Validation of a new noncontact catheter system for electroanatomic mapping of left ventricular endocardium

被引:136
作者
Gornick, CC
Adler, SW
Pederson, B
Hauck, J
Budd, J
Schweitzer, J
机构
[1] Univ Minnesota, Vet Affairs Med Ctr, Minneapolis, MN 55417 USA
[2] St Paul Heart Clin, St Paul, MN USA
[3] Endocardial Solut Inc, St Paul, MN USA
关键词
mapping; catheter ablation; arrhythmia; ventricles; electrophysiology;
D O I
10.1161/01.CIR.99.6.829
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Improvements in cardiac mapping are required to advance our understanding and treatment of arrhythmias. This study validated a new noncontact multielectrode array catheter and accompanying analysis system to provide electroanatomic mapping of the entire left ventricular (LV) endocardium during a single beat. Methods and Results-A 9F 64-electrode balloon array catheter with an inflated size of 1.8x4.6 cm was used to simultaneously record electrical potentials generated by the heart and locate a standard electrophysiology (EP) catheter within the same chamber. By use of the recorded location of the EP-catheter tip, LV geometry was determined. Array potentials served as inputs to a high-order boundary-element method to produce 3360 potential points on the endocardial surface translatable into electrograms or color-coded activation maps. Three methods of validation were used: (1) driven electrodes in an in vitro tank were located; (2) waveforms generated from the array catheter were compared with catheter contact waveforms in canine LV; and (3) sites of local LV endocardial activation were located and marked with radiofrequency lesions. Tank testing located a driven electrode to within 2.33+/-0.44 mm, Correlation of timing and morphology of computed versus contact electrograms was 0.966, Radiofrequency lesions marked 17 endocardial pacing sites to within 4.0+/-3.2 mm. Conclusions-This new system provides anatomically accurate endocardial isopotential mapping during a single cardiac cycle. The locator component enabled placement of a separate EP catheter to any site within the mapped chamber.
引用
收藏
页码:829 / 835
页数:7
相关论文
共 8 条
[1]   SIMULTANEOUS 60-ELECTRODE MAPPING OF VENTRICULAR-TACHYCARDIA USING PERCUTANEOUS CATHETERS [J].
DAVIS, LM ;
COOPER, M ;
JOHNSON, DC ;
UTHER, JB ;
RICHARDS, DAB ;
ROSS, DL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (03) :709-719
[2]   Percutaneous multielectrode endocardial mapping during ventricular tachycardia in the swine model [J].
Eldar, M ;
Fitzpatrick, AP ;
Ohad, D ;
Smith, MF ;
Hsu, S ;
Whayne, JG ;
Vered, Z ;
Rotstein, Z ;
Kordis, T ;
Swanson, DK ;
Chin, M ;
Scheinman, MM ;
Lesh, MD ;
Greenspon, AJ .
CIRCULATION, 1996, 94 (05) :1125-1130
[3]  
Gepstein L, 1997, CIRCULATION, V95, P1611
[4]   Successful radiofrequency catheter ablation of sustained ventricular tachycardia postmyocardial infarction in man guided by a multielectrode ''basket'' catheter [J].
Greenspon, AJ ;
Hsu, SS ;
Datorre, S .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (05) :565-570
[5]   RECURRENT SUSTAINED VENTRICULAR TACHYCARDIA .2. ENDOCARDIAL MAPPING [J].
JOSEPHSON, ME ;
HOROWITZ, LN ;
FARSHIDI, A ;
SPEAR, JF ;
KASTOR, JA ;
MOORE, EN .
CIRCULATION, 1978, 57 (03) :440-447
[6]  
Khoury DS, 1998, CIRCULATION, V97, P399
[7]   RECONSTRUCTION OF ENDOCARDIAL POTENTIALS AND ACTIVATION SEQUENCES FROM INTRACAVITARY PROBE MEASUREMENTS - LOCALIZATION OF PACING SITES AND EFFECTS OF MYOCARDIAL STRUCTURE [J].
KHOURY, DS ;
TACCARDI, B ;
LUX, RL ;
ERSHLER, PR ;
RUDY, Y .
CIRCULATION, 1995, 91 (03) :845-863
[8]   A NEW INTRACAVITARY PROBE FOR DETECTING THE SITE OF ORIGIN OF ECTOPIC VENTRICULAR BEATS DURING ONE CARDIAC CYCLE [J].
TACCARDI, B ;
ARISI, G ;
MACCHI, E ;
BARUFFI, S ;
SPAGGIARI, S .
CIRCULATION, 1987, 75 (01) :272-281