Contour scanning, multi-leaf collimation and the combination thereof for proton pencil beam scanning

被引:13
|
作者
Winterhalter, Carla [1 ,2 ]
Meier, Gabriel [1 ]
Oxley, David [1 ]
Weber, Damien C. [1 ,3 ,4 ]
Lomax, Antony J. [1 ,2 ]
Safai, Sairos [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, Villigen, Switzerland
[2] Swiss Fed Inst Technol, Phys Dept, Zurich, Switzerland
[3] Univ Hosp Bern, Radiat Oncol Dept, Bern, Switzerland
[4] Univ Hosp Zurich, Radiat Oncol Dept, Zurich, Switzerland
关键词
proton therapy; pencil beam scanning; penumbra; collimation; optimisation; PATIENT-SPECIFIC APERTURE; MONTE-CARLO; DYNAMIC COLLIMATION; THERAPY; SYSTEM; RECONSTRUCTION; PENUMBRA; QUALITY; IMPACT;
D O I
10.1088/1361-6560/aaf2e8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In proton therapy, the lateral fall-off is often used to spare critical organs. It is therefore crucial to improve the penumbra for proton pencil beam scanning. However, previous work has shown that collimation may not be necessary for depths of >15 cm in water. As such, in this work we investigate the effectiveness of a thin multi leaf collimator (just thick enough to completely stop protons with ranges of <15 cm in water) for energy layer specific collimation in patient geometries, when applied in combination with both grid and contour scanned PBS proton therapy. For this, an analytical model of collimated beam shapes, based solely on data available in the treatment planning system, has been included in the optimization, with the resulting optimised plans then being recalculated using Monte Carlo in order to most accurately simulate the full physics effects of the collimator. For grid based scanning, energy specific collimation has been found to reduce the V30 outside the PTV by 19.8% for an example patient when compared to the same pencil beam placement without collimation. V30 could be even reduced by a further 5.6% when combining collimation and contour scanning. In addition, mixed plans, consisting of contour scanning for deep fields (max range >15 cm WER) and collimated contour scanning for superficial fields (<15 cm), have been created for four patients, by which V30 could be reduced by 0.8% to 8.0% and the mean dose to the brain stem by 1.5% to 3.3%. Target dose homogeneity however is not substantially different when compared to the best un-collimated scenario. In conclusion, we demonstrate the potential advantages of a thin, multi leaf collimator in combination with contour scanning for energy layer specific collimation in PBS proton therapy.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] Theoretical Benefits of Dynamic Collimation in Pencil Beam Scanning Proton Therapy for Brain Tumors: Dosimetric and Radiobiological
    Moignier, Alexandra
    Gelover, Edgar
    Wang, Dongxu
    Smith, Blake
    Flynn, Ryan
    Kirk, Maura
    Lin, Liyong
    Solberg, Timothy
    Lin, Alexander
    Hyer, Daniel
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01): : 171 - 180
  • [32] Technical Note: An efficient daily QA procedure for proton pencil beam scanning
    Younkin, James E.
    Shen, Jiajian
    Bues, Martin
    Robertson, Daniel G.
    Mundy, Daniel W.
    Clouser, Edward
    Liu, Wei
    Ding, Xiaoning
    Stoker, Joshua B.
    MEDICAL PHYSICS, 2018, 45 (03) : 1040 - 1049
  • [33] Improving the precision and performance of proton pencil beam scanning
    Safai, Sairos
    Bula, Christian
    Meer, David
    Pedroni, Eros
    TRANSLATIONAL CANCER RESEARCH, 2012, 1 (03) : 196 - 206
  • [34] Clinical Implementation of Proton Therapy Using Pencil-Beam Scanning Delivery Combined With Static Apertures
    Baeumer, Christian
    Plaude, Sandija
    Khalil, Dalia Ahmad
    Geismar, Dirk
    Kramer, Paul-Heinz
    Kroninger, Kevin
    Nitsch, Christian
    Wulff, Joerg
    Timmermann, Beate
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [35] Automation of pencil beam scanning proton treatment planning for intracranial tumours
    Placidi, Lorenzo
    Righetto, Roberto
    Vecchi, Claudio
    Zara, Stefania
    Alparone, Alessandro
    Moretti, Roberto
    Amelio, Dante
    Scartoni, Daniele
    Schwarz, Marco
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 105
  • [36] Spot decomposition in a novel pencil beam scanning proton Computed Tomography
    Zhou, J.
    Li, X.
    Kabolizadeh, P.
    Yang, X.
    Yan, D.
    Stevens, C.
    Liu, T.
    Ding, X.
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [37] Clinical Outcome of Sacral Chordoma Patients Treated with Pencil Beam Scanning Proton Therapy
    Walser, M.
    Bojaxhiu, B.
    Kawashiro, S.
    Tran, S.
    Beer, J.
    Leiser, D.
    Pica, A.
    Bachtiary, B.
    Weber, D. C.
    CLINICAL ONCOLOGY, 2021, 33 (12) : E578 - E585
  • [38] Evaluation of the ray-casting analytical algorithm for pencil beam scanning proton therapy
    Winterhalter, Carla
    Zepter, Stefan
    Shim, Sojin
    Meier, Gabriel
    Bolsi, Alessandra
    Fredh, Anna
    Hrbacek, Jan
    Oxley, David
    Zhang, Ye
    Weber, Damien C.
    Lomax, Antony
    Safai, Sairos
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (06)
  • [39] A pre-absorber optimization technique for pencil beam scanning proton therapy treatments
    Fracchiolla, Francesco
    Fellin, Francesco
    Innocenzi, Marta
    Lipparini, Mirko
    Lorentini, Stefano
    Widesott, Lamberto
    Farace, Paolo
    Schwarz, Marco
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 : 145 - 152
  • [40] Proton pencil beam scanning craniospinal irradiation (CSI) with a single posterior brain beam: Dosimetry and efficiency
    Hu, Lei
    Zhai, Anna
    Chen, Qing
    Puri, Vandana
    Chen, Chin-Cheng
    Yu, Francis
    Fox, Jana
    Wolden, Suzanne
    Yang, Jonathan
    Simone II, Charles B.
    Lin, Haibo
    MEDICAL DOSIMETRY, 2024, 49 (01) : 25 - 29