Electrostatic microactuators for precise positioning of neural microelectrodes

被引:44
作者
Muthuswamy, J
Okandan, M
Jain, T
Gilletti, A
机构
[1] Arizona State Univ, Dept Bioengn, Tempe, AZ 85287 USA
[2] Sandia Natl Labs, MEMS Sci & Technol, Albuquerque, NM 87185 USA
[3] Arizona State Univ, Harrigton Dept Bioengn, Tempe, AZ 85287 USA
关键词
brain implants; MEMS; microdrive; multi-unit activity; prostheses;
D O I
10.1109/TBME.2005.855712
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Microelectrode arrays used for monitoring single and multineuronal action potentials often fail to record from the same population of neurons over a period of time likely due to micro-motion of neurons away from the microelectrode, gliosis around the recording site and also brain movement due to behavior. We report here novel electrostatic microactuated microelectrodes that will enable precise repositioning of the microelectrodes within the brain tissue. Electrostatic comb-drive microactuators and associated microelectrodes are fabricated using the SUMMiT V(TM) (Sandia's Ultraplanar Multilevel MEMS Technology) process, a five-layer polysilicon micromachining technology of the Sandia National labs, NM. The microfabricated microactuators; enable precise bidirectional positioning of the microelectrodes in the brain with accuracy in the order of 1 mu m. The microactuators allow for a linear translation of the microelectrodes of up to 5 mm in either direction making it suitable for positioning microelectrodes in deep structures of a rodent brain. The overall translation was reduced to approximately 2 mm after insulation of the microelectrodes with epoxy for monitoring multiunit activity. The microactuators; are capable of driving the microelectrodes in the brain tissue with forces in the order of several micro-Newtons. Single unit recordings were obtained from the somatosensory cortex of adult rats in acute experiments demonstrating the feasibility of this technology. Further optimization of the insulation, packaging and interconnect issues will be necessary before this technology can be validated in long-term experiments.
引用
收藏
页码:1748 / 1755
页数:8
相关论文
共 28 条
[1]   SOLID-STATE ELECTRODES FOR MULTICHANNEL MULTIPLEXED INTRACORTICAL NEURONAL RECORDING [J].
BEMENT, SL ;
WISE, KD ;
ANDERSON, DJ ;
NAJAFI, K ;
DRAKE, KL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :230-241
[2]   Ceramic-based multisite microelectrodes for electrochemical recordings [J].
Burmeister, JJ ;
Moxon, K ;
Gerhardt, GA .
ANALYTICAL CHEMISTRY, 2000, 72 (01) :187-192
[3]   Semi-chronic motorized microdrive and control algorithm for autonomously isolating and maintaining optimal extracellular action potentials [J].
Cham, JG ;
Branchaud, EA ;
Nenadic, Z ;
Greger, B ;
Andersen, RA ;
Burdick, JW .
JOURNAL OF NEUROPHYSIOLOGY, 2005, 93 (01) :570-579
[4]  
CHARLES HK, 1984, POLYIMIDES, V2, P1139
[5]   A NEW METHOD FOR THE INSERTION OF MULTIPLE MICROPROBES INTO NEURAL AND MUSCULAR TISSUE, INCLUDING FIBER ELECTRODES, FINE WIRES, NEEDLES AND MICROSENSORS [J].
ECKHORN, R ;
THOMAS, U .
JOURNAL OF NEUROSCIENCE METHODS, 1993, 49 (03) :175-179
[6]  
EICHENBAUM H, 1986, J ELECTROPHYSIOL TEC, V13, P189
[7]   Miniature motorized microdrive and commutator system for chronic neural recording in small animals [J].
Fee, MS ;
Leonardo, A .
JOURNAL OF NEUROSCIENCE METHODS, 2001, 112 (02) :83-94
[8]   SURFACE MICROMACHINED MICROENGINE [J].
GARCIA, EJ ;
SNIEGOWSKI, JJ .
SENSORS AND ACTUATORS A-PHYSICAL, 1995, 48 (03) :203-214
[9]  
Jacobsen S. C., 1989, Proceedings: IEEE Micro Electro Mechanical Systems. An Investigation of Micro Structures, Sensors, Actuators, Machines and Robots (IEEE Cat. No.89THO249-3), P17, DOI 10.1109/MEMSYS.1989.77953
[10]   Silicon-substrate intracortical microelectrode arrays for long-term recording of neuronal spike activity in cerebral cortex [J].
Kipke, DR ;
Vetter, RJ ;
Williams, JC ;
Hetke, JF .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2003, 11 (02) :151-155