摘要
A hydroxyapatite coated femoral stem, matching a retrieved recovered from a case of third body wear, was analysed to investigate particle release from the hydroxyapatite coating. The stem was sectioned into 12 pieces, mounted in epoxy resin and polished to reveal the microstructure. One section was analysed with X-ray diffraction and Fourier transform infra-red spectroscopy. It was found that the coating contained a large amorphous phase content indicating that the coating has the potential for rapid dissolution. The small crystalline particles in the coating were mainly in the 1-8 micron size range. Exposure to dilute nitric acid showed that the coating is capable of releasing particles into the peri-implant area.
引用
收藏
页码:183 / 186
页数:4
相关论文
共 7 条
[1]   Fretting wear behavior of thermal sprayed hydroxyapatite coating lubricated with bovine albumin [J].
Fu, YQ ;
Batchelor, AW ;
Khor, KA .
WEAR, 1999, 230 (01) :98-102
[2]  
Gross KA, 1997, INT J ORAL MAX IMPL, V12, P589
[3]  
LETHINEN R, 1990, J ORAL MAXILLOFAC SU, V48, P1075
[4]   Polyethylene wear, osteolysis and acetabular loosening with an HA-coated hip prosthesis - A follow-up of 94 consecutive arthroplasties [J].
Rokkum, M ;
Brandt, M ;
Bye, K ;
Hetland, KR ;
Waage, S ;
Reigstad, A .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (04) :582-589
[5]  
Sun JS, 1998, J BIOMED MATER RES, V39, P390, DOI 10.1002/(SICI)1097-4636(19980305)39:3<390::AID-JBM7>3.0.CO
[6]  
2-E
[7]  
WANG JS, 1994, CLIN ORTHOP RELAT R, P272