Measurement of the MTF of a Cone-Beam Breast Computed Tomography Laboratory Scanner

被引:20
|
作者
Mettivier, Giovanni [1 ,2 ]
Russo, Paolo [1 ,2 ]
机构
[1] Univ Naples Federico II, Lab Med Phys, Dipartimento Sci Fis, I-80126 Naples, Italy
[2] Ist Nazl Fis Nucl, Sez Napoli, I-80126 Naples, Italy
关键词
Computed tomography; image quality; mammography; PRESAMPLED MTF; CT; CALIBRATION; MISALIGNMENT; SYSTEM;
D O I
10.1109/TNS.2011.2105503
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We measured the spatial resolution of the prototype scanner developed for laboratory investigations on Cone-Beam Breast Computed Tomography (CBBCT). The scanner has potential for high spatial resolution, since it employs a minifocus X-ray tube (0.04 mm nominal focal spot size) and a fine pitch (0.05 mm) flat panel detector. With the available maximum tube current (0.25 mA) and detector field of view (120 x 120 mm(2)), adequate sample volume coverage and exposure requirements as in a clinical scan are obtained via increasing the total acquisition time and by combining laterally translated projections at each angular position. The Modulation Transfer Function (MTF) of the system was evaluated in step-and-shoot acquisitions with the tilted wire method, by imaging a test pattern made by an array of vertical and horizontal thin metal wires. MTF measurements in each tomographic plane were performed. Spatial frequency values of 3-3.5 mm(-1) at 10% MTF were measured. A lower spatial resolution performance is expected in continuous acquisitions, due to the gantry motion blur.
引用
收藏
页码:703 / 713
页数:11
相关论文
共 50 条
  • [21] Evaluating the mandible with cone-beam computed tomography
    Stratemann, Scott A.
    Huang, John C.
    Maki, Koutaro
    Hatcher, David C.
    Miller, Arthur J.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (04) : S58 - S70
  • [22] Simple ROI Cone-Beam Computed Tomography
    Maass, Clemens
    Knaup, Michael
    Sawall, Stefan
    Kachelriess, Marc
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3161 - 3168
  • [23] Operational principles for cone-beam computed tomography
    Hatcher, David C.
    JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2010, 141 (10): : 3S - 6S
  • [24] Cone-Beam Computed Tomography: Beyond Diagnostics
    Perel, Morton L.
    IMPLANT DENTISTRY, 2015, 24 (04) : 367 - 367
  • [25] Cone-Beam Computed Tomography (CBCT) to Computed Tomography (CT) Image Translation with Perceptual Loss to Eliminate Cone-Beam Artifacts
    Yoo, S.
    Choi, B.
    Park, I.
    Kim, J.
    Kim, H.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [26] 3D MTF estimation using sphere phantoms for cone-beam computed tomography systems
    Lee, Changwoo
    Song, Hoon-dong
    Baek, Jongduk
    MEDICAL PHYSICS, 2020, 47 (07) : 2838 - 2851
  • [27] 3-D MTF Estimation Using Sphere Phantoms for Cone-Beam Computed Tomography Systems
    Song, Hoon-dong
    Lee, Changwoo
    Baek, Jongduk
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [28] Newer Technologies in Breast Cancer Imaging: Dedicated Cone-Beam Breast Computed Tomography
    O'Connell, Avice M.
    Karellas, Andrew
    Vedantham, Srinivasan
    Kawakyu-O'Connor, Daniel T.
    SEMINARS IN ULTRASOUND CT AND MRI, 2018, 39 (01) : 106 - 113
  • [29] Computed gray levels in multislice and cone-beam computed tomography
    Azeredo, Fabiane
    de Menezes, Luciane Macedo
    Enciso, Reyes
    Weissheimer, Andre
    de Oliveira, Rogerio Belle
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2013, 144 (01) : 147 - 155
  • [30] Dose Distribution in Cone-Beam Breast Computed Tomography: An Experimental Phantom Study
    Russo, Paolo
    Lauria, Adele
    Mettivier, Giovanni
    Montesi, Maria Cristina
    Villani, Natalia
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 366 - 374