Microfluidic heavy metal immunoassay based on absorbance measurement

被引:23
作者
Date, Yasumoto [1 ,2 ]
Terakado, Shingo [1 ]
Sasaki, Kazuhiro [1 ]
Aota, Arata [1 ]
Matsumoto, Norio [1 ,2 ]
Shiku, Hitoshi [2 ,3 ]
Ino, Kosuke [1 ]
Watanabe, Yoshitomo [1 ,2 ]
Matsue, Tomokazu [2 ,3 ]
Ohmura, Naoya [1 ]
机构
[1] Cent Res Inst Elect Power Ind, Environm Sci Res Lab, Abiko, Chiba 2701194, Japan
[2] Tohoku Univ, Grad Sch Environm Studies, Sendai, Miyagi 9808579, Japan
[3] Tohoku Univ, World Premier Int Res Ctr Initiat, Adv Inst Mat Res WPI AIMR, Sendai, Miyagi 9808577, Japan
关键词
Immunoassay; Microfluidic device; Biosensor; Heavy metal; Absorbance measurement; Theoretical detection; MONOCLONAL-ANTIBODY; GOLD NANOPARTICLES; CADMIUM; IMMUNOCHROMATOGRAPHY; SYSTEMS; IONS; SPECTROMETRY; EXTRACTION; COMPLEXES; CHROMIUM;
D O I
10.1016/j.bios.2011.12.030
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A simple and rapid flow-based multioperation immunoassay for heavy metals using a microfluidic device was developed. The antigen-immobilized microparticles in a sub-channel were introduced as the solid phase into a main channel structures through a channel flow mechanism and packed into a detection area enclosed by dam-like structures in the microfluidic device. A mixture of a heavy metal and a gold nanoparticle-labeled antibody was made to flow toward the corresponding metal through the main channel and make brief contact with the solid phase. A small portion of the free antibody was captured and accumulated on the packed solid phase. The measured absorbance of the gold label was proportional to the free antibody portion and, thus, to the metal concentration. Each of the monoclonal antibodies specific for cadmium-EDTA, chromium-EDTA, or lead-DTPA was applied to the single-channel microfluidic device. Under optimized conditions of flow rate, volume, and antibody concentration, the theoretical (antibody K-d-limited) detection levels of the three heavy metal species were achieved within only 7 min. The dynamic range for cadmium, chromium, and lead was 0.57-60.06 ppb, 0.03-0.97 ppb, and 0.04-5.28 ppb, respectively. An integrated microchannel device for simultaneous multiflow was also successfully developed and evaluated. The multiplex cadmium immunoassay of four samples was completed within 8 min for a dynamic range of 0.42-37.48 ppb. Present microfluidic heavy metal immunoassays satisfied the Japanese environmental standard for cadmium, chromium and, lead, which provided in the soil contamination countermeasures act. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:106 / 112
页数:7
相关论文
共 32 条
[1]   Immunoassays for metal ions [J].
Blake, DA ;
Blake, RC ;
Khosraviani, M ;
Pavlov, AR .
ANALYTICA CHIMICA ACTA, 1998, 376 (01) :13-19
[2]   Metal binding properties of a monoclonal antibody directed toward metal-chelate complexes [J].
Blake, DA ;
Chakrabarti, P ;
Khosraviani, M ;
Hatcher, FM ;
Westhoff, CM ;
Goebel, P ;
Wylie, DE ;
Blake, RC .
JOURNAL OF BIOLOGICAL CHEMISTRY, 1996, 271 (44) :27677-27685
[3]   Antibody-based sensors for heavy metal ions [J].
Blake, DA ;
Jones, RM ;
Blake, RC ;
Pavlov, AR ;
Darwish, IA ;
Yu, HN .
BIOSENSORS & BIOELECTRONICS, 2001, 16 (9-12) :799-809
[4]   ENZYME-IMMUNOASSAY TO DETERMINE HEAVY-METALS USING ANTIBODIES TO SPECIFIC METAL-EDTA COMPLEXES - OPTIMIZATION AND VALIDATION OF AN IMMUNOASSAY FOR SOLUBLE INDIUM [J].
CHAKRABARTI, P ;
HATCHER, FM ;
BLAKE, RC ;
LADD, PA ;
BLAKE, DA .
ANALYTICAL BIOCHEMISTRY, 1994, 217 (01) :70-75
[5]   Recent progress in the application of microfluidic systems and gold nanoparticles in immunoassays [J].
Chen WenWen ;
Li TangSong ;
He Sha ;
Liu DingBin ;
Wang Zhuo ;
Zhang Wei ;
Jiang XingYu .
SCIENCE CHINA-CHEMISTRY, 2011, 54 (08) :1227-1232
[6]   Microfluidics-based diagnostics of infectious diseases in the developing world [J].
Chin, Curtis D. ;
Laksanasopin, Tassaneewan ;
Cheung, Yuk Kee ;
Steinmiller, David ;
Linder, Vincent ;
Parsa, Hesam ;
Wang, Jennifer ;
Moore, Hannah ;
Rouse, Robert ;
Umviligihozo, Gisele ;
Karita, Etienne ;
Mwambarangwe, Lambert ;
Braunstein, Sarah L. ;
van de Wijgert, Janneke ;
Sahabo, Ruben ;
Justman, Jessica E. ;
El-Sadr, Wafaa ;
Sia, Samuel K. .
NATURE MEDICINE, 2011, 17 (08) :1015-U138
[7]   Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) [J].
Duffy, DC ;
McDonald, JC ;
Schueller, OJA ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4974-4984
[8]   Solid-phase extraction of Mn(II), Co(II), Ni(II), Cu(II), Cd(II) and Pb(II) ions from environmental samples by flame atomic absorption spectrometry (FAAS) [J].
Duran, Celal ;
Gundogdu, Ali ;
Bulut, Volkan Numan ;
Soylak, Mustafa ;
Elci, Latif ;
Senturk, Hasan Basri ;
Tufekci, Mehmet .
JOURNAL OF HAZARDOUS MATERIALS, 2007, 146 (1-2) :347-355
[9]   Toward one-step point-of-care immunodiagnostics using capillary-driven microfluidics and PDMS substrates [J].
Gervais, Luc ;
Delamarche, Emmanuel .
LAB ON A CHIP, 2009, 9 (23) :3330-3337
[10]   Micro OS-ELISA: Rapid noncompetitive detection of a small biomarker peptide by open-sandwich enzyme-linked immunosorbent assay (OS-ELISA) integrated into microfluidic device [J].
Ihara, Masaki ;
Yoshikawa, Amane ;
Wu, Yushu ;
Takahashi, Hiroko ;
Mawatari, Kazuma ;
Shimura, Kiyohito ;
Sato, Kae ;
Kitamori, Takehiko ;
Ueda, Hiroshi .
LAB ON A CHIP, 2010, 10 (01) :92-100