摘要
In proton therapy, the lateral fall-off is often used to spare critical organs. It is therefore crucial to improve the penumbra for proton pencil beam scanning. However, previous work has shown that collimation may not be necessary for depths of >15 cm in water. As such, in this work we investigate the effectiveness of a thin multi leaf collimator (just thick enough to completely stop protons with ranges of <15 cm in water) for energy layer specific collimation in patient geometries, when applied in combination with both grid and contour scanned PBS proton therapy. For this, an analytical model of collimated beam shapes, based solely on data available in the treatment planning system, has been included in the optimization, with the resulting optimised plans then being recalculated using Monte Carlo in order to most accurately simulate the full physics effects of the collimator. For grid based scanning, energy specific collimation has been found to reduce the V30 outside the PTV by 19.8% for an example patient when compared to the same pencil beam placement without collimation. V30 could be even reduced by a further 5.6% when combining collimation and contour scanning. In addition, mixed plans, consisting of contour scanning for deep fields (max range >15 cm WER) and collimated contour scanning for superficial fields (<15 cm), have been created for four patients, by which V30 could be reduced by 0.8% to 8.0% and the mean dose to the brain stem by 1.5% to 3.3%. Target dose homogeneity however is not substantially different when compared to the best un-collimated scenario. In conclusion, we demonstrate the potential advantages of a thin, multi leaf collimator in combination with contour scanning for energy layer specific collimation in PBS proton therapy.
引用
收藏
页数:15
相关论文
共 50 条
  • [41] Pitfalls in the beam modelling process of Monte Carlo calculations for proton pencil beam scanning
    Winterhalter, Carla
    Aitkenhead, Adam
    Oxley, David
    Richardson, Jenny
    Weber, Damien C.
    MacKay, Ranald, I
    Lomax, Antony J.
    Safai, Sairos
    BRITISH JOURNAL OF RADIOLOGY, 2020, 93 (1107)
  • [42] Evaluation and second check of a commercial Monte Carlo dose engine for small-field apertures in pencil beam scanning proton therapy
    Holmes, Jason
    Shen, Jiajian
    Shan, Jie
    Patrick, Christopher L.
    Wong, William W.
    Foote, Robert L.
    Patel, Samir H.
    Bues, Martin
    Liu, Wei
    MEDICAL PHYSICS, 2022, 49 (05) : 3497 - 3506
  • [43] Absolute dosimetry for FLASH proton pencil beam scanning radiotherapy
    Lourenco, Ana
    Subiel, Anna
    Lee, Nigel
    Flynn, Sam
    Cotterill, John
    Shipley, David
    Romano, Francesco
    Speth, Joe
    Lee, Eunsin
    Zhang, Yongbin
    Xiao, Zhiyan
    Mascia, Anthony
    Amos, Richard A.
    Palmans, Hugo
    Thomas, Russell
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [44] Determination of surface dose in pencil beam scanning proton therapy
    Kern, A.
    Baeumer, C.
    Kroeninger, K.
    Mertens, L.
    Timmermann, B.
    Walbersloh, J.
    Wulff, J.
    MEDICAL PHYSICS, 2020, 47 (05) : 2277 - 2288
  • [45] Validation and clinical implementation of an accurate Monte Carlo code for pencil beam scanning proton therapy
    Huang, Sheng
    Kang, Minglei
    Souris, Kevin
    Ainsley, Christopher
    Solberg, Timothy D.
    McDonough, James E.
    Simone, Charles B.
    Lin, Liyong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 558 - 572
  • [46] Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy
    Han, Rong-Cheng
    Li, Yong-Jiang
    Pu, Yue-Hu
    NUCLEAR SCIENCE AND TECHNIQUES, 2020, 31 (02)
  • [47] Validating a Monte Carlo approach to absolute dose quality assurance for proton pencil beam scanning
    Winterhalter, C.
    Fura, E.
    Tian, Y.
    Aitkenhead, A.
    Bolsi, A.
    Dieterle, M.
    Fredh, A.
    Meier, G.
    Oxley, D.
    Siewert, D.
    Weber, D. C.
    Lomax, A.
    Safai, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (17)
  • [48] Neutrons in proton pencil beam scanning: parameterization of energy, quality factors and RBE
    Schneider, Uwe
    Halg, Roger A.
    Baiocco, Giorgio
    Lomax, Tony
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (16) : 6231 - 6242
  • [49] An advanced junction concept in pediatric craniospinal irradiation by proton pencil beam scanning
    Fellin, Francesco
    Fracchiolla, Francesco
    Rombi, Barbara
    Lipparini, Mirko
    Vennarini, Sabina
    Farace, Paolo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 60 : 168 - 173
  • [50] Utility of a patient-specific aperture collimator and multi-leaf collimator in line scanning proton therapy
    Ko, Mincheol
    Jeon, Chanil
    Ahn, Sung Hwan
    Han, Youngyih
    Chung, Kwangzoo
    Cho, Sungkoo
    Jo, Kwanghyun
    Shin, Eun Hyuk
    Lee, Sehyoung
    Lee, Woojin
    Oh, Yoonjin
    Kim, Juhye
    Kang, Seokyoon
    Park, Hee Chul
    Lee, Kisung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2024, 85 (09) : 751 - 762