Measurement of nasal nitric oxide by hand-held and stationary devices

被引:29
作者
Montella, Silvia [1 ]
Alving, Kjell [2 ]
Maniscalco, Mauro [3 ]
Sofia, Matteo [4 ]
De Stefano, Sara [1 ]
Raia, Valeria [1 ]
Santamaria, Francesca [1 ]
机构
[1] Univ Naples Federico II, Dept Pediat, I-80131 Naples, Italy
[2] Uppsala Univ, Dept Womens & Childrens Hlth, Uppsala, Sweden
[3] S Maria della Pieta Hosp, Sect Resp Med, Naples, Italy
[4] Univ Naples Federico II, Dept Resp Med, I-80131 Naples, Italy
关键词
Chemiluminescence; cystic fibrosis; electrochemical sensor; nasal nitric oxide; primary ciliary dyskinesia; PRIMARY CILIARY DYSKINESIA; CHILDREN;
D O I
10.1111/j.1365-2362.2011.02501.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Nasal nitric oxide (nNO) is assessed by nasal aspiration/insufflation via one nostril or by nasal silent exhalation through a facemask and is also measured during humming, a manoeuvre that results in increased nNO in the presence of a patent osteomeatal complex. Humming nNO peak is absent in primary ciliary dyskinesia (PCD) and in cystic fibrosis (CF). Hand-held devices are used successfully for exhaled or nNO analysis. No study compared nNO during silent and humming exhalation using hand-held and stationary analysers. Methods Thirty-eight subjects (14 PCD; 11 CF; 13 healthy individuals) measured nNO with a stationary and a hand-held analyser during silent and humming exhalations. Results No difference between nNO obtained from stationary or hand-held analyser during silent and humming exhalation was found (P > 0 05). Patients with PCD exhibited lower silent and humming nNO than CF or controls (P < 0 001). During both silent and humming exhalation, there was a significant correlation between nNO from the two analyzers both in the whole study population and within each group (r >= 0.7, P < 0 01). Bland-Altman plots confirmed this agreement. Using the hand-held device during humming, nNO values of 50, 81 and 21 ppb had sensitivity and specificity > 90% for discriminating PCD or CF from healthy subjects, and patients with PCD from patients with CF, respectively. Conclusions The hand-held device is as effective as the stationary analyzer for assessing nNO during silent and humming exhalation. Its wider use might result in an increased number of subjects suspected to have PCD.
引用
收藏
页码:1063 / 1070
页数:8
相关论文
共 29 条
  • [1] Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children
    Alving, K
    Janson, C
    Nordvall, L
    [J]. RESPIRATORY RESEARCH, 2006, 7 (1)
  • [2] ALVING K, 1993, EUR RESPIR J, V6, P1368
  • [3] ATS/ERS recommendations for standardized procedures for the online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide, 2005
    American Thoracic Society
    European Respiratory Society
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2005, 171 (08) : 912 - 930
  • [4] [Anonymous], EUR RESP J
  • [5] Nasal nitric oxide concentration in paranasal sinus inflammatory diseases
    Arnal, JF
    Flores, P
    Rami, J
    Murris-Espin, M
    Bremont, F
    Aguilla, MPI
    Serrano, E
    Didier, A
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1999, 13 (02) : 307 - 312
  • [6] Primary ciliary dyskinesia: a consensus statement on diagnostic and treatment approaches in children
    Barbato, A.
    Frischer, T.
    Kuehni, C. E.
    Snijders, D.
    Azevedo, I.
    Baktai, G.
    Bartoloni, L.
    Eber, E.
    Escribano, A.
    Haarman, E.
    Hesselmar, B.
    Hogg, C.
    Jorissen, M.
    Lucas, J.
    Nielsen, K. G.
    O'Callaghan, C.
    Omran, H.
    Pohunek, P.
    Strippoli, M-P. F.
    Bush, A.
    [J]. EUROPEAN RESPIRATORY JOURNAL, 2009, 34 (06) : 1264 - 1276
  • [7] Airway nitric oxide in asthmatic children and patients with cystic fibrosis
    Dotsch, J
    Demirakca, S
    Terbrack, HG
    Huls, G
    Rascher, W
    Kuhl, PG
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1996, 9 (12) : 2537 - 2540
  • [8] Novel hand-held device for exhaled nitric oxide-analysis in research and clinical applications
    Hemmingsson T.
    Linnarsson D.
    Gambert R.
    [J]. Journal of Clinical Monitoring and Computing, 2004, 18 (5-6) : 379 - 387
  • [9] Tutorial in biostatistics - Sample sizes for clinical trials with normal data
    Julious, SA
    [J]. STATISTICS IN MEDICINE, 2004, 23 (12) : 1921 - 1986
  • [10] Karadag B, 1999, EUR RESPIR J, V13, P1402, DOI 10.1183/09031936.99.13614069