In vitro evaluation of a novel pulsatile right heart assist device - the PERKAT system

被引:10
作者
Kretzschmar, Daniel [1 ]
Lauten, Alexander [2 ]
Ferrari, Markus W. [3 ]
机构
[1] Univ Jena, Jena Univ Hosp, Dept Internal Med 1, Jena, Germany
[2] Charite, Dept Cardiol, D-13353 Berlin, Germany
[3] Helios Kliniken, Clin Internal Med 1, HSK, Wiesbaden, Germany
关键词
Right ventricular failure; Right heart failure; Right ventricular mechanical support; Percutaneous right ventricular assist device; Ventricular assist device; RIGHT-VENTRICULAR FAILURE; MECHANICAL CIRCULATORY SUPPORT; PULMONARY ARTERIAL-HYPERTENSION; CARDIOGENIC-SHOCK; SYSTOLIC FUNCTION; INFARCTION; MANAGEMENT; EXCLUSION; REGISTRY; EVENTS;
D O I
10.5301/ijao.5000440
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: Acute right ventricular failure is a life-threatening condition with poor prognosis. It occurs as a result of right ventricular infarction, postcardiac transplantation, or postimplantation of a left ventricular assist device. Temporary mechanical right ventricular support could be a reasonable treatment option. Therefore, we developed a novel percutaneously implantable device. Methods: The PERKAT device consists of a self-expandable chamber covered with multiple inflow valves carrying foils. A flexible outlet tube with a pigtail tip is attached to the distal end. PERKAT is designed for percutaneous implantation through the femoral vein (18 French sheath). The chamber is expanded in the inferior vena cava while the outlet tube bypasses the right heart and the pigtail tip is lying in the pulmonary trunk. An IABP balloon is inserted into the chamber and connected to an IABP console. Balloon deflation generates blood flow from the vena cava into the chamber through the foil valves. During inflation blood is pumped through the tube into the pulmonary arteries. Results: In vitro experiments were performed using 30 mL and 40 mL IABP balloons. IABP inflation/deflation times were set to 80, 90, 100, and 110 per min with an afterload of 22 mmHg and 44 mmHg. PERKAT generated flow rates between 1.6 to 3.1 l/min, depending on balloon size, pump cycle, and afterload. Conclusions: The novel percutaneously implantable right ventricular assist device offers emergency support of up to 3 l/min. Based on the successful in vitro evaluation, we recommend the system as a promising approach for treatment of patients in need of RV support.
引用
收藏
页码:537 / 541
页数:5
相关论文
共 21 条
[11]   2014 ESC Guidelines on the diagnosis and management of acute pulmonary embolism The Task Force for the Diagnosis and Management of Acute Pulmonary Embolism of the European Society of Cardiology (ESC) [J].
Konstantinides, Stavros V. ;
Torbicki, Adam ;
Agnelli, Giancarlo ;
Danchin, Nicolas ;
Fitzmaurice, David ;
Galie, Nazzareno ;
Gibbs, J. Simon R. ;
Huisman, Menno V. ;
Humbert, Marc ;
Kucher, Nils ;
Lang, Irene ;
Lankeit, Mareike ;
Lekakis, John ;
Maack, Christoph ;
Mayer, Eckhard ;
Meneveau, Nicolas ;
Perrier, Arnaud ;
Pruszczyk, Piotr ;
Rasmussen, Lars H. ;
Schindler, Thomas H. ;
Svitil, Pavel ;
Noordegraaf, Anton Vonk ;
Zamorano, Jose Luis ;
Zompatori, Maurizio .
EUROPEAN HEART JOURNAL, 2014, 35 (43) :3033-3080
[12]   Neurological events during long-term mechanical circulatory support for heart failure - The Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure (REMATCH) experience [J].
Lazar, RM ;
Shapiro, PA ;
Jaski, BE ;
Parides, MK ;
Bourge, RC ;
Watson, JT ;
Damme, L ;
Dembitsky, W ;
Hosenpud, JD ;
Gupta, L ;
Tierney, A ;
Kraus, T ;
Naka, Y .
CIRCULATION, 2004, 109 (20) :2423-2427
[13]   Percutaneous left ventricular assistance in post cardiac arrest shock: Comparison of intra aortic blood pump and IMPELLA Recover LP2.5 [J].
Manzo-Silberman, Stephane ;
Fichet, Jerome ;
Mathonnet, Armelle ;
Varenne, Olivier ;
Ricome, Sylvie ;
Chaib, Aures ;
Zuber, Benjamin ;
Spaulding, Christian ;
Cariou, Alain .
RESUSCITATION, 2013, 84 (05) :609-615
[14]   Mechanical support for isolated right ventricular failure in patients after cardiotomy [J].
Moazami, N ;
Pasque, MK ;
Moon, MR ;
Herren, RL ;
Bailey, MS ;
Lawton, JS ;
Damiano, RJ .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2004, 23 (12) :1371-1375
[15]   PulseCath iVAC 3L™ hemodynamic performance for simple assisted flow [J].
Niclauss, Lars ;
von Segesser, Ludwig Karl .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2011, 12 (06) :912-913
[16]   Long-term use of a left ventricular assist device for end-stage heart failure [J].
Rose, EA ;
Gelijns, AC ;
Moskowitz, AJ ;
Heitjan, DF ;
Stevenson, LW ;
Dembitsky, W ;
Long, JW ;
Ascheim, DD ;
Tierney, AR ;
Levitan, RG ;
Watson, JT ;
Meier, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 345 (20) :1435-1443
[17]  
Rudiger A, 2012, SCHWEIZ MED FORUM, V12, P364
[18]   Left ventricular contributions to right ventricular systolic function during LVAD support [J].
Santamore, WP ;
Gray, LA .
ANNALS OF THORACIC SURGERY, 1996, 61 (01) :350-356
[19]   Feasibility and efficacy of bypassing the right ventricle and pulmonary circulation to treat right ventricular failure: an experimental study [J].
Spillner, Jan ;
Stoppe, Christian ;
Hatam, Nima ;
Amerini, Andrea ;
Menon, Ares ;
Nix, Christoph ;
Steinseifer, Ulrich ;
Abusabha, Yousef ;
Giessen, Hanna ;
Autschbach, Ruediger ;
Haushofer, Marcus .
JOURNAL OF CARDIOTHORACIC SURGERY, 2012, 7
[20]   Total right ventricular exclusion improves left ventricular function in patients with end-stage congestive right ventricular failure [J].
Takagaki, M ;
Ishino, K ;
Kawada, M ;
Ohtsuki, S ;
Hirota, M ;
Tedoriya, T ;
Tanabe, Y ;
Nakai, M ;
Sano, S .
CIRCULATION, 2003, 108 (10) :226-229