Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010

被引:4
|
作者
Beime, Beate [1 ]
Deutsch, Cornelia [1 ]
Krueger, Ralf [1 ]
Wolf, Andreas [2 ]
Mueller, Peter [2 ]
Hammel, Gertrud [3 ]
Bramlage, Peter [1 ]
机构
[1] Inst Pharmacol & Prevent Med, Bahnhofstr 20, D-49661 Cloppenburg, Germany
[2] Muller & Sebastiani GmbH, Ottobrunn, Germany
[3] Estimate GmbH, Augsburg, Germany
关键词
Validation; Blood pressure measuring device; Children; ESH; Custo screen pediatric; RECOMMENDATIONS; ADOLESCENTS; CHILDREN; GUIDELINES; MANAGEMENT; COMMITTEE;
D O I
10.1007/s00431-017-2874-3
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The purpose of the study was to validate the ambulatory blood pressure monitoring (ABPM) device custo screen pediatric in children aged 3 to 12 years according to the International Protocol of the European Society of Hypertension (ESH-IP revision 2010). Thirty-three children were included and systolic and diastolic blood pressure measurements were performed according to the ESH-IP. The protocol was modified for children considering data from the German Health Interview and Examination Survey for Children and Adolescents (KIGGS). The custo screen pediatric met all the requirements of the ESH-IP. The mean difference between the test device and the reference was -1.4 +/- 3.0 mmHg for systolic blood pressure (SBP) and -0.7 +/- 3.2 mmHg for diastolic blood pressure (DBP). For SBP and DBP, all 99 measurements were within the absolute difference of 10 mmHg between the test device and the reference. As to part 2 of the protocol, for DBP in all subjects, two out of three measurements were within 5 mmHg between the device and the standard, whereas for SBP in 32 of 33 subjects, two out of three measurements were within this range. Conclusion: The custo screen pediatric met all criteria of the ESH-IP review 2010, modified for children from 3 to about 12 years, and can be recommended for ABPM in children.
引用
收藏
页码:573 / 580
页数:8
相关论文
共 50 条
  • [1] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 573 - 580
  • [2] Erratum to: Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 835 - 836
  • [3] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010 (vol 176, pg 573, 2017)
    Beime, Beate
    Deutsch, Cornelia
    Kruger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Hammel, Gertrud
    Bramlage, Peter
    EUROPEAN JOURNAL OF PEDIATRICS, 2017, 176 (06) : 835 - 836
  • [4] Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010
    Bramlage, Peter
    Deutsch, Cornelia
    Krueger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Zwingers, Thomas
    Beime, Beate
    Mengden, Thomas
    VASCULAR HEALTH AND RISK MANAGEMENT, 2014, 10 : 303 - 309
  • [5] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    M Abou-Dakn
    C Döhmen
    S Wenzel
    Journal of Human Hypertension, 2017, 31 : 89 - 92
  • [6] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    Abou-Dakn, M.
    Doehmen, C.
    Wenzel, S.
    JOURNAL OF HUMAN HYPERTENSION, 2017, 31 (02) : 89 - 92
  • [7] Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Bilo, Grzegorz
    Zorzi, Cristina
    Munera, Juan E. Ochoa
    Torlasco, Camilla
    Giuli, Valentina
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 291 - 294
  • [8] Validation of the SCIAN LD-735 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2016, 21 (04) : 255 - 258
  • [9] Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (05) : 204 - 206
  • [10] Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Liu, Chang-Yuan
    Li, Yan
    Wang, Ji-Guang