Steady-state MR imaging sequences: Physics, classification, and clinical applications

被引:199
作者
Chavhan, Govind B. [1 ,2 ]
Babyn, Paul S. [1 ,2 ]
Jankharia, Bhavin G. [3 ]
Cheng, Hai-Ling M. [1 ,2 ]
Shroff, Manohar M. [1 ,2 ]
机构
[1] Hosp Sick Children, Dept Diagnost Imaging, Toronto, ON M5G 1X8, Canada
[2] Univ Toronto, Toronto, ON M5G 1X8, Canada
[3] Jankharia Imaging Ctr, Mumbai, Maharashtra, India
关键词
D O I
10.1148/rg.284075031
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Steady-state sequences are a class of rapid magnetic resonance (MR) imaging techniques based on fast gradient-echo acquisitions in which both longitudinal magnetization (LM) and transverse magnetization (TM) are kept constant. Both LM and TM reach a nonzero steady state through the use of a repetition time that is shorter than the T2 relaxation time of tissue. When TM is maintained as multiple radio-frequency excitation pulses are applied, two types of signal are formed once steady state is reached: preexcitation signal (S-) from echo reformation; and postexcitation signal (S+), which consists of free induction decay. Depending on the signal sampled and used to form an image, steady-state sequences can be classified as (a) postexcitation refocused (only S+ is sampled), (b) preexcitation refocused (only S- is sampled), and (c) fully refocused (both S+ and S- are sampled) sequences. All tissues with a reasonably long T2 relaxation time will show additional signals due to various refocused echo paths. Steady-state sequences have revolutionized cardiac imaging and have become the standard for anatomic functional cardiac imaging and for the assessment of myocardial viability because of their good signal-to-noise ratio and contrast-to-noise ratio and increased speed of acquisition. They are also useful in abdominal and fetal imaging and hold promise for interventional MR imaging. Because steady-state sequences are now commonly used in MR imaging, radiologists will benefit from understanding the underlying physics, classification, and clinical applications of these sequences. (C) RSNA, 2008.
引用
收藏
页码:1147 / 1160
页数:14
相关论文
共 42 条
[1]   Fat attenuation using a dual steady-state balanced-SSFP sequence with periodically variable flip angles [J].
Absil, J ;
Denolin, V ;
Metens, T .
MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (02) :343-351
[2]  
[Anonymous], 1999, MAGN RESON IMAGING
[3]   Differentiation of benign osteoporotic and neoplastic vertebral compression fractures with a diffusion-weighted, steady-state free precession sequence [J].
Baur, A ;
Huber, A ;
Dürr, HR ;
Nikolaou, K ;
Stäbler, A ;
Deimling, M ;
Reiser, M .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (01) :70-75
[4]   Three-dimensional breathhold SSFP coronary MRA: A comparison between 1.5T and 3.OT [J].
Bi, XM ;
Deshpande, V ;
Simonetti, O ;
Laub, G ;
Li, DB .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 22 (02) :206-212
[5]   Comparison of ECG-gated rectilinear vs. real-time radial K-space sampling schemes in cine true-FISP cardiac MRI [J].
Boll, DT ;
Merkle, EM ;
Seaman, DM ;
Gilkeson, RC ;
Larson, AP ;
Simonetti, OP ;
Duerk, JL ;
Lewin, JS .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2004, 6 (04) :793-802
[6]   Comparison of MR imaging sequences for liver and head and neck interventions: Is there a single optimal sequence for all purposes? [J].
Boll, DT ;
Lewin, JS ;
Duerk, JL ;
Aschoff, AJ ;
Merkle, EM .
ACADEMIC RADIOLOGY, 2004, 11 (05) :506-515
[7]   MR imaging abbreviations, definitions, and descriptions: A review [J].
Brown, MA ;
Semelka, RC .
RADIOLOGY, 1999, 213 (03) :647-662
[8]  
CARROLL TJ, 2006, CLIN MAGNETIC RESONA, P187
[9]   T2-weighted fast MR imaging with true FISP versus HASTE: Comparative efficacy in the evaluation of normal fetal brain maturation [J].
Chung, HW ;
Chen, CY ;
Zimmerman, RA ;
Lee, KW ;
Lee, CC ;
Chin, SC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (05) :1375-1380
[10]  
Chung YC, 1999, MAGNET RESON MED, V42, P335, DOI 10.1002/(SICI)1522-2594(199908)42:2<335::AID-MRM16>3.0.CO