Influence of stent length and heparin coating on platelet activation:: A flow cytometric analysis in a pulsed floating model

被引:32
作者
Beythien, C
Gutensohn, K
Bau, J
Hamm, CW
Kühnl, P
Meinertz, T
Terres, W
机构
[1] Univ Hamburg, Hosp Eppendorf, Med Clin, Dept Cardiol, D-20246 Hamburg, Germany
[2] Univ Hamburg, Hosp Eppendorf, Dept Transfus Med & Transplantat Immunol, D-20246 Hamburg, Germany
关键词
stents; platelets; glycoproteins; flow cytometry; heparin;
D O I
10.1016/S0049-3848(98)00198-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Platelets are involved in acute and subacute thrombotic occlusions of coronary stents and also may play a role in the pathophysiology of in-stent restenosis, This study sought to investigate the expression of activation dependent glycoproteins on platelets by flow cytometry and time until stent thrombosis in an in vitro model of stent thrombosis. Coronary stents were placed in parallel silicon tubings with circulating citrated platelet rich plasma to measure 1) influence of stent length on platelet antigens; 2) influence of heparin coating on platelet antigens; and 3) time until stent thrombosis. After recalcification aliquots of platelet-rich plasma were taken over 10 minutes in 2-minute intervals and immediately fixed and stabilized. For flow cytometric analysis monoclonal antibodies to CD41a (glycoprotein IIb/IIIa), CD42b (glycoprotein Ib-V-IX), CD62p (P-selectin), and CD63 (glycoprotein 53) were used. Within 2 minutes after start of circulation, the expression of CD62p and CD63 increased, Longer stents resulted in more platelet activation than shorter stents (25 mm vs. 15 mm; p<0.001. Time until stent thrombosis was reduced (25 mm vs. 15 mm; p<0.05). Heparin coating did not significantly influence flow cytometry detectable platelet activation but prolonged time until stent thrombosis (coated vs. uncoated; p<0.005). In control tubing systems without stents platelet activation was less pronounced (p<0.0001). Antibodies to CD41a and CD42b did not show significant changes. In this model platelet activation detected by flow cytometry and time until stent thrombosis were dependent on stent length and coating. In vitro testing could be useful to optimize stent design and material. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:79 / 86
页数:8
相关论文
共 37 条
[1]  
Aliabadi D., 1997, Journal of the American College of Cardiology, V29, p276A
[2]   IN-VITRO MODEL TO TEST THE THROMBOGENICITY OF CORONARY STENTS [J].
BEYTHIEN, C ;
TERRES, W ;
HAMM, CW .
THROMBOSIS RESEARCH, 1994, 75 (06) :581-590
[3]  
Chronos Nicolas A. F., 1996, Journal of the American College of Cardiology, V27, p84A
[4]   SUPPRESSION BY HEPARIN OF SMOOTH-MUSCLE CELL-PROLIFERATION IN INJURED ARTERIES [J].
CLOWES, AW ;
KARNOWSKY, MJ .
NATURE, 1977, 265 (5595) :625-626
[5]   INTRACORONARY STENTING WITHOUT ANTICOAGULATION ACCOMPLISHED WITH INTRAVASCULAR ULTRASOUND GUIDANCE [J].
COLOMBO, A ;
HALL, P ;
NAKAMURA, S ;
ALMAGOR, Y ;
MAIELLO, L ;
MARTINI, G ;
GAGLIONE, A ;
GOLDBERG, SL ;
TOBIS, JM .
CIRCULATION, 1995, 91 (06) :1676-1688
[6]  
CORASH L, 1990, BLOOD CELLS, V16, P97
[7]  
DSOUZA D, 1994, LANCET, V344, P991
[8]  
Fernandez-Aviles F., 1996, European Heart Journal, V17, P307
[9]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[10]   Platelet activation and coronary stent implantation - Effect of antithrombotic therapy [J].
Gawaz, M ;
Neumann, FJ ;
Ott, I ;
May, A ;
Schomig, A .
CIRCULATION, 1996, 94 (03) :279-285