Developmental validation of a point-of-care, salivary α-amylase biosensor

被引:64
作者
Shetty, Vivek [1 ]
Zigler, Corwin [1 ]
Robles, Theodore F. [2 ]
Elashoff, David [3 ]
Yamaguchi, Masaki [4 ]
机构
[1] Univ Calif Los Angeles, Sch Dent, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Psychol, Los Angeles, CA 90095 USA
[3] Univ Calif Los Angeles, Dept Med, Los Angeles, CA 90095 USA
[4] Iwate Univ, Grad Sch Engn, Morioka, Iwate 0208551, Japan
关键词
Salivary diagnostics; Stress biomarker; Salivary alpha amyase; Biosensor; Point-of-care measurement;
D O I
10.1016/j.psyneuen.2010.07.008
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The translation of salivary alpha-amylase (sAA) to the ambulatory assessment of stress hinges on the development of technologies capable of speedy and accurate reporting of sAA levels. Here, we describe the developmental validation and usability testing of a point-of-care, calorimetric, sAA biosensor. A disposable test strip allows for streamlined sample collection and a corresponding hand-held reader with integrated analytic capabilities permits rapid analysis and reporting of sAA levels. Bioanalytical validation utilizing saliva samples from 20 normal subjects indicates that, within the biosensor's linear range (10-230 U/ml), its accuracy (R(2) = 0.989), precision (CV < 9%), and measurement repeatability (range -3.1% to +3.1%) approach more elaborate laboratory-based, clinical analyzers. The truncated sampling-reporting cycle (<1 min) and the excellent performance characteristics of the biosensor has the potential to take sAA analysis out of the realm of dedicated, centralized laboratories and facilitate future sAA biomarker qualification studies. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:193 / 199
页数:7
相关论文
共 10 条
[1]   Salivary alpha-amylase as a measure of endogenous adrenergic activity [J].
Chatterton, RT ;
Vogelsong, KM ;
Lu, YC ;
Ellman, AB ;
Hudgens, GA .
CLINICAL PHYSIOLOGY, 1996, 16 (04) :433-448
[2]   Methodological considerations in the use of salivary α-amylase as a stress marker in field research [J].
DeCaro, Jason A. .
AMERICAN JOURNAL OF HUMAN BIOLOGY, 2008, 20 (05) :617-619
[3]   Salivary α-amylase in biobehavioral research -: Recent developments and applications [J].
Granger, Douglas A. ;
Kivlighan, Katie T. ;
el-Sheikh, Mona ;
Gordis, Elana B. ;
Stroud, Laura R. .
ORAL-BASED DIAGNOSTICS, 2007, 1098 :122-144
[4]   Clinical assessment applications of ambulatory biosensors [J].
Haynes, Stephen N. ;
Yoshioka, Dawn T. .
PSYCHOLOGICAL ASSESSMENT, 2007, 19 (01) :44-57
[5]   Salivary alpha-amylase as a non-invasive biomarker for the sympathetic nervous system: Current state of research [J].
Nater, U. M. ;
Rohleder, N. .
PSYCHONEUROENDOCRINOLOGY, 2009, 34 (04) :486-496
[6]   Human salivary alpha-amylase reactivity in a psychosocial stress paradigm [J].
Nater, UM ;
Rohleder, N ;
Gaab, J ;
Berger, S ;
Jud, A ;
Kirschbaum, C ;
Ehlert, U .
INTERNATIONAL JOURNAL OF PSYCHOPHYSIOLOGY, 2005, 55 (03) :333-342
[7]  
Robles T.F., BIOL PSYCHOL UNPUB
[8]   Psychosocial stress-induced activation of salivary alpha-amylase - An indicator of sympathetic activity? [J].
Rohleder, N ;
Nater, UM ;
Wolf, JM ;
Ehlert, U ;
Kirschbaum, C .
BIOBEHAVIORAL STRESS RESPONSE: PROTECTIVE AND DAMAGING EFFECTS, 2004, 1032 :258-263
[9]   Hand-held monitor of sympathetic nervous system using salivary amylase activity and its validation by driver fatigue assessment [J].
Yamaguchi, M ;
Deguchi, M ;
Wakasugi, J ;
Ono, S ;
Takai, N ;
Higashi, T ;
Mizuno, Y .
BIOSENSORS & BIOELECTRONICS, 2006, 21 (07) :1007-1014
[10]   Proposal of blood-collecting needle approach to semi-invasive method [J].
Yamaguchi, M ;
Kawabata, Y ;
Yamazaki, K ;
Kobayashi, M ;
Ito, T .
DIABETES RESEARCH AND CLINICAL PRACTICE, 2004, 66 :S179-S183