A study of the moving rate of positive results for use in a patient-based real-time quality control program on a procalcitonin point-of-care testing analyzer

被引:3
作者
He, Yili [1 ]
Gu, Daqing [1 ]
Kong, Xiangzhi [1 ]
Feng, Zhiqiang [1 ]
Lin, Weishang [1 ]
Cai, Yunfeng [1 ]
机构
[1] Fifth Peoples Hosp Panyu, Clin Lab, Guangzhou 511400, Peoples R China
关键词
internal quality control; moving rate of positive results; point-of-care testing; procalcitonin; AVERAGE; RISK; SUM;
D O I
10.1002/jcla.24320
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective To establish an applicable and highly sensitive patient-based real-time quality control (PBRTQC) program based on a data model constructed with patients' results of a procalcitonin point-of-care testing (POCT) analyzer. Methods Patients' results were retrospectively collected within one year. The Excel software was used to establish quality control (QC) programs of the moving average (MA) and the moving rate of positive results (MR). A Monte Carlo simulation was used to introduce positive and negative biases between 0.01 and 1 ng/ml at random points of the testing data set. Different parameters were used to detect the biases, and the detection efficiency was expressed using the median number of patient samples affected until error detection (MNPed). After comparing the MNPeds of different programs, MA and MR programs with appropriate parameters were selected, and validation plots were generated using MNPeds and maximum number of the patient samples affected (MAX). beta curves were generated using the power function of the programs, the performances were compared with that of the conventional QC program. Results Neither the conventional QC nor MA program was sensitive to small bias, While MR program can detect the minimum positive bias of 0.06 ng/ml and negative of 0.4 ng/ml at an average daily run size of 10 specimens, with FRs < 1.0%, beta s < 1%. Conclusion The MR program, which is more sensitive to small biases than conventional QC and MA programs, with low FR and beta. As such, it can be used as a PBRTQC program with high performance.
引用
收藏
页数:12
相关论文
共 25 条
[1]   Patient-Based Real-Time Quality Control: Review and Recommendations [J].
Badrick, Tony ;
Bietenbeck, Andreas ;
Cervinski, Mark A. ;
Katayev, Alex ;
van Rossum, Huub H. ;
Loh, Tze Ping .
CLINICAL CHEMISTRY, 2019, 65 (08) :962-971
[2]   A primer on patient-based quality control techniques [J].
Badrick, Tony ;
Cervinski, Mark ;
Loh, Tze Ping .
CLINICAL BIOCHEMISTRY, 2019, 64 :1-5
[3]   Understanding Patient-Based Real-Time Quality Control Using Simulation Modeling [J].
Bietenbeck, Andreas ;
Cervinski, Mark A. ;
Katayev, Alex ;
Loh, Tze Ping ;
van Rossum, Huub H. ;
Badrick, Tony .
CLINICAL CHEMISTRY, 2020, 66 (08) :1072-1083
[4]   ASSESSMENT OF AVERAGE OF NORMALS QUALITY-CONTROL PROCEDURES AND GUIDELINES FOR IMPLEMENTATION [J].
CEMBROWSKI, GS ;
CHANDLER, EP ;
WESTGARD, JO .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1984, 81 (04) :492-499
[5]   Plan for Quality to Improve Patient Safety at the Point of Care [J].
Ehrmeyer, Sharon S. .
ANNALS OF SAUDI MEDICINE, 2011, 31 (04) :342-346
[6]  
Huang W-C., 2021, ERJ OPEN RES, V7, P1
[7]   Sample size and rejection limits for detecting reagent lot variability: analysis of the applicability of the Clinical and Laboratory Standards Institute (CLSI) EP26-A protocol to real-world clinical chemistry data [J].
Kim, Sollip ;
Chang, Jeonghyun ;
Kim, Soo-Kyung ;
Park, Sholhui ;
Huh, Jungwon ;
Jeong, Tae-Dong .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2021, 59 (01) :127-138
[8]   Application of procalcitonin, white blood cell count and neutrophil-to-lymphocyte ratio in the diagnosis of systemic lupus erythematosus with a bacterial infection [J].
Li, Zhihong ;
Xiao, Yan ;
Zhang, Lina .
ANNALS OF PALLIATIVE MEDICINE, 2020, 9 (06) :3870-3876
[9]   Patient-based quality control for glucometers: using the moving sum of positive patient results and moving average [J].
Lim, Chun Yee ;
Badrick, Tony ;
Loh, Tze Ping .
BIOCHEMIA MEDICA, 2020, 30 (02) :1-11
[10]   Moving standard deviation and moving sum of outliers as quality tools for monitoring analytical precision [J].
Liu, Jiakai ;
Tan, Chin Hon ;
Badrick, Tony ;
Loh, Tze Ping .
CLINICAL BIOCHEMISTRY, 2018, 52 :112-116