MRI and dental implantology: Two which do not exclude each other

被引:40
作者
Korn, P. [1 ]
Elschner, C. [2 ]
Schulz, M. C. [1 ]
Range, U. [3 ]
Mai, R. [1 ]
Scheler, U. [2 ]
机构
[1] Univ Klinikum Carl Gustav Carus Dresden, Klin & Poliklin Mund Kiefer & Gesichtschirurg, D-01307 Dresden, Germany
[2] Leibniz Inst Polymerforsch Dresden eV, D-01069 Dresden, Germany
[3] Tech Univ Dresden, Med Fak Carl Gustav Carus, Inst Med Informat & Biometrie, D-01307 Dresden, Germany
关键词
Dental implant; Polyetheretherketone; Coating; Osseointegration; Histology; Magnetic resonance imaging; TITANIUM IMPLANTS; BONE; OSSEOINTEGRATION; PEEK; SOFT; BIOMATERIALS; DENTISTRY; ACCURACY; IMPROVE; TISSUE;
D O I
10.1016/j.biomaterials.2015.02.114
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Currently, histological techniques are used to analyse implant-tissue-interactions. However, these methods are destructive and time-consuming. Furthermore, they require a large number of animals as longitudinal observations in one individual are not possible. The evaluation by non-destructive imaging techniques provides the opportunity to study the osseous integration with a reduced number of animals and a decreased biological variability. The present study examined the suitability of magnetic resonance imaging (MRI) to assess peri-implant bone formation exemplarily for a dental implant in a minipig model. Due to its compatibility to MR imaging polyetheretherketone (PEEK) coated with a thin layer of titanium was applied as implant material. Osseointegration was analysed within different pea-implant regions quantifying bone volume density and soft tissue content, which were assessed by MRI and histology, likewise. It could be proven that the examined regions showed differences in bone formation; the region adjacent to the implant apex turned out to be the most dynamic. Both methods led to comparable results with no significant differences regarding to the assessed parameters. Moreover, it was demonstrated that titanium coated PEEK showed a sufficient osseointegration and MRI provides a promising application in monitoring bone formation. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:634 / 645
页数:12
相关论文
共 47 条
[1]   Accuracy of magnetic resonance imaging compared with computed tomography for implant planning [J].
Aguiar, Marcelo F. ;
Marques, Alexandre P. ;
Carvalho, Antonio Carlos P. ;
Cavalcanti, Marcelo G. P. .
CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (04) :362-365
[2]   MR Image Based Approach for Metal Artifact Reduction in X-Ray CT [J].
Anderla, Andras ;
Culibrk, Dubravko ;
Delso, Gaspar ;
Mirkovic, Milan .
SCIENTIFIC WORLD JOURNAL, 2013,
[3]   Are ceramic implants a viable alternative to titanium implants? A systematic literature review [J].
Andreiotelli, Marina ;
Wenz, Hans J. ;
Kohal, Ralf-Joachim .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 :32-47
[4]   Imaging challenges in biomaterials and tissue engineering [J].
Appel, Alyssa A. ;
Anastasio, Mark A. ;
Larson, Jeffery C. ;
Brey, Eric M. .
BIOMATERIALS, 2013, 34 (28) :6615-6630
[5]  
Bernhardt R, 2004, Eur Cell Mater, V7, P42
[6]   COMPARISON OF BONE-IMPLANT CONTACT AND BONE-IMPLANT VOLUME BETWEEN 2D-HISTOLOGICAL SECTIONS AND 3D-SRμCT SLICES [J].
Bernhardt, Ricardo ;
Kuhlisch, Eberhard ;
Schulz, Matthias C. ;
Eckelt, Uwe ;
Stadlinger, Bernd .
EUROPEAN CELLS & MATERIALS, 2012, 23 :237-248
[7]   Current concepts in maxillofacial imaging [J].
Boeddinghaus, Rudolf ;
Whyte, Andy .
EUROPEAN JOURNAL OF RADIOLOGY, 2008, 66 (03) :396-418
[8]   OSSEOINTEGRATION AND ITS EXPERIMENTAL BACKGROUND [J].
BRANEMARK, PI .
JOURNAL OF PROSTHETIC DENTISTRY, 1983, 50 (03) :399-410
[9]   Systematic review with meta-analysis: magnetic resonance enterography signs for the detection of inflammation and intestinal damage in Crohn's disease [J].
Church, P. C. ;
Turner, D. ;
Feldman, B. M. ;
Walters, T. D. ;
Greer, M. -L. ;
Amitai, M. M. ;
Griffiths, A. M. .
ALIMENTARY PHARMACOLOGY & THERAPEUTICS, 2015, 41 (02) :153-166
[10]  
Cook S D, 1995, J Oral Implantol, V21, P176