Thermal changes and drill wear in bovine bone during implant site preparation. A comparative in vitro study: twisted stainless steel and ceramic drills

被引:90
作者
Oliveira, Natalia [1 ]
Alaejos-Algarra, Fernando [2 ]
Mareque-Bueno, Javier [2 ]
Ferres-Padro, Eduard [2 ]
Hernandez-Alfaro, Federico [2 ]
机构
[1] Univ Int Catalunya, Fac Dent, Int Program Implant Dent Surg & Prosthodont, Barcelona 08195, Spain
[2] Univ Int Catalunya, Fac Dent, Dept Oral Med Oral & Maxillofacial Surg & Implant, Barcelona 08195, Spain
关键词
bone drilling; ceramic drill; dental implants; heat generation; implant drills; implant preparation; oxide zirconia drill; thermal osseonecrosis; thermocouple; SELF-TAPPING SCREWS; HEAT-GENERATION; CORTICAL BONE; TEMPERATURE; SPEED; IRRIGATION; PARAMETERS; RABBIT; FORCE;
D O I
10.1111/j.1600-0501.2011.02248.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The purpose was to assess thermal changes and drill wear in bovine bone tissue with the use of twisted stainless steel and zirconia-based drills, during implant site preparation. Methods: A total of 100 implant site preparations were performed on bovine ribs using a surgical unit linked to a testing device, in order to standardize/simulate implant drilling procedures. Bone temperature variations and drilling force were recorded when drilling at a depth of 8 and 10 mm. A constant irrigation of 50 ml/min. (21 +/- 1 degrees C) and drilling speed of 800 r.p.m. were used. Scanning electron microscopy analysis was preformed prior and after drilling. Results: Mean temperature increase with both drills at 8 mm was 0.9 degrees C and at 10 mm was 2 degrees C (P<0.0001). Statistical significant higher bone temperatures were obtained with stainless steel drill (1.6 degrees C), when comparing with the ceramic drill (1.3 degrees C) (P<0.05). Temperature increase was correlated with higher number of perforations (P<0.05) and drilling load applied. There was no significant association between drilling force applied and temperature increase by either drill or at either depth. No severe signs of wear of either drill were detected after 50 uses. Conclusions: Drill material and design, number of uses, depth and drilling load applied appear to influence bone temperature variations during implant site preparation. Drilling depth was a predominant factor in bone temperature increase. Both drills can be used up to 50 times without producing harmful temperatures to bone tissue or severe signs of wear and deformation. To cite this article:? Oliveira N, Alaejos-Algarra F, Mareque-Bueno J, Ferres-Padro E, Hernandez-Alfaro F. Thermal changes and drill wear in bovine bone during implant site preparation. A comparative in vitro study: twisted stainless steel and ceramic drills?Clin. Oral Impl. Res. 23, 2012; 963969?doi: 10.1111/j.1600-0501.2011.02248.x
引用
收藏
页码:963 / 969
页数:7
相关论文
共 40 条
[1]  
Abouzgia M. B., 1996, J ORAL MAXILLOFACIAL, V53, P1315
[2]   MEASUREMENTS OF SHAFT SPEED WHILE DRILLING THROUGH BONE [J].
ABOUZGIA, MB ;
JAMES, DF .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1995, 53 (11) :1308-1315
[3]   Effect of drill speed on bone temperature [J].
Abouzgia, MB ;
Symington, JM .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1996, 25 (05) :394-399
[4]  
Abouzgia MB, 1997, INT J ORAL MAX IMPL, V12, P342
[5]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[6]  
Albrektsson T, 1985, Clin Orthop Relat Res, P311
[7]   Effects of repeated drill use on temperature of bone during preparation for osteosynthesis self-tapping screws [J].
Allan, W ;
Williams, ED ;
Kerawala, CJ .
BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2005, 43 (04) :314-319
[8]   Thermal osteonecrosis and bone drilling parameters revisited [J].
Augustin, Goran ;
Davila, Slavko ;
Mihoci, Kristijan ;
Udiljak, Toma ;
Vedrina, Denis Stjepan ;
Antabak, Anko .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2008, 128 (01) :71-77
[9]   The effects of drilling force on cortical temperatures and their duration: an in vitro study [J].
Bachus, KN ;
Rondina, MT ;
Hutchinson, DT .
MEDICAL ENGINEERING & PHYSICS, 2000, 22 (10) :685-691
[10]  
Bayerfein T., 2006, Folia Morphologica, V65, P72