Endourological Training Using 3D-Printed Bladder Phantoms: Development and Prospective Evaluation

被引:9
作者
Waldbillig, Frank [1 ,2 ]
von Rohr, Lennard [1 ]
Nientiedt, Malin [1 ]
Gruene, Britta [1 ,2 ]
Hein, Simon [2 ,3 ]
Suarez-Ibarrola, Rodrigo [2 ,3 ]
Miernik, Arkadiusz [2 ,3 ]
Ritter, Manuel [2 ,4 ]
Kriegmair, Maximilian C. [1 ,2 ]
机构
[1] Heidelberg Univ, Univ Med Ctr Mannheim, Dept Urol & Urosurg, Theodor Kutzer Ufer 1-3, D-68167 Mannheim, Germany
[2] German Fed Minist Educ & Res BMBF, RaVeNNA 4pi Consortium, Mannheim, Germany
[3] Univ Freiburg, Med Ctr, Fac Med, Dept Urol, Freiburg, Germany
[4] Univ Bonn, Univ Med Ctr Bonn, Dept Urol & Pediat Urol, Bonn, Germany
关键词
cystoscopy; endoscopy; surgical training; education; urology; simulator;
D O I
10.1089/end.2020.0900
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background: To create and evaluate a realistic, anatomically accurate, and user-friendly bladder phantom for reproducible endourological training purposes and endoscope mastery. Materials and Methods: The anatomy of full bladders was mapped from human computed tomography datasets. After a 3D model development process, content evidence and response process evidence (RPE) of the phantom were evaluated using the system usability scale (SUS), 5-point Likert scale questionnaires, and task execution of experienced urologists (U) and endoscopy-naive medical students (MS) in two training sessions (first vs second). Required validation cohort sizes (1:10) of the evaluating urologists (n = 12) and students (n = 115) were precalculated. Time measurements were recorded. Students were additionally evaluated by a validated global psychomotor assessment score (GPSS). Group comparisons were calculated by the Mann-Whitney U test. All tests were two sided with p < 0.05 considered statistically significant. Results: Content evidence was assessed by urologists with an "excellent" SUS score of 89.4 +/- 5.9 and an average "agreement" of >= 4 pts in the Likert scale questionnaires. RPE was assessed by intra- and intergroup time comparison for the execution of endoscopic tasks (cystoscopy [CY], guidewire insertion, and tumor biopsy). For CY, U: first 17.6 +/- 4.4 seconds vs second 12.4 +/- 2.0 seconds, p = 0.002; MS: first 56.6 +/- 28.2 seconds vs second 28.6 +/- 14.7 seconds, p < 0.001; U vs MS: first U 17.6 +/- 4.4 seconds vs first MS 56.6 +/- 28.2 seconds, p < 0.001, second U 12.4 +/- 2.0 seconds vs second MS 28.6 +/- 14.7 seconds, p < 0.001. Significant time differences were documented for all tasks and sessions (p < 0.001). Additionally, significant GPSS differences were recorded between the sessions (GPSS: first 20.4 +/- 5.1 pts vs second 24.7 +/- 4.0 pts, p < 0.001). Conclusions: Our low-fidelity 3D-printed bladder, called BladCap, is an easy-to-assemble, inexpensive, and robust phantom. We present data, which establish construct validity to support use as a clinical training device.
引用
收藏
页码:1257 / 1264
页数:8
相关论文
共 25 条
[11]   Validation of computer-based training in ureterorenoscopy [J].
Knoll, T ;
Trojan, L ;
Haecker, A ;
Alken, P ;
Michel, MS .
BJU INTERNATIONAL, 2005, 95 (09) :1276-1279
[12]   A call for the utilization of consensus standards in the surgical education literature [J].
Korndorffer, James R., Jr. ;
Kasten, Steven J. ;
Downing, Steven M. .
AMERICAN JOURNAL OF SURGERY, 2010, 199 (01) :99-104
[13]   The effect of bench model fidelity on endourological skills: A randomized controlled study [J].
Matsumoto, ED ;
Hamstra, SJ ;
Radomski, SB ;
Cusimano, MD .
JOURNAL OF UROLOGY, 2002, 167 (03) :1243-1247
[14]   Virtual reality ureteroscopy simulator as a valid tool for assessing endourological skills [J].
Matsumoto, Edward D. ;
Pace, Kenneth T. ;
Honey, R. John D'A .
INTERNATIONAL JOURNAL OF UROLOGY, 2006, 13 (07) :896-901
[15]   Evaluation of the Endo-Uro trainer for semi-rigid ureteroscopy training [J].
Palaneer, Sharanya ;
Aydin, Abdullatif ;
Al Janabi, Hasaneen ;
Al-Jabir, Ahmed ;
Macchione, Nicola ;
Khan, Muhammad Shamim ;
Dasgupta, Prokar ;
Ahmed, Kamran .
THERAPEUTIC ADVANCES IN UROLOGY, 2019, 11
[16]   The Learning Curve for Transurethral Resection of Bladder Tumour: How Many is Enough to be Independent, Safe and Effective Surgeon? [J].
Poletajew, Slawomir ;
Krajewski, Wojciech ;
Kaczmarek, Krystian ;
Kopczynski, Bartlomiej ;
Stamirowski, Remigiusz ;
Tukiendorf, Andrzej ;
Zdrojowy, Romuald ;
Slojewski, Marcin ;
Radziszewski, Piotr .
JOURNAL OF SURGICAL EDUCATION, 2020, 77 (04) :978-985
[17]   The impact of video games on training surgeons in the 21st century [J].
Rosser, James C., Jr. ;
Lynch, Paul J. ;
Cuddihy, Laurie ;
Gentile, Douglas A. ;
Klonsky, Jonathan ;
Merrell, Ronald .
ARCHIVES OF SURGERY, 2007, 142 (02) :181-186
[18]   Multi-institutional Surgical Education Interventions A Scoping Review [J].
Salles, Arghavan ;
Milam, Laurel ;
Sevdalis, Nick ;
Alseidi, Adnan ;
Mellinger, John ;
Stefanidis, Dimitrios ;
Nahmias, Jeffry ;
Kulaylat, Afif N. ;
Kim, Roger H. ;
Falcone, John L. ;
Arora, Tania K. ;
Phitayakorn, Roy ;
Cochran, Amalia .
ANNALS OF SURGERY, 2019, 270 (02) :257-269
[19]   Historical review of surgical simulation - A personal perspective [J].
Satava, Richard M. .
WORLD JOURNAL OF SURGERY, 2008, 32 (02) :141-148
[20]   Validation and implementation of surgical simulators: a critical review of present, past, and future [J].
Schout, B. M. A. ;
Hendrikx, A. J. M. ;
Scheele, F. ;
Bemelmans, B. L. H. ;
Scherpbier, A. J. J. A. .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2010, 24 (03) :536-546