Dynamic shielding of the magnetic fields

被引:22
作者
Baltag, Octavian [1 ,2 ]
Costandache, Doina [1 ]
Rau, Miuta [1 ]
Iftemie, Anca [1 ]
Rau, Ion [1 ]
机构
[1] Gr T Popa Univ Med & Pharm, RO-700115 Iasi, Romania
[2] Terraflux Control Ltd, RO-700397 Iasi, Romania
关键词
active shielding; biomagnetics; dynamic shielding; magnetic field; magnetic shielding; LOW-FREQUENCY; PERFORMANCE; COMPENSATION; ROOM;
D O I
10.4316/AECE.2010.04022
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
The paper presents a comparative study of the methods used to control and compensate the direct and alternative magnetic fields. Two frequently used methods in the electromagnetic compatibility of the complex biomagnetism installations were analyzed. The two methods refer to the use of inductive magnetic field sensors (only for alternative fields) and of fluxgate magnetometers as active transducers which measures both the direct and alternative components of the magnetic field. The applications of the dynamic control of the magnetic field are: control of the magnetic field of the military ships, control of parasite magnetic field produced by power transformers and the electrical networks, protection of the mass spectrometers, electronic microscopes, SQUID and optical pumping magnetometers for applications in biomagnetism.
引用
收藏
页码:135 / 142
页数:8
相关论文
共 37 条
[1]   Magnetic field cancellation techniques for the Mars Global Surveyor solar array [J].
Acuna, MH ;
Scheifele, JL ;
Stella, P ;
Kloss, C ;
Smith, B ;
Heinsohn, G ;
Sharmit, K .
CONFERENCE RECORD OF THE TWENTY FIFTH IEEE PHOTOVOLTAIC SPECIALISTS CONFERENCE - 1996, 1996, :325-328
[2]  
[Anonymous], NEUROLOGY CLIN NEURO
[3]  
BALTAG O, 2003, MAGNETOMETRIE
[4]  
BALTAG O, 1995, P 11 INT S EMC ZUR
[5]   Magnetic field stabilization for magnetically shielded volumes by external field coils [J].
Brys, T ;
Czekaj, S ;
Daum, M ;
Fierlinger, P ;
George, D ;
Henneck, R ;
Hochman, Z ;
Kasprzak, M ;
Kohlik, K ;
Kirch, K ;
Kuzniak, M ;
Kuehne, G ;
Pichlmaier, A ;
Siodmok, A ;
Szelc, A ;
Tanner, L .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 2005, 110 (03) :173-178
[6]   Characterization of large-area photomultipliers under low magnetic fields: Design and performance of the magnetic shielding for the Double Chooz neutrino experiment [J].
Calvo, E. ;
Cerrada, M. ;
Fernandez-Bedoya, C. ;
Gil-Botella, I. ;
Palomares, C. ;
Rodriguez, I. ;
Toral, F. ;
Verdugo, A. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3) :222-230
[7]  
COHEN D, 2002, BIOM 2002 P 13 INT C
[8]   Principle and Performance of a Dual-Band Search Coil Magnetometer: A New Instrument to Investigate Fluctuating Magnetic Fields in Space [J].
Coillot, C. ;
Moutoussamy, J. ;
Lebourgeois, R. ;
Ruocco, S. ;
Chanteur, G. .
IEEE SENSORS JOURNAL, 2010, 10 (02) :255-260
[9]   Dynamic Shielding in Biomagnetism [J].
Costandache, D. ;
Banarescu, A. ;
Baltag, O. ;
Rau, I. ;
Rau, M. ;
Ojica, S. .
INTERNATIONAL CONFERENCE ON ADVANCEMENTS OF MEDICINE AND HEALTH CARE THROUGH TECHNOLOGY, 2009, 26 :121-124
[10]  
DAVID V, 1999, MASURARI BIOMEDICINA