion, this new method is simple, accurate, and useful for the determination of TTX and should be of benefit to both forensic and clinical toxicology.
引用
收藏
页码:46 / 49
页数:4
相关论文
共 14 条
[1]   DETERMINATION OF DRUGS BY DIRECT-INJECTION OF PLASMA INTO A BIOCOMPATIBLE EXTRACTION COLUMN BASED ON A PROTEIN-ENTRAPPED HYDROPHOBIC PHASE [J].
HERMANSSON, J ;
GRAHN, A .
JOURNAL OF CHROMATOGRAPHY A, 1994, 660 (1-2) :119-129
[2]  
Kawabata T., 1978, Food Hygiene Examination Manual, V2, P232
[3]  
MAHJOUB AE, 2000, J CHROMATOGR B, V742, P381
[4]  
MORIYA F, 1992, JPN J LEGAL MED, V46, P117
[5]  
NARITA H, 1981, B JPN SOC SCI FISH, V47, P935
[6]   FLUORIMETRIC METHOD TO DETERMINE TETRODOTOXIN [J].
NUNEZ, MT ;
FISCHER, S ;
JAIMOVICH, E .
ANALYTICAL BIOCHEMISTRY, 1976, 72 (1-2) :320-325
[7]  
Ohno Y, 1998, J TOXICOL-TOXIN REV, V17, P1
[8]   A MONOCLONAL ANTIBODY-BASED IMMUNOASSAY FOR DETECTING TETRODOTOXIN IN BIOLOGICAL SAMPLES [J].
RAYBOULD, TJG ;
BIGNAMI, GS ;
INOUYE, LK ;
SIMPSON, SB ;
BYRNES, JB ;
GROTHAUS, PG ;
VANN, DC .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 1992, 6 (02) :65-72
[9]   Screening method for seventy psychoactive drugs or drug metabolites in serum based on high-performance liquid chromatography-electrospray ionization mass spectrometry [J].
Rittner, M ;
Pragst, F ;
Bork, WR ;
Neumann, J .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2001, 25 (02) :115-124
[10]   Electrospray ionization mass spectrometry of tetrodotoxin and its analogs: Liquid chromatography/mass spectrometry, tandem mass spectrometry, and liquid chromatography/tandem mass spectrometry [J].
Shoji, Y ;
Yotsu-Yamashita, M ;
Miyazawa, T ;
Yasumoto, T .
ANALYTICAL BIOCHEMISTRY, 2001, 290 (01) :10-17