Repeatability of Foveal Measurements Using Spectralis Optical Coherence Tomography Segmentation Software

被引:96
作者
Ctori, Irene [1 ]
Huntjens, Byki [1 ]
机构
[1] City Univ London, Henry Wellcome Labs Vis Sci, Appl Vis Res Ctr, London EC1V 0HB, England
关键词
RETINAL THICKNESS MEASUREMENTS; MACULAR THICKNESS; TIME-DOMAIN; MULTIPLE-SCLEROSIS; LAYER THICKNESS; REPRODUCIBILITY; OCT; HEALTHY; PIGMENT; MORPHOLOGY;
D O I
10.1371/journal.pone.0129005
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To investigate repeatability and reproducibility of thickness of eight individual retinal layers at axial and lateral foveal locations, as well as foveal width, measured from Spectralis spectral domain optical coherence tomography (SD-OCT) scans using newly available retinal layer segmentation software. Methods High-resolution SD-OCT scans were acquired for 40 eyes of 40 young healthy volunteers. Two scans were obtained in a single visit for each participant. Using new Spectralis segmentation software, two investigators independently obtained thickness of each of eight individual retinal layers at 0 degrees, 2 degrees and 5 degrees eccentricities nasal and temporal to foveal centre, as well as foveal width measurements. Bland-Altman Coefficient of Repeatability (CoR) was calculated for inter-investigator and inter-scan agreement of all retinal measurements. Spearman's. indicated correlation of manually located central retinal thickness (RT0) with automated minimum foveal thickness (MFT) measurements. In addition, we investigated nasal-temporal symmetry of individual retinal layer thickness within the foveal pit. Results Inter-scan CoR values ranged from 3.1 mu m for axial retinal nerve fibre layer thickness to 15.0 mu m for the ganglion cell layer at 5 degrees eccentricity. Mean foveal width was 2550 mu m +/- 322 mu m with a CoR of 13 mu m for inter-investigator and 40 mu m for inter-scan agreement. Correlation of RT0 and MFT was very good (rho = 0.97, P < 0.0005). There were no significant differences in thickness of any individual retinal layers at 2 degrees nasal compared to temporal to fovea (P > 0.05); however this symmetry could not be found at 5 degrees eccentricity. Conclusions We demonstrate excellent repeatability and reproducibility of each of eight individual retinal layer thickness measurements within the fovea as well as foveal width using Spectralis SDOCT segmentation software in a young, healthy cohort. Thickness of all individual retinal layers were symmetrical at 2 degrees, but not at 5 degrees eccentricity away from the fovea.
引用
收藏
页数:14
相关论文
共 56 条
[1]   Ultrahigh resolution optical coherence tomography of the monkey fovea. Identification of retinal sublayers by correlation with semithin histology sections [J].
Anger, EM ;
Unterhuber, A ;
Hermann, B ;
Sattmann, H ;
Schubert, C ;
Morgan, JE ;
Cowey, A ;
Ahnelt, PK ;
Drexler, W .
EXPERIMENTAL EYE RESEARCH, 2004, 78 (06) :1117-1125
[2]  
[Anonymous], 2013, SPECTR HRA OCT US MA
[3]   Effect of image quality on tissue thickness measurements obtained with spectral domain-optical coherence tomography [J].
Balasubramanian, Madhusudhanan ;
Bowd, Christopher ;
Vizzeri, Gianmarco ;
Weinreb, Robert N. ;
Zangwill, Linda M. .
OPTICS EXPRESS, 2009, 17 (05) :4019-4036
[4]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   Reproducibility of Spectral-Domain Optical Coherence Tomography Retinal Thickness Measurements and Conversion to Equivalent Time-Domain Metrics in Diabetic Macular Edema [J].
Bressler, Susan B. ;
Edwards, Allison R. ;
Chalam, Kakarla V. ;
Bressler, Neil M. ;
Glassman, Adam R. ;
Jaffe, Glenn J. ;
Melia, Michele ;
Saggau, David D. ;
Plous, Oren Z. .
JAMA OPHTHALMOLOGY, 2014, 132 (09) :1113-1122
[7]   Correlation in foveal thickness measurements between spectral-domain and time-domain optical coherence tomography in normal individuals [J].
Carpineto, P. ;
Nubile, M. ;
Toto, L. ;
Gnama, A. Aharrh ;
Marcucci, L. ;
Mastropasqua, L. ;
Ciancaglini, M. .
EYE, 2010, 24 (02) :251-258
[8]   The Royal College of Ophthalmologists Guidelines on AMD: Executive Summary [J].
Chakravarthy, U. ;
Williams, M. .
EYE, 2013, 27 (12) :1429-1431
[9]   Validated Automatic Segmentation of AMD Pathology Including Drusen and Geographic Atrophy in SD-OCT Images [J].
Chiu, Stephanie J. ;
Izatt, Joseph A. ;
O'Connell, Rachelle V. ;
Winter, Katrina P. ;
Toth, Cynthia A. ;
Farsiu, Sina .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (01) :53-61
[10]   Repeatability of Spectralis OCT Measurements of Macular Thickness and Volume in Diabetic Macular Edema [J].
Comyn, Oliver ;
Heng, Ling Zhi ;
Ikeji, Felicia ;
Bibi, Kanom ;
Hykin, Philip G. ;
Bainbridge, James W. ;
Patel, Praveen J. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (12) :7754-7759