Image Quality in Multidetector CT of Paranasal Sinuses: Potential of Dose Reduction Using an Adaptive Post-Processing Filter

被引:12
作者
Kropil, P. [1 ]
Cohnen, M. [2 ]
Andersen, K. [2 ]
Heinen, W. [1 ]
Stegmann, V. [1 ]
Modder, U. [1 ]
机构
[1] Univ Klinikum Dusseldorf, Inst Radiol, D-40225 Dusseldorf, Germany
[2] Stadt Kliniken Neuss Lukaskrankenhaus GmbH, Inst Klin Radiol, Neuss, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2010年 / 182卷 / 11期
关键词
post processing; filter; dose reduction; image quality; CT; paranasal sinuses; COMPUTED-TOMOGRAPHY; OPTIMIZATION; LESIONS;
D O I
10.1055/s-0029-1245586
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Evaluation of subjective image quality in dose-reduced multi-detector CT (MDCT) of paranasal sinuses using a 2D non-linear adaptive post-processing filter (2D-NLAF) Materials and Methods MDCT of paranasal sinuses was simulated using a human head phantom at a Somatom Sensation Cardiac 64 (Siemens Eilangen) At constant collimation (64 x 0 6 mm) und pitch (p = 1) the tube current (50 100 200 mAs) and tube potential (80 100 120 kVp) were modified The radiation exposure was represented by CTDIvol Four Independent blinded radiologists evaluated the image quality of axial 2 mm images and coronal reformations concerning the assessment of fractures and soft tissue processes The subjective image quality of original and post-processed images using a 2D-NLAF (SharpViewCT (R) Sweden) was graded on a 5-point scale ( 1 excellent - 5 not adequate) and compared Results Compared to the protocol with the best image quality (120kVp/200 mAs) 2D-NLAF led to a significant improvement in the subjective image quality at 100 kVp/100 mAs (score 1 4 with filter versus 22 without) and 120 kVp/50 mAs ( 1 6 versus 20) (p < 0 03) particularly for high contrasts ( fractures p < 0 001) In soft tissue processes 2D-NLAF provided improved quality from 21 to 14 (p < 0 04) at 100 kVp/100 mAs Down to a CTDIvol of 8 mGy the image quality was rated good and down to 5 mGy diagnostic Conclusion The phantom study indicates a dose reduction potential in MDCT of paranasal sinuses up to 58% compared to a standard dose protocol using a 2D-NLAF without an essential loss of image quality 2D-NLAF is particularly effective at 100 kVp/100 mAs and 120 kVp/50 mAs
引用
收藏
页码:973 / 978
页数:6
相关论文
共 18 条
  • [1] ABULKASIM K, 2009, AM J OTOLARYNGO 1009
  • [2] Multidetector CT of the paranasal sinus: Potential for radiation dose reduction
    Brem, Matthias H.
    Zamani, Amir A.
    Riva, Roberto
    Zou, Kelly H.
    Rumboldt, Zoran
    Hennig, Friedrich F.
    Kikinis, Ron
    Norbash, Alexander M.
    Schoepf, U. Joseph
    [J]. RADIOLOGY, 2007, 243 (03) : 847 - 852
  • [3] *BUND STRAHL STRAH, 2009, BUND STRAHL STRAHL D
  • [4] Multislice spiral CT of the paranasal sinuses.
    Dammann, F
    Bode, A
    Heuschmid, M
    Kopp, A
    Georg, C
    Pereira, PL
    Claussen, CD
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (08): : 701 - 706
  • [5] Multislice helical CT (MSCT) for mid-facial trauma: Optimization of parameters for scanning and reconstruction
    Dammert, S
    Funke, M
    Merten, HA
    Obernauer, S
    Grabbe, E
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (07): : 874 - 879
  • [6] Computed tomography practice Germany - Results of a national survey in 1999 - Preface
    Galanski, M
    Nagel, HD
    Stamm, G
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2001, 173 (10): : R1 - R1
  • [7] Technical approaches to the optimisation of CT
    Kalender, Willi A.
    Buchenau, Stefanie
    Deak, Paul
    Kellermeier, Markus
    Langner, Oliver
    van Straten, Marcel
    Vollmar, Sabrina
    Wilharm, Sylvia
    [J]. PHYSICA MEDICA, 2008, 24 (02) : 71 - 79
  • [8] Detection and characterization of lesions on low-radiation-dose abdominal CT images postprocessed with noise reduction filters
    Kalra, MK
    Maher, MM
    Blake, MA
    Lucey, BC
    Karau, K
    Toth, TL
    Avinash, G
    Halpern, EF
    Saini, S
    [J]. RADIOLOGY, 2004, 232 (03) : 791 - 797
  • [9] Can noise reduction filters improve low-radiation-dose chest CT images? Pilot study
    Kalra, MK
    Wittram, C
    Maher, MM
    Sharma, A
    Avinash, GB
    Karau, K
    Toth, TL
    Halpern, E
    Saini, S
    Shepard, JA
    [J]. RADIOLOGY, 2003, 228 (01) : 257 - 264
  • [10] Dose Reduction and Image Quality in MDCT of the Upper Abdomen: Potential of an Adaptive Post-Processing Filter
    Kroepil, P.
    Lanzman, R. S.
    Walther, C.
    Roehlen, S.
    Godehardt, E.
    Moedder, U.
    Cohnen, M.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2010, 182 (03): : 248 - 253