Kinematic Analysis of a Posterior-stabilized Knee Prosthesis

被引:11
作者
Zhao, Zhi-Xin [1 ]
Wen, Liang [2 ]
Qu, Tie-Bing [2 ]
Hou, Li-Li [3 ]
Xiang, Dong [3 ]
Bin, Jia [1 ]
机构
[1] Beijing DCN Orthopaed Hosp, Dept Orthoped, Beijing 100143, Peoples R China
[2] Capital Med Univ, Beijing Chao Yang Hosp, Dept Orthoped, Beijing 100020, Peoples R China
[3] Naton Inst Med Technol, Dept Biomech, Beijing 100094, Peoples R China
关键词
Arthroplasty; Knee; Kinematics; Posterior-stabilized Prosthesis; Replacement; REPLACEMENT; JOINT;
D O I
10.4103/0366-6999.149205
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The goal of total knee arthroplasty (TKA) is to restore knee kinematics. Knee prosthesis design plays a very important role in successful restoration. Here, kinematics models of normal and prosthetic knees were created and validated using previously published data. Methods: Computed tomography and magnetic resonance imaging scans of a healthy, anticorrosive female cadaver were used to establish a model of the entire lower limbs, including the femur, tibia, patella, fibula, distal femur cartilage, and medial and lateral menisci, as well as the anterior cruciate, posterior cruciate, medial collateral, and lateral collateral ligaments. The data from the three-dimensional models of the normal knee joint and a posterior-stabilized (PS) knee prosthesis were imported into finite element analysis software to create the final kinematic model of the TKA prosthesis, which was then validated by comparison with a previous study. The displacement of the medial/lateral femur and the internal rotation angle of the tibia were analyzed during 0-135 degrees flexion. Results: Both the output data trends and the measured values derived from the normal knee's kinematics model were very close to the results reported in a previous in vivo study, suggesting that this model can be used for further analyses. The PS knee prosthesis underwent an abnormal forward displacement compared with the normal knee and has insufficient, or insufficiently aggressive, "rollback" compared with the lateral femur of the normal knee. In addition, a certain degree of reverse rotation occurs during flexion of the PS knee prosthesis. Conclusions: There were still several differences between the kinematics of the PS knee prosthesis and a normal knee, suggesting room for improving the design of the PS knee prosthesis. The abnormal kinematics during early flexion shows that the design of the articular surface played a vital role in improving the kinematics of the PS knee prosthesis.
引用
收藏
页码:216 / 221
页数:6
相关论文
共 21 条
  • [1] Three-dimensional dynamic behaviour of the human knee joint under impact loading
    Abdel-Rahman, EM
    Hefzy, MS
    [J]. MEDICAL ENGINEERING & PHYSICS, 1998, 20 (04) : 276 - 290
  • [2] In vivo three-dimensional knee kinematics using a biplanar image-matching technique
    Asano, T
    Akagi, M
    Tanaka, K
    Tamura, J
    Nakamura, T
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (388) : 157 - 166
  • [3] 2003 Hap Paul Award paper of the international society for technology in arthroplasty
    Banks, SA
    Hodge, WA
    [J]. JOURNAL OF ARTHROPLASTY, 2004, 19 (07) : 809 - 816
  • [4] Beverland David, 2010, Orthopedics, V33, P657, DOI 10.3928/01477447-20100722-23
  • [5] RECRUITMENT OF KNEE-JOINT LIGAMENTS
    BLANKEVOORT, L
    HUISKES, R
    DELANGE, A
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1991, 113 (01): : 94 - 103
  • [6] Buechel Frederick F Sr, 2002, J Knee Surg, V15, P84
  • [7] PATIENT OUTCOMES FOLLOWING TRICOMPARTMENTAL TOTAL KNEE REPLACEMENT - A METAANALYSIS
    CALLAHAN, CM
    DRAKE, BG
    HECK, DA
    DITTUS, RS
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1994, 271 (17): : 1349 - 1357
  • [8] Intraoperative evaluation of total knee replacement: kinematic assessment with a navigation system
    Casino, Daniela
    Zaffagnini, Stefano
    Martelli, Sandra
    Lopomo, Nicola
    Bignozzi, Simone
    Iacono, Francesco
    Russo, Alessandro
    Marcacci, Maurilio
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2009, 17 (04) : 369 - 373
  • [9] In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics
    Dennis, DA
    Mahfouz, MR
    Komistek, RD
    Hoff, W
    [J]. JOURNAL OF BIOMECHANICS, 2005, 38 (02) : 241 - 253
  • [10] A multicenter analysis of axial femorotibial rotation after total knee arthroplasty
    Dennis, DA
    Komistek, RD
    Mahfouz, MR
    Walker, SA
    Tucker, A
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) : 180 - 189