Safety Analysis of Medical Implants in the Human Head Exposed to a Wireless Power Transfer System

被引:18
作者
Shah, Izaz Ali [1 ]
Basir, Abdul [1 ]
Cho, Youngdae [1 ]
Yoo, Hyoungsuk [1 ,2 ]
机构
[1] Hanyang Univ, Dept Elect Engn, Seoul 04763, South Korea
[2] Hanyang Univ, Dept Biomed Engn, Seoul 04763, South Korea
关键词
Implants; Magnetic heads; Biomedical imaging; Coils; Brain modeling; Lead; Bones; Exposure evaluation; human model; medical implants; portable electronics; specific absorption rate (SAR); wireless power transfer; METALLIC IMPLANTS; HUMAN-BODY; DESIGN; COILS; MRI; SAR; DEVICE;
D O I
10.1109/TEMC.2022.3144468
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The interactions of medical implants in the human body with electromagnetic fields from newly introduced high-field technologies such as near-field wireless power transfer (WPT) are of great concern. In this study, the effects of implants on the specific absorption rate (SAR) in a head model were computationally evaluated in the immediate vicinity of a WPT system operating at 6.78 MHz with a transferred power of 50 W. The SAR in the head model located 30 mm above the WPT system was evaluated with a skull plate, bone plates (different shapes), miniplate, fixtures, and a deep brain stimulator. The results indicated that even small implants had notable effects on the SAR distribution and corresponding local peak and mass-averaged SAR values. Apart from the size and shape of the implant, the crucial factor influencing the SAR was the implant's position with respect to the WPT system. Therefore, various position configurations were simulated for the WPT system to determine the worst case scenario for each specific implant. Although the local peak SAR value was increased by a factor of 600 in the worst-case scenario (for the skull plate), the SAR$_{\text{1}g}$ and SAR$_{\text{10}g}$ were only increased by factors of seven and four, respectively. Compliance with international safety limits was studied, followed by computing the maximum allowable transmit power (MATP). It was found that, without the implant, the MATP satisfying SAR$_{\text{1}g}$ was 43 W for the designed WPT system, and was reduced to 6.9 W in the presence of the skull plate.
引用
收藏
页码:640 / 649
页数:10
相关论文
共 32 条
[11]   A Novel All-Direction Antimisalignment Wireless Power Transfer System Designed by Truncated Region Eigenfunction Expansion Method [J].
Dong, Zifan ;
Li, Xiaoming ;
Liu, Sheng ;
Xu, Ziwei ;
Yang, Lin .
IEEE TRANSACTIONS ON POWER ELECTRONICS, 2021, 36 (11) :12456-12467
[12]   Transmitter Coils Design for Free-Positioning Omnidirectional Wireless Power Transfer System [J].
Feng, Junjie ;
Li, Qiang ;
Lee, Fred C. ;
Fu, Minfan .
IEEE TRANSACTIONS ON INDUSTRIAL INFORMATICS, 2019, 15 (08) :4656-4664
[13]   Feasibility of Using Linearly Polarized Rotating Birdcage Transmitters and Close-Fitting Receive Arrays in MRI to Reduce SAR in the Vicinity of Deep Brain Simulation Implants [J].
Golestanirad, Laleh ;
Keil, Boris ;
Angelone, Leonardo M. ;
Bonmassar, Giorgio ;
Mareyam, Azma ;
Wald, Lawrence L. .
MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (04) :1701-1712
[14]   Development of a new generation of high-resolution anatomical models for medical device evaluation: the Virtual Population 3.0 [J].
Gosselin, Marie-Christine ;
Neufeld, Esra ;
Moser, Heidi ;
Huber, Eveline ;
Farcito, Silvia ;
Gerber, Livia ;
Jedensjoe, Maria ;
Hilber, Isabel ;
Di Gennaro, Fabienne ;
Lloyd, Bryn ;
Cherubini, Emilio ;
Szczerba, Dominik ;
Kainz, Wolfgang ;
Kuster, Niels .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (18) :5287-5303
[15]   Computational and experimental investigation of RF-induced heating for multiple orthopedic implants [J].
Guo, Ran ;
Zheng, Jianfeng ;
Wang, Yu ;
Zeng, Qi ;
Wang, Qingyan ;
Yang, Rui ;
Kainz, Wolfgang ;
Chen, Ji .
MAGNETIC RESONANCE IN MEDICINE, 2019, 82 (05) :1848-1858
[16]   Free-Positioning Wireless Power Transfer Using a 3D Transmitting Coil for Portable Devices [J].
Ha-Van, Nam ;
Le-Huu, Hoang ;
Minh Thuy Le ;
Park, Kwangsuk ;
Seo, Chulhun .
JOURNAL OF ELECTROMAGNETIC ENGINEERING AND SCIENCE, 2020, 20 (04) :270-276
[17]  
Hayat S., 2021, IEEE ACCESS, V9
[18]   Gaps in Knowledge Relevant to the "Guidelines for Limiting Exposure to Time-Varying Electric and Magnetic Fields (1 Hz-100 kHz)" [J].
Ziegelberger G. ;
Marino C. ;
Croft R. ;
Feychting M. ;
Green A.C. ;
Hirata A. ;
d'Inzeo G. ;
Miller S. ;
Oftedal G. ;
Okuno T. ;
Röösli M. ;
Sienkiewicz Z. ;
van Rongen E. ;
Watanabe S. ;
Lagroye I. ;
Karipidis K. ;
Mattsson M.-O. ;
Wood A. .
HEALTH PHYSICS, 2020, 118 (05) :533-542
[19]   Evaluations of the MRI RF-Induced Heating for Helical Stents Under a 1.5T MRI System [J].
Ji, Xiaohe ;
Zheng, Jianfeng ;
Yang, Rui ;
Kainz, Wolfgang ;
Chen, Ji .
IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2019, 61 (01) :57-64
[20]   Design Methodology of a Printed WPT System for HF-Band Mid-Range Applications Considering Human Safety Regulations [J].
Koohestani, Mohsen ;
Zhadobov, Maxim ;
Ettorre, Mauro .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2017, 65 (01) :270-279