Inappropriate shock delivery due to ventricular double detection with a biventricular pacing implantable cardioverter defibrillator

被引:15
作者
Schreieck, J
Zrenner, B
Kolb, C
Ndrepepa, G
Schmitt, C
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Med Klin 1, D-80636 Munich, Germany
[2] Tech Univ Munich, Deutsch Herzzentrum Munchen, Klin Herzkreislauferkrankungen, D-80636 Munich, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2001年 / 24卷 / 07期
关键词
ICDs; biventricular pacing; double detection; inappropriate discharges;
D O I
10.1046/j.1460-9592.2001.01154.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This report describes a patient with advanced heart failure, pronounced intraventricular conduction delay, and ventricular tachycardias who underwent implantation of a multisite pacing ICD. Pacing leads were placed in the right atrium, right ventricular apex, and to the left ventricular posterior wall via a coronary sinus vein. The system proved to have correct sensing and pacing function in an atria] synchronized biventricular pacing mode and an appropriate detection of ventricular fibrillation. However, 1 month after implantation the patient received an inappropriate shock delivery due to double detection of ventricular premature beats. The inherent detection problem of dual ventricular sensing is discussed.
引用
收藏
页码:1154 / 1157
页数:4
相关论文
共 3 条
[1]   Cardioverter-defibrillator oversensing due to double counting of ventricular tachycardia electrograms [J].
Boriani, G ;
Biffi, M ;
Frabetti, L ;
Parlapiano, M ;
Galli, R ;
Branzi, A ;
Magnani, B .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1998, 66 (01) :91-95
[2]  
Le Franc P, 1998, PACE, V21, P1672
[3]  
Stellbrink C, 1999, AM J CARDIOL, V83, p143D