Automated function imaging: a new operator-independent strain method for assessing left ventricular function

被引:100
作者
Belghiti, H. [1 ]
Brette, S. [1 ]
Lafitte, S. [1 ]
Reant, P. [1 ]
Picard, F. [1 ]
Serri, K. [1 ]
Lafitte, M. [1 ]
Courregelongue, M. [1 ]
Dos Santos, P. [1 ]
Douard, H. [1 ]
Roudaut, R. [1 ]
DeMaria, A. [2 ]
机构
[1] Hop Cardiol Haut Leveque, Dept Cardiol Chirug, F-33600 Pessac, France
[2] Univ Calif San Diego, San Diego, CA 92103 USA
关键词
speckle tracking; regional left; ventricular function; strain; AFI;
D O I
10.1016/S1875-2136(08)71798-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. - Speckle tracking is a new technique based on pure 2D grayscale ultrasound acquisition allowing calculation of segmental strains. To facilitate clinical application, speckle tracking has been integrated into the most recent echocardiographic systems for quick, automated evaluation of left ventricular function (Automated Function Imaging, AFI). Objective. - To evaluate the feasibility, calculation time, accuracy and reproducibility of global longitudinal strain (GLS) from AFI in determining LV function in comparison to reference echocardiographic and angiographic methods-whatever the operator's experience. Methods. - Echocardiography was performed in 65 patients scheduled for cardiac catheterization using a Vivid 7 system. They were divided into 3 groups according to EF (> 55%, 35sEFS55%, < 35%). Image quality, global LV function parameters (ejection fraction, aortic flow, dp/dt) and segmental contraction were analyzed by one experienced operator and one beginner. GLS was obtained from apical 2, 3 and 4 chamber views. GLS was compared to both echocardiographic and angiographic EF, as well as to other echocardiographic parameters. Results. - GLS was obtained successfully in 97% of patients. Mean calculation time including correction of endocardial. detection was less than 60 seconds. GLS was significantly different between the 3 groups, respectively -18.1 +/- 2.5 %, -11.5 +/- 2.1% and -6.0 +/- 1.2% (p < 0.01). Strong correlations were observed between GLS and LV function (r from 0.68 to 0.87) with a high level of reproducibility. No difference was observed between experienced and inexperienced operators. Conclusion. - AFI is clinically applicable and an effective means of assessing LV function due to its short acquisition time, feasibility and accuracy, whatever the experience of the operator. (c) 2008 Published by Elsevier Masson SAS.
引用
收藏
页码:163 / 169
页数:7
相关论文
共 23 条
  • [1] Noninvasive myocardial strain measurement by speckle tracking echocardiography - Validation against sonomicrometry and tagged magnetic resonance imaging
    Amundsen, BH
    Helle-Valle, T
    Edvardsen, T
    Torp, H
    Crosby, J
    Lyseggen, E
    Stoylen, A
    Ihlen, H
    Lima, JAC
    Smiseth, OA
    Slordahl, SA
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (04) : 789 - 793
  • [2] The combined effect of nonlinear filtration and window size on the accuracy of tissue displacement estimation using detected echo signals
    Behar, V
    Adam, D
    Lysyansky, P
    Friedman, Z
    [J]. ULTRASONICS, 2004, 41 (09) : 743 - 753
  • [3] Comparison of two-dimensional speckle and tissue velocity based strain and validation with harmonic phase magnetic resonance imaging
    Cho, Goo-Yeong
    Chan, Jonathan
    Leano, Rodel
    Strudwick, Mark
    Marwick, Thomas H.
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (11) : 1661 - 1666
  • [4] Echocardiographic strain and strain-rate imaging: A new tool to study regional myocardial function
    D'hooge, J
    Bijnens, B
    Thoen, J
    Van de Werf, F
    Sutherland, GR
    Suetens, P
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (09) : 1022 - 1030
  • [5] Two-dimensional ultrasonic strain rate measurement of the human heart in vivo
    D'hooge, J
    Konofagou, E
    Jamal, F
    Heimdal, A
    Barrios, L
    Bijnens, B
    Thoen, J
    Van de Werf, F
    Sutherland, G
    Suetens, P
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (02) : 281 - 286
  • [6] Quantitative assessment of intrinsic regional myocardial deformation by Doppler strain rate echocardiography in humans - Validation against three-dimensional tagged magnetic resonance imaging
    Edvardsen, T
    Gerber, BL
    Garot, J
    Bluemke, DA
    Lima, JAC
    Smiseth, OA
    [J]. CIRCULATION, 2002, 106 (01) : 50 - 56
  • [7] TRANSMURAL MYOCARDIAL DEFORMATION IN CANINE LEFT-VENTRICULAR WALL
    FENTON, TR
    CHERRY, JM
    KLASSEN, GA
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1978, 235 (05): : H523 - H530
  • [8] RELATIONSHIP BETWEEN MYOCARDIAL FIBER DIRECTION AND SEGMENT SHORTENING IN THE MIDWALL OF THE CANINE LEFT-VENTRICLE
    FREEMAN, GL
    LEWINTER, MM
    ENGLER, RL
    COVELL, JW
    [J]. CIRCULATION RESEARCH, 1985, 56 (01) : 31 - 39
  • [9] Quantification of regional contractile function after infarction:: Strain analysis superior to wall thickening analysis in discriminating infarct from remote myocardium
    Götte, MJW
    van Rossum, AC
    Twisk, JWR
    Kuijer, JPA
    Marcus, JT
    Visser, CA
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (03) : 808 - 817
  • [10] Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy
    Grothues, F
    Smith, GC
    Moon, JCC
    Bellenger, NG
    Collins, P
    Klein, HU
    Pennell, DJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (01) : 29 - 34