Sensitive determination of free benzophenone-3 in human urine samples based on an ionic liquid as extractant phase in single-drop microextraction prior to liquid chromatography analysis

被引:110
作者
Vidal, Lorena [1 ]
Chisvert, Alberto [1 ]
Canals, Antonio [1 ]
Salvador, Amparo [2 ]
机构
[1] Univ Alicante, Fac Ciencias, Dept Quim Analit, E-03080 Alicante, Spain
[2] Univ Valencia, Fac Quim, Dept Quim Analit, E-46100 Burjassot, Spain
关键词
benzophenone-3; UV filter; sunscreen; cosmetics; urine; single-drop microextraction (SDME); ionic liquid; multivariate optimization; liquid chromatography;
D O I
10.1016/j.chroma.2007.07.077
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Benzophenone-3 (BZ3), one of the compounds most commonly used as UV filter in cosmetic products, can be absorbed through the skin into the human body, since it can be found at trace levels in urine from users of cosmetic products that contain BZ3. Moreover, different undesirable effects have been attributed to this compound. Thus, sensitive analytical methods to monitor urinary excretion of this compound should be developed. This paper presents a selective and sensitive methodology for BZ3 determination at ultratrace levels in human urine samples. The methodology is based on a novel microextraction technique, known as single-drop microextraction (SDME). An ionic liquid (IL) has been used as extractant phase instead of an organic solvent. After the microextraction process, the extractant phase was injected into a liquid chromatography system. The variables of interest in the SDME process were optimized using a multivariate optimization approach. A Plackett-Burman design for screening and a circumscribed central composite design for optimizing the significant variables were applied. Ionic strength, extraction time, stirring speed, pH, ionic liquid type, drop volume and sample volume were the variables studied. The optimum experimental conditions found were: sodium chloride concentration, 13% (w/v); extraction time, 25 min; stirring speed, 900 rpm; pH, 2; ionic liquid type, 1-hexyl-3-methylimidazolium hexafluorophosphate ([C6MIM][PF6); drop volume, 5 mu L; and sample volume, 10 mL. The proposed method requires a standard addition calibration approach, and it has been successfully employed to determine free BZ3 in urine samples coming from human volunteers who applied a sunscreen cosmetic containing this UV filter. The limit of detection was in the order of 1.3 ng mL(-1) and repeatability of the method, expressed as relative standard deviation, was 6% (n = 8). (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:95 / 103
页数:9
相关论文
共 37 条
[1]   DEVELOPMENT AND VALIDATION OF A HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF BENZOPHENONE-3 IN RATS [J].
ABDELNABI, IM ;
KADRY, AM ;
DAVIS, RA ;
ABDELRAHMAN, MS .
JOURNAL OF APPLIED TOXICOLOGY, 1992, 12 (04) :255-259
[2]   Sequential-injection determination of traces of disodium phenyl dibenzimidazole tetrasulphonate in urine from users of sunscreens by on-line solid-phase extraction coupled with a fluorimetric detector [J].
Balaguer, A ;
Chisvert, A ;
Salvador, A .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 40 (04) :922-927
[3]   Percutaneous absorption of Mexoryl SX® in human volunteers:: Comparison with in vitro data [J].
Benech-Kieffer, F ;
Meuling, WJA ;
Leclerc, C ;
Roza, L ;
Leclaire, J ;
Nohynek, G .
SKIN PHARMACOLOGY AND APPLIED SKIN PHYSIOLOGY, 2003, 16 (06) :343-355
[4]  
Chisvert A, 2007, ANALYSIS OF COSMETIC PRODUCTS, P83, DOI 10.1016/B978-044452260-3/50028-0
[5]   Robustness testing for a capillary electrophoresis method using the "short-end injection" technique [J].
Fabre, H ;
Mesplet, N .
JOURNAL OF CHROMATOGRAPHY A, 2000, 897 (1-2) :329-338
[6]   Determination of benzophenone-3 and metabolites in water and human urine by solid-phase microextraction and quadrupole ion trap GC-MS [J].
Felix, T ;
Hall, BJ ;
Brodbelt, JS .
ANALYTICA CHIMICA ACTA, 1998, 371 (2-3) :195-203
[7]   UV filters: From sunscreens to human body and the environment [J].
Giokas, Dimosthenis L. ;
Salvador, Amparo ;
Chisvert, Alberto .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2007, 26 (05) :360-374
[8]   Percutaneous absorption of the sunscreen benzophenone-3 after repeated whole-body applications, with and without ultraviolet irradiation [J].
Gonzalez, H ;
Farbrot, A ;
Larkö, O ;
Wennberg, AM .
BRITISH JOURNAL OF DERMATOLOGY, 2006, 154 (02) :337-340
[9]   Percutaneous absorption of benzophenone-3, a common component of topical sunscreens [J].
Gonzalez, HG ;
Farbrot, A ;
Larkö, O .
CLINICAL AND EXPERIMENTAL DERMATOLOGY, 2002, 27 (08) :691-694
[10]   Systemic absorption of sunscreen after topical application [J].
Hayden, CGJ ;
Roberts, MS ;
Benson, HAE .
LANCET, 1997, 350 (9081) :863-864