Halfpipe Coaxial Cannula for Self-Contained Vacuum-Assisted Biopsy Systems: Feasibility in a Pig Breast Model

被引:0
作者
Teubner, Tobias [1 ,2 ]
Haefner, Matthias F. [1 ]
Schimmele, Max [1 ]
Teubner, Joachim [1 ,2 ]
机构
[1] St Elizabeth Hosp, Diagnost Breast Ctr, D-69121 Heidelberg, Germany
[2] Univ Heidelberg, Dept Radiol & Nucl Med, Univ Hosp Mannheim, D-6800 Mannheim, Germany
关键词
breast; coaxial systems; stereotactic biopsy; vacuum-assisted biopsy; CORE BIOPSY; NEEDLE;
D O I
10.2214/AJR.09.2876
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this article is to discuss the development of a dedicated halfpipe coaxial cannula for stereotactic vacuum-assisted biopsy. We evaluated the system by retrieving 18 copper targets from a pig breast model in the upright position via vertical and lateral approaches. CONCLUSION. Sampling was successful in 15 of 18 cases. Errors occurred only in superficial lesions biopsied via the vertical approach. The halfpipe coaxial cannula shows promise for improving positioning accuracy, avoiding target dislocation, and obviating repeated needle repositioning.
引用
收藏
页码:W563 / W566
页数:4
相关论文
共 8 条
  • [1] Stereotactic breast biopsy of atypical ductal hyperplasia and ductal carcinoma in situ lesions: Improved accuracy with directional, vacuum-assisted biopsy
    Burbank, F
    [J]. RADIOLOGY, 1997, 202 (03) : 843 - 847
  • [2] Are malignant cells displaced by large-gauge needle core biopsy of the breast?
    Diaz, LK
    Wiley, EL
    Venta, LA
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (05) : 1303 - 1313
  • [3] Stereotactic biopsy of the breast using an upright unit, a vacuum-suction needle, and a lateral arm-support system
    Georgian-Smith, D
    D'Orsi, C
    Morris, E
    Clark, CF
    Liberty, E
    Lehman, CD
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 178 (04) : 1017 - 1024
  • [4] Coaxial technique: Approach to breast core biopsies
    Helbich, TH
    Mayr, W
    Schick, S
    Youssefzadeh, S
    Rudas, M
    Taucher, S
    Wagner, T
    Kelkar, P
    Wolf, G
    Thurnher, M
    Mostbeck, GH
    [J]. RADIOLOGY, 1997, 203 (03) : 684 - 690
  • [5] Interdisciplinary consensus on the use and technique of vacuum-assisted stereotactic breast biopsy
    Heywang-Köbrunner, SH
    Schreer, I
    Decker, T
    Böcker, W
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2003, 47 (03) : 232 - 236
  • [6] US-GUIDED CORE BIOPSY OF THE BREAST WITH A COAXIAL SYSTEM
    KAPLAN, SS
    RACENSTEIN, MJ
    WONG, WS
    HANSEN, GC
    MCCOMBS, MM
    BASSETT, LW
    [J]. RADIOLOGY, 1995, 194 (02) : 573 - 575
  • [7] Stereotactic vacuum-assisted breast biopsy in 2874 patients -: A Multicenter study
    Kettritz, U
    Rotter, K
    Schreer, I
    Murauer, M
    Schulz-Wendtiand, R
    Peter, D
    Heywang-Köbrunner, SH
    [J]. CANCER, 2004, 100 (02) : 245 - 251
  • [8] First experiences with a new vacuum-assisted device for breast biopsy
    Schulz-Wendtland, R
    Krämer, S
    Bautz, W
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2003, 175 (11): : 1496 - 1499