Screening determination of pharmaceutical pollutants in different water matrices using dual-channel capillary electrophoresis coupled with contactless conductivity detection

被引:36
作者
Minh Duc Le [1 ]
Hong Anh Duong [1 ]
Manh Huy Nguyen [1 ]
Saiz, Jorge [2 ]
Hung Viet Pham [1 ]
Thanh Duc Mai [1 ]
机构
[1] Hanoi Univ Sci, Ctr Environm Technol & Sustainable Dev CETASD, Nguyen Trai St 334, Hanoi, Vietnam
[2] CSIC, Inst Gen Organ Chem IQOG, Calle Juan de la Cierva,3, Madrid 28006, Spain
关键词
Capacitively coupled contactless conductivity detection ((CD)-D-4); Capillary electrophoresis (CE); Dual-channel CE; Water analysis; Pharmaceutically active compounds (PhACs); Water samples; NONSTEROIDAL ANTIINFLAMMATORY DRUGS; PERSONAL CARE PRODUCTS; SAMPLE STACKING STRATEGIES; SOLID-PHASE EXTRACTION; LIQUID-CHROMATOGRAPHY; AQUATIC ENVIRONMENT; CONCURRENT DETERMINATION; ONLINE PRECONCENTRATION; DRINKING-WATER; UV DETECTION;
D O I
10.1016/j.talanta.2016.07.032
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, the employment of purpose-made dual-channel compact capillary electrophoresis (CE) instrument with capacitively coupled contactless conductivity detection ((CD)-D-4) as a simple and inexpensive solution for screening determination of various pharmaceutical pollutants frequently occurring in surface water and hospital wastewater in Hanoi, Vietnam is reported. Five negatively charged pharmaceutically active compounds, namely ibuprofen, diclofenac, bezafibrate, ketoprofen and mefenamic acid were determined using the first channel whereas three positively charged ones, namely diphenhydramine, metoprolol and atenolol were determined with the second channel of the CE-(CD)-D-4 instrument. Two different background electrolytes (BGEs) were used in these two CE channels independently. The best detection limits achieved were in the range of 0.2-0.8 mg/L without sample pre-concentration. Enrichment factors up to 200 were obtainable with the inclusion of a solid phase extraction step. Good agreement between results obtained from CE-(CD)-D-4 and those with the standard confirmation method (HPLC-DAD) was achieved, with correlation coefficients higher than 0.98. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:512 / 520
页数:9
相关论文
共 51 条
[1]   Determination of drug residues in water by the combination of liquid chromatography or capillary electrophoresis with electrospray mass spectrometry [J].
Ahrer, W ;
Scherwenk, E ;
Buchberger, W .
JOURNAL OF CHROMATOGRAPHY A, 2001, 910 (01) :69-78
[2]   Detection and quantification of residues and metabolites of medicinal products in environmental compartments, food commodities and workplaces. A review [J].
Bottoni, Paola ;
Caroli, Sergio .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2015, 106 :3-24
[3]  
Boxall ABA, 2012, ENV HLTH PERSPECT, V120, P9
[4]   Current approaches to trace analysis of pharmaceuticals and personal care products in the environment [J].
Buchberger, Wolfgang W. .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (04) :603-618
[5]   Pharmaceuticals, Personal Care Products and Endocrine Disrupting Agents in the Environment - A Review [J].
Caliman, Florentina Anca ;
Gavrilescu, Maria .
CLEAN-SOIL AIR WATER, 2009, 37 (4-5) :277-303
[6]   Simultaneous determination of caffeine, paracetamol, and ibuprofen in pharmaceutical formulations by high-performance liquid chromatography with UV detection and by capillary electrophoresis with conductivity detection [J].
Cunha, Rafael R. ;
Chaves, Sandro C. ;
Ribeiro, Michelle M. A. C. ;
Torres, Livia M. F. C. ;
Munoz, Rodrigo A. A. ;
Dos Santos, Wallans T. P. ;
Richter, Eduardo M. .
JOURNAL OF SEPARATION SCIENCE, 2015, 38 (10) :1657-1662
[7]   Simultaneous determination of diclofenac and its common counter-ions in less than 1 minute using capillary electrophoresis with contactless conductivity detection [J].
Cunha, Rafael R. ;
Gimenes, Denise T. ;
Munoz, Rodrigo A. A. ;
do Lago, Claudimir L. ;
Richter, Eduardo M. .
ELECTROPHORESIS, 2013, 34 (9-10) :1423-1428
[8]   Electrokinetic supercharging for on-line preconcentration of seven non-steroidal anti-inflammatory drugs in water samples [J].
Dawod, Mohamed ;
Breadmore, Michael C. ;
Guijt, Rosanne M. ;
Haddad, Paul R. .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1189 (1-2) :278-284
[9]   Determination of pharmaceuticals in drinking water by CD-modified MEKC: Separation optimization using experimental design [J].
Drover, Vincent J. ;
Bottaro, Christina S. .
JOURNAL OF SEPARATION SCIENCE, 2008, 31 (21) :3740-3748
[10]  
Ducey SB, 2010, ACS SYM SER, V1048, P345