Comparison of new-generation renal artery denervation systems: assessing lesion size and thermodynamics using a thermochromic liquid crystal phantom model

被引:7
作者
Al Raisi, Sara I. [1 ,2 ]
Barry, Michael T. [1 ,2 ]
Qian, Pierre [1 ,2 ]
Bhaskaran, Abhishek [2 ]
Pouliopoulos, Jim [1 ,2 ]
Kovoor, Pramesh [1 ,2 ]
机构
[1] Westmead Hosp, Dept Cardiol, POB 533,Cnr Hawkesbury & Darcy Rd, Sydney, NSW 2145, Australia
[2] Univ Sydney, Sydney, NSW, Australia
基金
英国医学研究理事会; 澳大利亚国家健康与医学研究理事会;
关键词
catheter ablation; renal sympathetic denervation; resistant hypertension; RADIOFREQUENCY ABLATION; CATHETER ABLATION; ELECTRODE; TEMPERATURE; SYMPLICITY; TISSUE;
D O I
10.4244/EIJ-D-17-00278
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: The aim of this study was to evaluate and compare lesion dimensions and thermodynamics of the new-generation multi-electrode Symplicity Spyral and the new-generation multi-electrode EnligHTN renal artery denervation systems, using a thermochromic liquid crystal phantom model. Methods and results: A previously described renal artery phantom model was used as a platform for radiofrequency ablation. A total of 32 radiofrequency ablations were performed using the multi-electrode Symplicity Spyral (n=16) and the new-generation EnligHTN systems (n=16). Both systems were used as clinically recommended by their respective manufacturer. Lesion borders were defined by the 51 degrees C isotherm. Lesion size (depth and width) was measured and compared between the two systems. Mean lesion depth was 2.15 +/- 0.02 mm for the Symplicity Spyral and 2.32 +/- 0.02 mm for the new-generation EnligHTN (p-value <0.001). Mean lesion width was 3.64 +/- 0.08 mm and 3.59 +/- 0.05 mm (p-value=0.61) for the Symplicity Spyral and the new-generation EnligHTN, respectively. Conclusions: The new-generation EnligHTN system produced lesions of greater depth compared to the Symplicity Spyral under the same experimental conditions. Lesion width was similar between both systems. Achieving greater lesion depth by use of the new-generation EnligHTN may result in better efficacy of renal artery denervation.
引用
收藏
页码:1242 / 1247
页数:6
相关论文
共 26 条
[1]   Evaluation of lesion and thermodynamic characteristics of Symplicity and EnligHTN renal denervation systems in a phantom renal artery model [J].
Al Raisi, Sara I. ;
Pouliopoulos, Jim ;
Barry, Michael T. ;
Swinnen, John ;
Thiagalingam, Aravinda ;
Thomas, Stuart P. ;
Sivagangabalan, Gopal ;
Chow, Clara ;
Chong, James ;
Kizana, Eddy ;
Kovoor, Pramesh .
EUROINTERVENTION, 2014, 10 (02) :277-284
[2]   Randomised, blinded and controlled comparative study of chemical and radiofrequency-based renal denervation in a porcine model [J].
Bertog, Stefan ;
Fischell, Tim A. ;
Vega, Felix ;
Ghazarossian, Vartan ;
Pathak, Atul ;
Vaskelyte, Laura ;
Kent, Darrin ;
Sievert, Horst ;
Ladich, Elena ;
Yahagi, Kazuyuki ;
Virmani, Renu .
EUROINTERVENTION, 2017, 12 (15) :1898-1906
[3]   A Controlled Trial of Renal Denervation for Resistant Hypertension [J].
Bhatt, Deepak L. ;
Kandzari, David E. ;
O'Neill, William W. ;
D'Agostino, Ralph ;
Flack, John M. ;
Katzen, Barry T. ;
Leon, Martin B. ;
Liu, Minglei ;
Mauri, Laura ;
Negoita, Manuela ;
Cohen, Sidney A. ;
Oparil, Suzanne ;
Rocha-Singh, Krishna ;
Townsend, Raymond R. ;
Bakris, George L. .
NEW ENGLAND JOURNAL OF MEDICINE, 2014, 370 (15) :1393-1401
[4]   RENAL AFFERENT DENERVATION PREVENTS THE PROGRESSION OF RENAL-DISEASE IN THE RENAL ABLATION MODEL OF CHRONIC-RENAL-FAILURE IN THE RAT [J].
CAMPESE, VM ;
KOGOSOV, E ;
KOSS, M .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1995, 26 (05) :861-865
[5]   High Spatial Resolution Thermal Mapping of Radiofrequency Ablation Lesions Using a Novel Thermochromic Liquid Crystal Myocardial Phantom [J].
Chik, William W. B. ;
Barry, Michael Anthony ;
Thavapalachandran, Sujitha ;
Midekin, Christine ;
Pouliopoulos, Jim ;
Lim, Toon Wei ;
Sivagangabalan, Gopal ;
Thomas, Stuart P. ;
Ross, David L. ;
McEwan, Alistair L. ;
Kovoor, Pramesh ;
Thiagalingam, Aravinda .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (11) :1278-1286
[6]  
Cosman E.R., 2009, TXB STEREOTACTIC FUN, P1359
[7]   Electric and thermal field effects in tissue around radiofrequency electrodes [J].
Cosman, ER ;
Cosman, ER .
PAIN MEDICINE, 2005, 6 (06) :405-424
[8]   Factors That Affect Radiofrequency Heat Lesion Size [J].
Cosman, Eric R., Jr. ;
Dolensky, Joseph R. ;
Hoffman, Ryan A. .
PAIN MEDICINE, 2014, 15 (12) :2020-2036
[9]   Physiology in perspective:: The Wisdom of the Body.: Neural control of the kidney [J].
DiBona, GF .
AMERICAN JOURNAL OF PHYSIOLOGY-REGULATORY INTEGRATIVE AND COMPARATIVE PHYSIOLOGY, 2005, 289 (03) :R633-R641
[10]   MEASUREMENT OF TOTAL AND ORGAN-SPECIFIC NOREPINEPHRINE KINETICS IN HUMANS [J].
ESLER, M ;
JENNINGS, G ;
KORNER, P ;
BLOMBERY, P ;
SACHARIAS, N ;
LEONARD, P .
AMERICAN JOURNAL OF PHYSIOLOGY, 1984, 247 (01) :E21-E28