Validation of a microtiter plate ELISA for screening of postmortem blood for opiates and benzodiazepines

被引:15
|
作者
Kemp, P
Sneed, G
Kupiec, T
Spiehler, V
机构
[1] Off Med Examiner, Oklahoma City, OK USA
[2] Analyt Res Labs Inc, Oklahoma City, OK USA
[3] Spiehler & Associates, Newport Beach, CA USA
关键词
D O I
10.1093/jat/26.7.504
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The object of this study was to evaluate the suitability of the Neogen Corp. microtiter plate enzyme-linked immunoassays (ELISA) for opiates and benzodiazepines for screening of postmortem blood. Ninety postmortem whole blood specimens were obtained from drug-involved deaths which had been screened and confirmed positive for opiates and/or benzodiazepines. Forty negative specimens were obtained from non-opiate-involved deaths. Specimens were tested using the Neogen Opiates Group and Neogen Benzodiazepines Group microtiter plate ELISA assays. No matrix effects were found for whole blood in these assays and a dilution of 1:5 was chosen to facilitate pipetting and to bring the IC50 of the microtiter plate ELISA assay within the range of opiates and benzodiazepines encountered in medical examiner specimens. True positive, true negatives, false positives, and false negatives were determined and graphed for the ELISA results against gas chromatography-mass spectrometry (GC-MS), gas chromatography-nitrogen-phosphorus detection and case histories. From these graphs and the ROC curves, the optimal cut-off for the Neogen Opiates Group ELISA was found to be between 20 and 50 ng/mL morphine equivalents and the optimum cut-off for the Neogen Benzodiazepines Group ELISA was between 20 and 50 ng/mL temazepam equivalents. The Neogen Opiates Group ELISA had a sensitivity of 95.2% ± 2.7% and a specificity of 92.2% ± 3.4% versus GC-MS at a cut-off of 20-ng/mL cut-off and a sensitivity of 88.8% ± 3.9% and specificity of 96.8% ± 2.1% versus GC-MS at a 50-ng/mL morphine equivalents cut-off. The Neogen Benzodizepines Group ELISA had a sensitivity of 100% ± 1.3% and a specificity of 94.6% ± 2.9% versus GC-MS (20-ng/mL temazepam equivalents cut-off) and a sensitivity of 95.8% ± 2.5% and specificity of 98.2% ± 1.8% versus GC-MS at a 50-ng/mL cut-off.
引用
收藏
页码:504 / 512
页数:9
相关论文
共 50 条
  • [1] Performance of a microtiter plate ELISA for screening of postmortem blood for cocaine and metabolites
    Spiehler, V
    Isenschmid, DS
    Matthews, P
    Kemp, P
    Kupiec, T
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2003, 27 (08) : 587 - 591
  • [2] Validation of an ELISA method for screening methadone in postmortem blood
    Juhascik, Matthew
    Habbel, Steven
    Barron, William
    Behonick, George
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2006, 30 (08) : 617 - 620
  • [3] 'Cold' incubation in ELISA microtiter plate assays
    Becker, U
    BLOOD COAGULATION & FIBRINOLYSIS, 1997, 8 (01) : 72 - 72
  • [4] Screening postmortem whole blood for oxycodone by ELISA response ratios
    Spiehler, V
    DeCicco, L
    McCutcheon, JR
    Kupiec, T
    Kemp, P
    JOURNAL OF FORENSIC SCIENCES, 2004, 49 (03) : 621 - 626
  • [5] Choice of an ELISA assay for screening postmortem blood for amphetamine and/or methamphetamine
    Kupiec, T
    DeCicco, L
    Spiehler, V
    Sneed, G
    Kemp, P
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2002, 26 (07) : 513 - 518
  • [6] POSTMORTEM STABILITY OF BENZODIAZEPINES IN BLOOD AND TISSUES
    LEVINE, B
    BLANKE, RV
    VALENTOUR, JC
    JOURNAL OF FORENSIC SCIENCES, 1983, 28 (01) : 102 - 115
  • [7] Development of a microtiter plate ELISA and a dipstick ELISA for the determination of the organophosphorus insecticide fenthion
    Cho, YA
    Kim, YJ
    Hammock, BD
    Lee, YT
    Lee, HS
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2003, 51 (27) : 7854 - 7860
  • [8] Validation of the Cozart® microplate ELISA for detection of opiates in hair
    Cooper, G
    Wilson, L
    Reid, C
    Baldwin, D
    Hand, C
    Spiehler, V
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2003, 27 (08) : 581 - 586
  • [9] A MICROTITER ALT ASSAY FOR BLOOD SCREENING
    THOMPSON, L
    FERRERA, C
    STURGE, J
    BEATY, L
    TRANSFUSION, 1987, 27 (06) : 523 - 523
  • [10] THIN-LAYER CHROMATOGRAPHIC SCREENING FOR METHAQUALONE, PHENOTHIAZINES, OPIATES AND BENZODIAZEPINES
    BREITER, J
    HELGER, R
    INTERSCHICK, E
    WUST, H
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1978, 16 (02): : 127 - 134