Dosimetry of high intensity electron beams produced with dedicated accelerators in Intra-Operative Radiation Therapy (IORT)

被引:0
|
作者
Lamanna, E. [1 ,6 ]
Fiorillo, A. S. [1 ,6 ]
Bruno, C. [2 ]
Santaniello, A. [2 ]
Siaka, Y. F. T. [2 ,3 ]
Berdondini, A. [4 ]
Bettuzzi, M. [4 ]
Brancaccio, R. [4 ]
Casali, F. [4 ]
Morigi, M. P. [4 ]
Barca, G. [5 ]
Castrovillari, F. [5 ]
机构
[1] Magna Graecia Univ Catanzaro, Fac Med, I-88100 Catanzaro, Italy
[2] Univ Calabria, Dept Phys, Cosenza, Italy
[3] Univ Douala, Ctr Atom Mol Phys & Quantum Opt, Douala, Cameroon
[4] Univ Bologna, Dept Phys, INFN, Bologna, Italy
[5] Azienda Ospedaliera, Med Phys, Cosenza, Italy
[6] INFN, Gruppo Collegato, Cosenza, Italy
关键词
IORT; dosimetry; scintillating fibers; depthdose; measurement; 3D;
D O I
10.1109/NSSMIC.2007.4436435
中图分类号
O59 [应用物理学];
学科分类号
摘要
The technique of High Dose Intra-Operative Radiation Therapy (HDR-IORT) consists in the delivery of irradiation immediately following the removal of a cancerous mass, where the same incision is used to direct the radiation to the tumour bed. Given its particular characteristics, IORT requires dose measurements that are different from those requested in external radiotherapy treatments. The main reason lies in the fact that in this case a single high dose must be delivered to a volume target whose extension and depths Will be determined directly during the operation. Since the possibility of devising a treatment plan using a TPS (Treatment Planning System) is not available, it is necessary to know the physical and geometric characteristics of the beam. Defining the physical characteristics of the beam entails both measuring the delivered dose and defining (monitoring) procedures. In any case a much higher dose will be released than occurs with conventional external accelerators. The ionization chamber recommended by the standard protocols for radiotherapy cannot be used because of the ion recombination inside the gas. In this work we propose the use of a calorimetric phantom, the Dosiort, to measure the beam properties. We describe the main characteristics and some preliminary results of the Dosiort System, which is proposed within the framework of a research project of the INFN (Italian National Institute of Nuclear Physics). The set-up is a solid phantom of density approaching 1 g/cc with sensitive layers of scintillating fibres at fixed a position in a calorimetric configuration for the containment of electrons of energy 4-12 MeV. The prototype will be able to define the physical and geometrical characteristics of the electron beam (quality, isotropy, homogeneity, etc) and to measure the parameters needed to select the energy, the intensity and the Monitor Units (MU) for the exposition: Percentage Depth Dose; Beam profiles; Isodose curves; Values of dose for MU.
引用
收藏
页码:732 / 738
页数:7
相关论文
共 50 条
  • [1] Dosimetry of High Intensity Electron Beams Produced by Dedicated Accelerators in Intra-Operative Radiation Therapy (IORT)
    Lamanna, Ernesto
    Fiorillo, Antonino Secondo
    Bruno, Carlo
    Santaniello, Anna
    Siaka, Yvette Flore Tchuente
    Berdondini, Andrea
    Bettuzzi, Matteo
    Brancaccio, Rosa
    Casali, Franco
    Morigi, Maria Pia
    Barca, Giuliana
    Castrovillari, Francesca
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (01) : 66 - 72
  • [2] Radiobiological characterization of the very high dose rate and dose per pulse electron beams produced by an IORT (intra operative radiation therapy) dedicated linac
    Scampoli, Paola
    Carpentieri, Carmela
    Giannelli, Marco
    Magaddino, Vera
    Manti, Lorenzo
    Moriello, Carmen
    Piliero, Maria Antonietta
    Righi, Sergio
    Di Martino, Fabio
    TRANSLATIONAL CANCER RESEARCH, 2017, 6 : S761 - S768
  • [3] Results of IORT (Intra-Operative Radiation Therapy) trials
    Orecchia, R.
    RADIOTHERAPY AND ONCOLOGY, 2021, 158 : S59 - S59
  • [4] MONTE CARLO SIMULATION OF THE ELECTRON BEAMS PRODUCED BY A LINEAR ACCELERATOR FOR INTRA-OPERATIVE RADIATION THERAPY
    Mihailescu, D.
    Borcia, C.
    ROMANIAN REPORTS IN PHYSICS, 2014, 66 (01) : 61 - 74
  • [5] Intra-operative radiation therapy (IORT) in recurrent head and neck cancer
    De Felice, Francesca
    Musio, Daniela
    Tombolini, Vincenzo
    ORAL ONCOLOGY, 2016, 55 : E1 - E1
  • [6] Study and development of an innovative electron beam imaging system for dosimetry in IORT (Intra Operative Radiation Therapy)
    Brancaccio, R
    Bettuzzi, M
    Casali, F
    Cornacchia, S
    Morigi, MP
    Pasini, A
    Ronsivalle, C
    Tata, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 562 (02): : 1020 - 1023
  • [7] Real-time system for dosimetry in IORT (intra operative radiation therapy)
    Brancaccio, R.
    Bettuzzi, M.
    Casali, F.
    Cornacchia, S.
    Morigi, M.
    Pasini, A.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (05) : 2844 - 2848
  • [8] Study and development of two innovative electron beam imaging systems for dosimetry in IORT (Intra Operative Radiation Therapy)
    Brancaccio, R
    Bettuzzi, M
    Casali, F
    Cornacchia, S
    Monigi, MP
    Pasini, A
    Romani, D
    Ronsivalle, C
    Tata, A
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2561 - 2564
  • [9] General purpose MOSFETs for the dosimetry of electron beams used in intra-operative radiotherapy
    Martinez-Garcia, M. S.
    Simancas, F.
    Palma, A. J.
    Lallena, A. M.
    Banqueri, J.
    Carvajal, M. A.
    SENSORS AND ACTUATORS A-PHYSICAL, 2014, 210 : 175 - 181