Towards High-Resolution X-Ray Imaging Using a Structured Scintillator

被引:9
作者
Hormozan, Yashar [1 ]
Yun, Sang-Ho [1 ]
Svenonius, Olof [2 ]
Linnros, Jan [1 ]
机构
[1] Royal Inst Technol KTH, SE-10044 Stockholm, Sweden
[2] Scint X AB, SE-16440 Stockholm, Sweden
关键词
Scintillation detectors; waveguides; X-ray imaging technologies; PERFORMANCE; DETECTOR; GUIDES; ARRAYS;
D O I
10.1109/TNS.2011.2177477
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Structured scintillators, by light-guiding secondary emitted visible photons to a pixel in a CCD or CMOS image sensor, improve the lateral resolution of X-ray imaging detectors. In this work we have fabricated pore arrays in a silicon wafer and subsequently filled them with CsI(Tl) by a melting process. The goal was to down-scale the pore geometry for increased resolution. The results show that although pore depth must be reduced to comply with achievable aspect ratio of the Inductively Coupled Plasma (ICP) etching, melting into the pores is possible. The time and temperature has, however, to be optimized to prevent thallium loss during the melting. By correlating light yield measurements with the X-ray absorption in samples of various geometries, we find that the efficiency is slightly reduced for pore diameters approaching one micron. Nevertheless, the increased absorption in deep pores will lead to a significantly improved quantum efficiency compared to thin films currently used to achieve the same lateral resolution.
引用
收藏
页码:19 / 23
页数:5
相关论文
共 11 条
  • [1] Performance of scintillating waveguides for CCD-based X-ray detectors
    Badel, X
    Norlin, B
    Kleimann, P
    Williams, L
    Moody, SJ
    Tyrrell, GC
    Fröjdh, C
    Linnros, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 3 - 8
  • [2] Metallized and oxidized silicon macropore arrays filled with a scintillator for CCD-based X-ray imaging detectors
    Badel, X
    Linnros, J
    Kleimann, P
    Norlin, B
    Koskiahde, E
    Valpas, K
    Nenonen, S
    Petersson, CS
    Fröjdh, C
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (03) : 1001 - 1005
  • [3] BADEL X, 2005, THESIS ROYAL I TECHN
  • [4] MICROFABRICATION OF CELLULAR PHOSPHORS
    DECKMAN, HW
    DUNSMUIR, JH
    GRUNER, SM
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1989, 7 (06): : 1832 - 1835
  • [5] An x-ray imaging pixel detector based on a scintillating guides screen
    Kleimann, P
    Linnros, J
    Fröjdh, C
    Petersson, CS
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (04) : 1483 - 1486
  • [6] Structured CsI(Tl) scintillators for X-ray imaging applications
    Nagarkar, VV
    Gupta, TK
    Miller, SR
    Klugerman, Y
    Squillante, MR
    Entine, G
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 492 - 496
  • [7] Development of a high-efficiency high-resolution imaging detector for 30-80 keV X-rays
    Olsen, U. L.
    Badel, X.
    Linnros, J.
    Di Michiel, M.
    Martin, T.
    Schmidt, S.
    Poulsen, H. F.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 576 (01) : 52 - 55
  • [8] Olsen U. L., 2009, THESIS TECHNICAL U L
  • [9] X-ray imaging performance of scintillator-filled silicon pore arrays
    Simon, Matthias
    Engel, Klaus Juergen
    Menser, Bernd
    Badel, Xavier
    Linnros, Jan
    [J]. MEDICAL PHYSICS, 2008, 35 (03) : 968 - 981
  • [10] Performance of an X-ray imaging detector based on a structured scintillator
    Svenonius, Olof
    Sahlholm, Anna
    Wiklund, Per
    Linnros, Jan
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 607 (01) : 138 - 140