DEVELOPMENT OF A VALIDATED METHOD FOR SIMULTANEOUS QUANTIFICATION OF CLARITHROMYCIN AND LEVOFLOXACIN BY HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY-DIODE ARRAY DETECTOR

被引:2
|
作者
Ijaz, Qazi Amir [1 ]
Abbas, Nasir [1 ]
Latif, Sumera [2 ]
Hussain, Amjad [1 ]
Arshad, Muhammad Sohail [3 ]
Ali, Ejaz [1 ]
Rashid, Memoona [1 ]
Sohail, Kashif [4 ]
Riaz, Sohail [4 ]
Bukhari, Nadeem Irfan [1 ]
机构
[1] Univ Punjab, Univ Coll Pharm, Lahore, Pakistan
[2] Lahore Coll Women Univ, Fac Pharmaceut & Allied Hlth Sci, Inst Pharm, Lahore, Pakistan
[3] Bahauddin Zakariya Univ, Dept Pharmaceut, Multan, Pakistan
[4] Mirpur Univ Sci & Technol, Akron Coll Pharm, Mirpur, AJ&K, Pakistan
来源
ACTA POLONIAE PHARMACEUTICA | 2020年 / 77卷 / 06期
关键词
clarithromycin; levofloxacin; high-performance liquid chromatography; diode array detector; validation; simultaneous quantification; HPLC METHOD; IN-VITRO; RP-HPLC; PHARMACOKINETICS; COMBINATIONS; IMPURITIES; BULK;
D O I
10.32383/appdr/131197
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
This study aims at the development of a validated reverse phase-high performance liquid chromatography (RP-HPLC) method for the simultaneous quantification of levofloxacin (LVX) and clarithromycin (CAM). Eclipse X DB C18 column (5 mu m, 4.6 x250 mm) was used as a stationary phase, and monobasic potassium phosphate (KH2PO4) buffer (0.035 M): acetonitrile (75 : 25) was used as an isocratic mobile phase. The pH of the mobile phase was adjusted to 4.5 with the help of diluted orthophosphoric acid. The diode array detector (DAD) was operated at 205 nm and 294 nm for the detection of CAM and LVX, respectively. Limit of detection (LOD) values for CAM and LVX were 1.22 mu g/mL and 0.79 mu g/mL, respectively. Limit of quantification(LOQ) values for CAM and LVX were 4.08 mu g/mL and 2.64 mu g/mL, respectively. The method was precise as% relative standard deviation (% RSD) values were less than 2% for all types of precision. This method was successfully applied to determine the percentage contents of both drugs in the pharmaceutical preparation.
引用
收藏
页码:841 / 847
页数:7
相关论文
共 50 条
  • [1] Simultaneous quantification of Gemcitabine and Irinotecan hydrochloride in rat plasma by using high performance liquid chromatography-diode array detector
    Malatesta, L.
    Cosco, D.
    Paolino, D.
    Cilurzo, F.
    Costa, N.
    Di Tullio, A.
    Fresta, M.
    Celia, C.
    Di Marzio, L.
    Locatelli, M.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 159 : 192 - 199
  • [2] Development and validation of an high-performance liquid chromatography-diode array detector method for the simultaneous determination of six phenolic compounds in abnormal savda munziq decoction
    Tian, Shuge
    Liu, Wenxian
    Liu, Feng
    Zhang, Xuejia
    Upur, Halmuart
    PHARMACOGNOSY MAGAZINE, 2015, 11 (41) : 157 - 162
  • [3] Quantification of maltol in Korean ginseng (Panax ginseng) products by high-performance liquid chromatography-diode array detector
    Jeong, Hyun Cheol
    Hong, Hee-Do
    Kim, Young-Chan
    Rhee, Young Kyoung
    Choi, Sang Yoon
    Kim, Kyung-Tack
    Kim, Sung Soo
    Lee, Young-Chul
    Cho, Chang-Won
    PHARMACOGNOSY MAGAZINE, 2015, 11 (43) : 657 - 664
  • [4] High Performance Liquid Chromatography-Diode Array Detector Method for the Simultaneous Determination of Five Compounds in the Pulp and Seed of Sea Buckthorn
    Zhao, Lu
    Wen, E.
    Upur, Halmuart
    Tian, Shuge
    PHARMACOGNOSY MAGAZINE, 2017, 13 (49) : 136 - 140
  • [5] Development and validation of high-performance liquid chromatography method for determination of clarithromycin in pharmaceutical tablets
    Mahmoudi, Abdelghani
    De Francia, Silvia
    Paul, Prasanta
    JOURNAL OF SEPARATION SCIENCE, 2023, 46 (21)
  • [6] Simultaneous Determination of Tylosin Tartrate and Tylvalosin Tartrate in Wastewater by the Solid-Phase Extraction and High-Performance Liquid Chromatography-Diode Array Detector Method
    Chai, Ruina
    Cheng, Jianguo
    Zhang, Shuheng
    Ma, Litong
    Duan, Jianguo
    Zhao, Wenyuan
    Wang, Xiaoxia
    Fang, Chen
    SEPARATION SCIENCE PLUS, 2025, 8 (01):
  • [7] Curcuminoid's Content and Fingerprint Analysis for Authentication and Discrimination of Curcuma xanthorrhiza from Curcuma longa by High-Performance Liquid Chromatography-Diode Array Detector
    Rafi, Mohamad
    Wulansari, Laela
    Heryanto, Rudi
    Darusman, Latifah Kosim
    Lim, Lee Wah
    Takeuchi, Toyohide
    FOOD ANALYTICAL METHODS, 2015, 8 (09) : 2185 - 2193
  • [8] Simultaneous Determination of Four Marker Components in Yukmijihwang Tang by High Performance Liquid Chromatography/Diode Array Detector
    Won, Jin Bae
    Ma, Jin Yeul
    Ma, Choong Je
    ARCHIVES OF PHARMACAL RESEARCH, 2010, 33 (04) : 619 - 623
  • [9] Simultaneous determination of four marker components in Yukmijihwang Tang by high performance liquid chromatography/diode array detector
    Jin Bae Won
    Jin Yeul Ma
    Choong Je Ma
    Archives of Pharmacal Research, 2010, 33 : 619 - 623
  • [10] A novel high-performance liquid chromatography with diode array detector method for the simultaneous quantification of the enzyme-reactivating oximes obidoxime, pralidoxime, and HI-6 in human plasma
    Kranawetvogl, Tamara
    Steinritz, Dirk
    Thiermann, Horst
    John, Harald
    DRUG TESTING AND ANALYSIS, 2020, 12 (07) : 938 - 947